| | Class 1 Recall Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, 5 Lumen |  |
| Date Posted |
February 23, 2011 |
| Recall Number |
Z-1133-2011 |
| Product |
Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Triple Lumen Polyurethane, 5.0 Fr /15 cm, C-UTLMY-501J- LSC-ABRM-HC-FST, G44016, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN. |
| Code Information |
Part Number: C-UTLMY-501J-LLSC-ABRM-HC-FST; Global Part Number G44016, lots 2239666, 2242701, 2248940, 2253262, 2254312, 2255610, 2262114, 2263390, 2275431, 2280744, 2285512, 2309427, 2321454 and 2332349.
|
Recalling Firm/ Manufacturer |
Cook, Inc. 750 Daniels Way Bloomington, Indiana 47404-9120 |
| For Additional Information Contact |
Rita A. Harden 812-339-2235
|
Reason for Recall |
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
|
| Action |
Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form. |
| Quantity in Commerce |
51,473 total kits/trays |
| Distribution |
Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia. |
| |
|
|
|
|