• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Central Venous Tray, Triple Lumen Polyurethane

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Central Venous Tray, Triple Lumen Polyurethane see related information
Date Posted February 23, 2011
Recall Status1 Terminated on June 21, 2011
Recall Number Z-1180-2011
Recall Event ID 57663
510(K)Number K081113 
Product Classification Peripherally Inserted, Central Venous Catheter Tray - Product Code FOZ
Product Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated,
Triple Lumen Polyurethane, 9.0 Fr/ 20cm, C-UTLMY-901J-RSC-ABRM-FSTA-RD, G49010, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Code Information Part number: C-UTLMY-901J-RSC-ABRM-FST-A-RD, Global part number G49010;

Lot numbers: 2238976.
Recalling Firm/
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington IN 47404-9120
For Additional Information Contact Rita A. Harden
812-339-2235
Manufacturer Reason
for Recall
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
FDA Determined
Cause 2
Nonconforming Material/Component
Action Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity in Commerce 51,473 total kits/trays
Distribution Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = FOZ and Original Applicant = COOK, INC.
-
-