 Class 1 Recall
Central Venous Catheter Tray,
see related information
Date Posted February 23, 2011
Recall Status1 Terminated on June 21, 2011
Recall Number Z-1137-2011
Recall Event ID 57663
Premarket Notification
510(K) Number
K081113 
Product Classification Catheter,Intravascular,Therapeutic,Short-Term Less Than 30 Days - Product Code FOZ
Product Central Venous Catheter Tray, Double Lumen Polyurethane 5.0 Fr/ 12 cm, C-UDLMY-501J- RSC-CHILDRENS-HCT, G44041, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Code Information Part number: C-UDLMY-501J- RSC-CHILDRENS-HCT, Global part number: 44041 Lot number: F2337715 and F2355280.
Recalling Firm/
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington, Indiana 47404-9120
For Additional Information Contact Rita A. Harden
812-339-2235
Manufacturer Reason
for Recall
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
FDA Determined
Cause 2
COMPONENT CONTROLS (GMP - GOOD MANUFACTURING PRACTICE): Nonconforming Material/Component
Action Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity in Commerce 51,473 total kits/trays
Distribution Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = FOZ and Original Applicant = COOK, INC.
-
-