| | Class 2 Recall Water Soluble Lubricant |  |
| Date Posted |
April 18, 2011 |
| Recall Number |
Z-2013-2011 |
| Product |
Lubricating Jelly packaged as follows:
(1) DeRoyal(R) Water Soluble Lubricant, Size: 5 Gram, Rx Only, REF 31-702, STERILE EO, Distributed by: DeRoyal - 200 DeBusk Lane, Powell, TN 37849, USA;
(2) DeRoyal(R) Water Soluble Lubricant, 1 / Pack, Size: 2 oz Rx Only, REF 31-704, STERILE EO, Distributor: DeRoyal, 200 DeBusk Lane, Powell, TN 37849, USA;
(3) DeRoyal(R) Lubricant, Size: 5 Gram, Rx Only, REF 31-938, STERILE EO, Distributed by: DeRoyal - 200 DeBusk Lane, Powell, TN 37849, USA
Lubricating body orifices. |
| Code Information |
31-702:
18040107, 18125741, 18150322, 18368014, 18605067, 18605067, 19032694, 19032694, 19136741, 19557491, 20023453,
20102637, 20190729, 20190729, 20023453, 20190729, 20510139, 20023453, 20190729, 21049840, 21049840, 21517201, 21780683, 21780683, 21517201, 21780683, 22025604, 22147169, 21517201, 21517201, 22269799, 22952133, 23474014, 23474014, 23474014, 24194532, 24194532, 24406620, 17950117, 19136741, 21517201, 21780683, 22147169, 22025604, 22595121, 23918771, 24406620, 20190729, 18040107, 18605067, 19032694, 19136741, 20190729, 18150322, 19032694, 20102637, 20190729, 20190729, 21517201, 22269799, 22269799, 24406620, 24406620, 17296856, 17950117, 18368014, 18368014, 18368014, 18605067, 19557491, 19557491, 20510139, 20634166, 21168988, 22952133, 22595121, 22898042, 22952133, 22269799, 24194532, 24194532, 24406620, 19557491, 19032694, 19032694, 20023453, 20023453, 20190729, 20510139, 20634166, 21517201, 21517201, 21517201, 22898042, 21517201, 23918771, 18040107, 18605067, 20023453, 20634166, 21049840, 21168988, 21517201, 22595121, 17950117, 17296856, 19136741, 19032694, 20190729, 20190729, 20190729, 20190729, 20190729, 20023453, 20634166, 21517201, 22025604, 21517201, 21517201, 23918771, 17950117, 18040107, 18368014, 18605067, 18605067, 19032694, 19218034, 19136741, 19557491, 20190729, 20190729, 20190729, 20510139, 21049840, 21049840, 21517201, 21780683, 22025604, 21517201, 21517201, 21517201, 21517201, 22952133, 22595121, 21517201, 23918771, 23474014, 24194532, 23474014, 24892314, 24194532, 22025604, 18040107, 18125741, 18368014, 18605067, 19032694, 19136741, 19218034, 20023453, 20102637, 20190729, 20190729, 20510139, 20634166, 21168988, 21517201, 21588464, 21517201, 20634166, 21517201, 21780683, 22025604, 22595121, 21517201, 22269799, 23474014, 24194532, 24194532, 20023453, 20190729, 20634166, 22595121, 22898042, 21517201, 23474014, 24641355
31-704:
17407430, 18395135, 18605075, 18605075, 19476391, 19476391, 19557520, 20510155, 20914037, 21049866, 21238801, 1397561, 21588481, 21746987, 22269810, 23474022, 23474022, 23750736, 23750736, 24406638, 24406638, 24406638, 17407430, 17407430, 17407430, 17950133, 18234024, 18395143, 18368031, 18368031, 18368031, 18605075, 18962021, 18962021, 18605075, 18962021, 18605075, 18962021, 18605075, 19089121, 19089121, 19476391, 19476391, 19476391, 19476391, 19792802, 19792802, 19557520, 19792802, 19557520, 19792802, 20510155, 20510155, 20510155, 20656576, 20656576, 20656576, 20914037, 20510155, 20656576, 20914037, 21049866, 20656576, 21049866, 21049866, 21238801, 21238801, 21397561, 21588481, 21588481, 21588481, 21746987, 21746987, 21746987, 22037111, 22037111, 22269810, 22531646, 22531646, 22531646, 22758218, 22758218, 22758218, 22269810, 23192682, 23192682, 23474022, 23474022, 23750736, 23474022, 23750736, 24175892, 24175892, 24175892, 24406638, 24406638, 24406638, 24764187, 17950133, 18234024, 18368031, 18368031, 18368031, 18605075, 18605075, 18962021, 18962021, 19089121, 19089121, 19476391, 19792802, 19557520, 19557520, 20510155, 20656576, 20656576, 20656576, 21049866, 21238801, 21238801, 21397561, 21397561, 21588481, 21746987, 21746987, 22037111, 22037111, 22269810, 22531646, 22758218, 22269810, 23474022, 23750736, 23750736, 23750736, 24175892, 23750736, 24175892, 24175892, 24175892, 24175892, 24406638, 24175892, 24175892, 24406638, 24406638, 24764187, 22269810, 18962021, 17407430, 18234024, 18234024, 18368031, 18395135, 18605075, 18962021, 19089121, 19089121, 19476391, 19476391, 19792802, 17407430, 18234024, 19089121, 18605075, 19476391, 20510155, 20656576, 21397561, 21588481, 21588481, 21746987, 22531646, 22758218, 22269810, 22269810, 23192682, 23474022, 23192682, 23474022, 24406638, 24764187, 18605075, 19089121, 19476391, 19557520, 19792802, 20510155, 20914037, 21049866, 21049866, 21049866, 21238801, 23750736, 24406638, 24406638, 24406638, 23192682, 18962021, 21049866, 21588481, 23750736, 17407430, 17950133, 18234024, 18234024, 20510155, 21049866, 21238801, 21746987, 22037111, 20914037, 18605075, 23474022, 22758218, 17407430, 17407430, 17407430, 17950133, 17950133, 18234024, 18368031, 18368031, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 18962021, 18605075, 18962021, 19476391, 19476391, 19476391, 19476391, 19792802, 19792802, 19557520, 19792802, 20510155, 20510155, 19557520, 20510155, 20656576, 20656576, 20914037, 20914037, 20656576, 20656576, 21049866, 21049866, 21506318, 21746987, 21746987, 22037111, 22037111, 22269810, 22758218, 22758218, 22758218, 22269810, 22269810, 23192682, 23192682, 23192682, 23474022, 23474022, 23474022, 23474022, 23474022, 23474022, 23750736, 23750736, 24175892, 24175892, 24406638, 24406638, 24406638, 24764187, 24406638, 21397561, 22269810, 23474022, 23750736, 24175892, 24764187, 19792802, 17407430, 17950133, 18234024, 18395143, 18368031, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 19476391, 19476391, 19792802, 19792802, 19557520, 19557520, 19792802, 20510155, 20510155, 20510155, 20510155, 20656576, 20656576, 20656576, 20914037, 20656576, 21238801, 21506318, 21397561, 21397561, 21588481, 21397561, 22037111, 22531646, 22531646, 22758218, 22269810, 22269810, 22269810, 23192682, 23192682, 23192682, 23192682, 23474022, 23474022, 23474022, 23750736, 23474022, 23750736, 23750736, 24175892, 24175892, 23750736, 24406638, 24406638, 24406638, 24406638, 24764187, 23192682, 17407430, 18605075, 18605075, 18605075, 19476391, 19476391, 20510155, 20914037, 21506318, 21588481, 22037111, 22758218, 22269810, 23474022, 23750736, 24175892, 24764187, 17950133, 18234024, 18368031, 18234024, 18605075, 19089121, 19089121, 18605075, 19476391, 1792802, 19557520, 20510155, 20510155, 20510155, 20914037, 21238801, 21238801, 21506318, 21397561, 21588481, 21746987, 21746987, 21746987, 22037111, 22269810, 22531646, 22758218, 22269810, 22758218, 22269810, 22531646, 23192682, 23192682, 23750736, 23750736, 24175892, 24175892, 24764187, 18605075, 17950133, 17950133, 18395143, 18368031, 18962021, 18605075, 19089121, 18605075, 19476391, 18605075, 19557520, 19557520, 20510155, 20510155, 20656576, 20656576, 20656576, 21238801, 21397561, 21588481, 21746987, 22269810, 22531646, 22758218, 22269810, 23192682, 23474022, 23750736, 24406638, 24406638, 24406638, 24764187, 19476391, 21049866, 22037111, 17950133, 18395143, 18368031, 18605075, 17950133, 18368031, 18605075, 19089121, 19476391, 19557520, 20510155, 20656576, 21238801, 21397561, 21588481, 21746987, 22037111, 22758218, 22758218, 23192682, 23750736, 24175892, 24406638, 18605075, 20510155, 20510155, 20656576, 21049866, 21049866, 21746987, 21746987, 22269810, 23192682, 23192682, 24406638, 17950133, 18368031, 18605075, 19792802, 20510155, 20914037, 21049866, 21397561, 21588481, 21588481, 21588481, 22269810, 23192682, 23192682, 23192682, 23474022, 23750736, 24175892, 24175892, 24406638, 24764187, 24764187, 18234024, 18234024, 18234024, 18234024, 18962021, 19089121, 19476391, 19476391, 19557520, 19557520, 20510155, 20510155, 20510155, 22758218, 23750736, 19476391, 19089121, 19476391, 19089121, 19476391, 20510155, 20510155, 20656576, 21238801, 18605075, 22531646, 22758218, 20656576, 22037111, 17407430, 17950133, 18395143, 18234024, 18368031, 18368031, 18605075, 18605075, 18605075, 18605075, 18605075, 18962021, 19089121, 18962021, 18605075, 18605075, 19476391, 19476391, 19089121, 19792802, 19557520, 19557520, 20510155, 20510155, 20510155, 20510155, 20914037, 20914037, 18234024, 20656576, 20914037, 21049866, 21049866, 20656576, 21049866, 21238801, 21238801, 21397561, 21397561, 21397561, 21746987, 21746987, 21746987, 22037111, 22269810, 22531646, 22531646, 22531646, 22758218, 22758218, 22269810, 22269810, 22269810, 23192682, 23192682, 23474022, 23474022, 23474022, 23750736, 23750736, 23750736, 23750736, 23750736, 24175892, 24175892, 24406638, 24406638, 24406638, 24406638, 24764187, 17407430, 23192682, 23474022, 24175892, 24175892, 21397561, 21588481, 21746987, 22037111, 22037111, 22269810, 22269810, 22269810, 22531646, 22269810, 23474022, 24406638, 24764187, 24406638, 24764187, 17407430, 18234024, 18368031, 18395135, 18395135, 18605075, 18605075, 18962021, 18962021, 18962021, 18962021, 19089121, 19089121, 19476391, 19476391, 19792802, 20510155, 20510155, 20656576, 20656576, 20656576, 21238801, 21397561, 21506318, 17950133, 17950133, 17950133, 18234024, 18368031, 18605075, 18605075, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 19476391, 19476391, 19792802, 19557520, 19557520, 20510155, 20510155, 20510155, 20656576, 20914037, 20914037, 24406638, 24764187, 18962021, 19476391, 19792802, 19792802, 20510155, 20656576, 21588481, 21746987, 22037111, 22531646, 23192682, 23474022, 23750736, 18234024, 19476391, 21238801, 22037111, 23192682, 24175892, 21397561, 21746987, 22531646, 22269810, 23474022, 24175892, 21049866
31-938:
18368057, 17070224, 17407448, 17407448, 18368057, 20656584, 21588499, 22147193, 20510180, 21747007, 24194559, 18500214, 20656584, 22269836, 23750744, 21747007, 24046927, 24194559, 24406646, 20069646, 21672797, 21672797, 18368057, 24406646, 17407448, 17407448, 18368057, 18500214, 18500214, 18500214, 19089148, 19089148, 19089148, 19089148, 19089148, 18500214, 20069646, 19940879, 20069646, 20510180, 20656584, 20656584, 20656584, 20069646, 20510180, 21672797, 21747007, 21747007, 21907509, 22147193, 22269836, 21907509, 22269836, 22853565, 21747007, 21907509, 23382419, 23750744, 24046927, 24194559, 24406646, 24194559, 17407448, 18500214, 18500214, 19089148, 20069646, 19940879, 20656584, 21672797, 22269836, 23750744, 19089148, 20656584, and 22269836.
|
Recalling Firm/ Manufacturer |
DeRoyal Processing Center 1703 Highway 33 S New Tazewell, Tennessee 37825-5043 |
| Consumer Instructions |
Contact the recalling firm for information |
| For Additional Information Contact |
Mr. Tracy Edmundson 865-362-2334
|
Reason for Recall |
Possible contamination due to sterilization issue. Inspection of contract manufacturer found cGMP deficiencies.
|
| Action |
The firm notified their consignees by a Re: DeRoyal Recall Of Triad Group Lubricating Jelly letter, dated Friday, March 18,2011. The letter discussed the reason for recall and the affected products. The letter asked customers to identify inventory using the attached spreadsheet and quarantine to prevent further use. Customers are to complete the spreadsheet regardless if they have effected product or not. The completed form can be faxed to 865-362-3716 or emailed to recalls@deroyal.com. All affected products should be return to DeRoyal at the address provided. Distributors are to forward a copy of the notice to their end-users. If there are questions or assistance is needed with the recall, Stephanie DeBusk should be contacted at 865-362-1037 or Sharon Cook at 865-362-6157. |
| Quantity in Commerce |
323,440 units |
| Distribution |
Nationwide Distribution |
| |
|
|
|
|