Class 2 Device Recall Signa Excite

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Signa Excite see related information
Date Posted February 24, 2011
Recall Status1 Terminated on July 03, 2012
Recall Number Z-1429-2011
Recall Event ID 57084
510(K)Number K041476 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
Product 6) Signa Excite (K041476 Signa Excite 1.5T and 3.0T)
Signa Excite, 12x (K041476 Signa Excite 1.5T and 3.0T)

The GE Signa¿ HDe MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa HDe MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa¿ HDe system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.


Code Information SERIAL
00000277692MR0
00000276632MR7
00000280465MR6
00000284315MR9
00000276732MR5
00000280892MR1
00000280464MR9
00000281796MR3
00000278623MR4
00000278196MR1
00000275595MR7
00000281935MR7
00000278626MR7
00000282909MR1
00000280461MR5
00000283074MR3
00000274891MR1
00000279922MR9
00000282548MR7
00000281267MR5
00000273517MR3
00000281020MR8
00000280954MR9
00000282105MR6
00000283998MR3
00000285302MR6
00000281346MR7
00000279502MR9
00000285599MR7
00000273536MR3
00000281185MR9
00000281312MR9
00000281456MR4
00000283663MR3
00000273487MR9
00000280009MR2
00000282253MR4
00000280276MR7
00000279925MR2
00000279887MR4
00000278472MR6
00000281445MR7
00000276633MR5
00000283316MR8
00000276728MR3
00000282695MR6
2284300
00000273537MR1
00000283452MR1
00000283468MR7
00000273265MR9
00000283345MR7
00000284360MR5
00000287579MR7
00000278245MR6
00000279503MR7
00000279078MR0
00000280435MR9
00000282830MR9
00000283077MR6
00000283653MR4
00000283437MR2
00000281681MR7
00000278622MR6
00000280620MR6
00000274711MR1
00000281022MR4
00000274713MR7
00000281314MR5
00000281273MR3
00000280274MR2
00000283665MR8
00000279915MR3
00000283365MR5
00000277599MR7
00000281265MR9
00000027862MR0
00000283530MR4
00000273541MR3
00000284786MR1
00000275937MR1
00000274937MR2
00000287317MR2
00000283624MR5
00000282788MR9
00000273516MR5
00000277763MR9
00000278726MR5
00000281227MR9
00000277965MR0
00000280094MR4
00000281347MR5
00000276330MR8
00000283377MR0
00000279076MR4
00000274717MR8
00002181458MR0
00000276735MR8
00000274931MR5
00000274170MR0
00000282463MR9
00000286406MR4
00000282524MR8
00000277788MR6
00000281937MR3
00000027888MR3
00000287864MR3
00000277959MR3
00000283151MR9
00000280278MR3
00000279627MR4
00000283473MR7
00000274930MR7
00000278619MR2
00000278246MR4
00000277700MR1
00000274889MR5
00000274933MR1
00000282849MR9
00000283471MR1
00000273514MR0
00000276737MR4
00000278561MR6
00000274939MR8
00000287086MR3
00000282812MR7
00000282492MR8
00000281457MR2
00000284903MR2
00000281526MR4
00000285664MR9
00000281776MR5
00000281703MR9
00000281188MR3
00000280374MR0
00000286309MR0
00000280462MR3
00000281442MR4
00000275597MR3
00000281829MR2
00000274709MR5
00000275474MR5
00000287063MR2
00000282792MR1
00000287617MR5
00000280069MR6
00000286391MR8
00000282107MR2
00000282106MR4
00000281644MR5
00000285951MR0
00000275936MR3
00000281678MR3
00000275938MR9
00000289205MR7
00000282857MR2
00000283506MR4
00000280767MR5
00000286526MR9
00000280095MR1
00000282961MR2
00000281266MR7
00000286425MR4
00000273008MR3
00000281840MR9
00000283435MR6
00000280434MR2
00000279080MR6
00000280526MR5
00000282080MR1
00000286282MR9
00000283317MR6
00000279892MR4
00000286476MR7
00000280903MR6
00000283357MR2
00000283385MR3
00000285384MR4
00000287396MR6
00000274892MR9
00000283696MR3
00000285023MR8
00000286267MR0
00000283474MR5
00000282911MR7
00000284214MR4
00000275475MR2
00000286149MR0
00000275736MR7
00000283697MR1
00000277600MR3
00000289162MR0
00000286250MR6
00000280619MR8
00000281056MR2
00000280713MR9
00000280759MR2
00000281313MR7
00000275473MR7
00000281865MR6
00000281857MR3
00000278307MR4
00000282079MR3
00000283150MR1
00000285301MR8
00000283755MR7
00000282464MR7
00000282958MR8
00000282083MR5
00000273009MR1
00000282029MR8
00000274710MR3
00000274150MR2
00000275935MR5
00000276332MR4
00000282084MR3
00000283300MR2
00000285914MR8
00000285300MR0
00000286060MR9
00000286058MR3
00000286061MR7
00000278463MR5
00000277602MR9
00000284236MR7
00000285020MR4
00000276276MR3
00000276273MR0
00000288212MR4
00000273518MR1
00000286146MR6
00000279077MR2
00000274708MR7
00000278305MR8
00000273267MR5
00000272999MR4
00000274169MR2
00000275735MR9
00000275599MR9
00000275737MR5
00000274173MR4
00000279404MR8
00000273489MR5
00000277961MR9
00000284464MR5
00000287399MR0
00000278469MR2
00000279407MR1
00000283358MR0
0000028
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
262-513-4122
For Additional Information Contact
262-513-4122
Manufacturer Reason
for Recall
GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer does not perform the geometry check as required in the service procedure, images m
FDA Determined
Cause 2
Nonconforming Material/Component
Action On November 9, 2010, the firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated November 5, 2010 to all Consignees/Customers. The letter described the product, problem and action to be taken by the firm. GE instructed the customers to contact their local GEHC support Engineer, if they have any questions about the image orientation, or the procedures for geometry checks on thier system. Additionally, a GE Healthcare representative will install a device, which will reduce the potiential for the gradient cables to be swapped. Contact your local GEHC Support Engineer if you have any questions concerning this notification or phone: 262-548-2731.
Quantity in Commerce 1219
Distribution Worldwide distribution: USA including DC and PR; and countries including: VIET NAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, FORMER YUGOSLAV REPUBLIC OF MACEDONIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, ROMANIA, REUNION, REPUBLIC OF KOREA, PORTUGAL POLAND, PHILIPPINES, PERU,PARAGUAY, PANAMA, PAKISTAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MEXICO, MAURITIUS, MARTINIQUE,MALTA, MALAYSIA, LUXEMBOURG,LITHUANIA, LIBYAN ARAB JAMAHIRIYA, LEBANON, LATVIA, KUWAIT, KENYA, KAZAKHSTAN, JORDAN, JAPAN, JAMAICA, ITALY, ISRAEL, ISLAMIC REPUBLIC OF IRAN, IRELAND, INDONESIA, INDIA, ICELAND, HUNGARY, HONG KONG, HONDURAS, GUYANA, GUATEMALA, GUADELOUPE, GREECE, GERMANY, GEORGIA, FRENCH POLYNESIA, FRANCE, FINLAND, EL SALVADOR, EGYPT, ECUADOR, DOMINICAN REPUBLIC, DENMARK, CECH REPUBLIC, CYPRUS, CROATIA, COLOMBIA, CHINA, CHILE, CANDA, BULGARIA, BRAZIL, BOSNIA AND HERZEGOVINA, BOLIVIA, BELGIUM, BELARUS, AZERBAIJAN, AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS, LLC
-
-