| | Class 1 Recall Cook Medical Central Venous Catheter Tray, 5 Lumen Polyurethane |  |
| Date Posted |
February 23, 2011 |
| Recall Number |
Z-1115-2011 |
| Product |
Cook Medical Central Venous Catheter Tray, 5 Lumen Polyurethane, 10Fr/15cm, C-UQLMY-1001J-CCT, G32392, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN. |
| Code Information |
Part Number: C-UQLMY-1001J-CCT; Global Part Number G32392, Lots: 2274646, 2276336.
|
Recalling Firm/ Manufacturer |
Cook, Inc. 750 Daniels Way Bloomington, Indiana 47404-9120 |
| For Additional Information Contact |
Rita A. Harden 812-339-2235
|
Reason for Recall |
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
|
| Action |
Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form. |
| Quantity in Commerce |
51,473 total kits/trays |
| Distribution |
Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia. |
| |
|
|
|
|