Class 2 Device Recall CARESCAPE Monitor B850 |
|
Date Initiated by Firm |
March 23, 2011 |
Date Posted |
April 08, 2011 |
Recall Status1 |
Terminated 3 on February 04, 2012 |
Recall Number |
Z-1931-2011 |
Recall Event ID |
58275 |
510(K)Number |
K092927
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
GE Healthcare, CARESCAPE Monitor B850
The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B650 is indicated for monitoring and recording of, and to generate alarms for, hemodynamic (including EGG, ST segment, arrhythmia detection, EGG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and mixed venous oxygen saturation), impedance respiration, airway gases (C02, 02, N20 and anesthetic agents), spirometry, gas exchange, and neurophysiological (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission) status. The CARESCAPE Monitor B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B650 is not intended for use during MRI. |
Code Information |
Parent SN SED10536360GA SED10536361GA SED10536362GA SED10536363GA SED10536364GA SED10536365GA SED10536368GA SED10536369GA SED10536370GA SED10536371GA SED10536372GA SED10536376GA SED10536377GA SED10536378GA SED10536379GA SED10536380GA SED10536381GA SED10536382GA SED10536383GA SED10536384GA SED10536385GA SED10536386GA SED10536387GA SED10536388GA SED10536389GA SED10536390GA SED10536391GA SED10536392GA SED10515982GA SED11056912GA SED11056923GA SED11056924GA SED11056925GA SED11056926GA SED11056927GA SED11056929GA SED11056930GA SED11056931GA SED11056932GA SED11056933GA SED11056959GA SED11056960GA SED11067132GA SED11067133GA SED11067134GA SED11067135GA SED11067136GA SED11067137GA SED11067139GA SED11067140GA SED11077142GA SED10526112GA SED10526244GA SED10526245GA SED10526253GA SED10526255GA SED10526256GA SED10526263GA SED10526265GA SED10526266GA SED10526293GA SED10526294GA SED10526295GA SED10526296GA SED10526297GA SED10526298GA SED10526301GA SED10526303GA SED10526305GA SED10526307GA SED10526308GA SED10526310GA SED10516098GA SED10526133GA SED10526134GA SED10526142GA SED10526143GA SED11046852GA SEC09440747GA SEC09340300GA SEC09340274GA SEC09340294GA SEC09430557GA SEC09340313GA SED10526115GA SED10536325GA SED10536328GA SED10536330GA SED10536331GA SED10536349GA SED10536350GA SED10536351GA SED10536352GA SED11067056GA SED11067072GA SED11067083GA SEC09430555GA SED10495275GA SED10495309GA SEC09430578GA SEC09340281GA SEC09340279GA SEC09340314GA SED09390860GA.DE; SED10454809GA SED10454812GA SED10454813GA SED10454814GA SED10454815GA SED10515833GA SED10515837GA SED10515838GA SED10515840GA SED10515841GA SED10515842GA SED10515843GA SED10515844GA SED10515845GA SED10515860GA SED10515872GA SED10515878GA SED10515879GA SED10515880GA SED10516089GA SED10526136GA SED10526141GA SED10526146GA SED10536397GA SED10536398GA SED10536413GA SED10536418GA SED10536420GA SED10536424GA SED10536428GA SED10536430GA SED10536432GA SED11057066GA SED11057069GA SED11057070GA SED11097478GA DTK08130668 SEC09340318GA SEC09340306GA SEC09340302GA SEC09430583GA SEC09340336GA SEC09340304GA SEC09340316GA SEC09340277GA SEC09430570GA SED10515834GA SED10515853GA SED10515869GA SED10515886GA SED10515894GA SED10515919GA SEC09430573GA SED10515826GA SED10515998GA SED11036635GA SED11036641GA SED11036669GA SED11036680GA SED11036705GA SED11046709GA SED11046711GA SED11046712GA SED11046713GA SED11046717GA SED11046718GA SED11046719GA SED11046726GA SED11046728GA SED11046733GA SED10404245GA SED10404246GA SED10404249GA SED10404250GA SED10404251GA SED10404252GA SED10404253GA SED10505482GA SED10505500GA SED10505503GA SED10505530GA SED10505531GA SED10505538GA SED10505539GA SED10505551GA SED10505554GA SED10505561GA SED10505569GA SED10505601GA SED11046737GA SEC09340335GA SEC09340312GA SEC09340320GA SED11067074GA SED11097452GA SEC09340310GA SED10464941GA SEC09340341GA SEC09340338GA SEC09340308GA SEC09340337GA SEC09420529GA SEC09430562GA SEC09430577GA SEC09430575GA SEC09420524GA SEC09430619GA SEC09420534GA SED11056963GA SED11056970GA SED11056981GA SED11056982GA SED11056983GA SEC09430567GA SEC09420532GA SED10495168GA SED10495169GA SED10495170GA SED10404247GA SED10404248GA SED10464842GA SED10464844GA SED10464845GA SED10464846GA SED10464847GA SED10464848GA SED10464849GA SED10505450GA SED10505507GA SED10505508GA SED10505509GA SED10505511GA SED10505513GA SED10505515GA SED10505516GA SED10505519GA SED10505520GA SED10505521GA SED10505523GA SED10505524GA SED10505528GA SED10505529GA SED10505532GA SED10505533GA SED10505535GA SED10505536GA SED10505537GA SED10505540GA SED10505542GA SED10505543GA SED10505545GA SED10505546GA SED10505548GA SED10505550GA SED10505552GA SED10505553GA SED10505555GA SED10505557GA SED10505559GA SED10505560GA SED10505562GA SED10505563GA SED10505564GA SED10505565GA SED10505566GA SED10505567GA SED10505568GA SED10505571GA SED10505573GA SED10505574GA SED10505575GA SED10505576GA SED10505577GA SED10505579GA SED10505580GA SED10505581GA SED10505582GA SED10505583GA SED10505585GA SED10505586GA SED10505587GA SED10505588GA SED10505589GA SED10505590GA SED10505592GA SED10505593GA SED10505594GA SED10505595GA SED10505596GA SED10505602GA SED10505604GA SED10505605GA SED10505606GA SED10505607GA SED10505608GA SED10505609GA SED10505610GA SED10505613GA SED10505614GA SED10505616GA SED10505619GA SED10505620GA SED10505621GA SED10505622GA SED10505623GA SED10505627GA SED10505628GA SED10505632GA SED10505633GA SED10505635GA SED10505636GA SED10505637GA SED10505638GA SED10505640GA SED10505641GA SED10505642GA SED10505643GA SED10505644GA SED10505645GA SED10505646GA SED10505647GA SED10505648GA SED10505649GA SED10505650GA SED10505651GA SED10505652GA SED10505654GA SED10505656GA SED10505657GA SED10505658GA SED10505659GA SED10505662GA SED10505663GA SED10505664GA SED10505665GA SED10505666GA SED10505667GA SED10505668GA SED10505669GA SED10505670GA SED10505671GA SED10505672GA SED10505673GA SED10505677GA SED10505679GA SED10505680GA SED10505681GA SED10505682GA SED10505684GA SED10505685GA SED10505686GA SED10505687GA SED10505688GA SED10505689GA SED10505690GA SED10505691GA SED10505692GA SED10505693GA SED10505697GA SED10505698GA SED10505699GA SED10505700GA SED10505701GA SED10505702GA SED10505703GA SED10505704GA SED10505705GA SED10505706GA SED10505708GA SED10505709GA SED10505713GA SED10505718GA SED10505720GA SED10505723GA SED10505724GA SED10505725GA SED10515727GA SED10515728GA SED10515729GA SED10515730GA SED10515732GA SED10515733GA SED10515734GA SED10515736GA SED10515738GA SED10515739GA SED10515740GA SED10515741GA SED10515742GA SED10515743GA SED10515744GA SED10515746GA SED10515747GA SED10515748GA SED10515749GA SED10515752GA SED10515813GA SED10515817GA SED10515821GA SED10515824GA SED10515836GA SED10515866GA SED10515867GA SED10515868GA SED10515870GA SED10515885GA SED10515887GA SED10515888GA SED10515898GA SED10515905GA SED10515907GA SED10515912GA SED10515915GA SED10515917GA SED10515922GA SED10515924GA SED10515928GA SED10515929GA SED10515930GA SED10515931GA SED10515932GA SED10515933GA SED10515934GA SED10515935GA SED10515936GA SED10515937GA SED10515938GA SED10515939GA SED10515940GA SED10515941GA SED10515942GA SED10515943GA SED10515944GA SED10515945GA SED10515946GA SED10515947GA SED10515948GA SED10515949GA SED10515950GA SED10515951GA SED10515952GA SED10515953GA SED10515954GA SED10515955GA SED10515956GA SED10515957GA SED10515958GA SED10515959GA SED10515960GA SED10515961GA SED10515962GA SED10515963GA SED10515964GA SED10515965GA SED10515975GA SED10515980GA SED10515984GA SED10515989GA SED10515990GA SED10515991GA SED10515992GA SED10515993GA SED10515994GA SED10515996GA SED10515997GA SED10516002GA SED10516007GA SED10516009GA SED11067089GA SED11067093GA SED11067099GA SED11067103GA SED11067117GA SED11067123GA SEC09430560GA SEC09430568GA SEC09430576GA SED11067130GA SED11077164GA SED11087352GA SED11097417GA SED11097426GA SED11097442GA SED11097446GA SED10475024GA SED10475025GA SED10475026GA 6621436 6621466 SA309538029GA SA309538034GA SA309538043GA SA309538046GA SA309538050GA SA309538093GA SA310028189GA SA310038380GA SA310038441GA SA310048620GA SEG10042545HX SED11046738GA SED11046752GA SED11046754GA SED11046755GA SED11046756GA SED11046757GA SED11046758GA SED11046759GA SED11046763GA SED11046765GA SED11046781GA SED11046782GA SED11046784GA SEC09420528GA SEC09430572GA SEC09430554GA SED11036689GA SED11036692GA SED11036701GA SED11036702GA SED11036703GA SED11036704GA SED11046710GA SED11046716GA SED11046720GA SED11046742GA SED11046745GA SED11046747GA SED11046749GA SED11046768GA SED11046771GA SED11046774GA SED11046775GA SED11046777GA SED11046780GA SED11046783GA SED10485085GA SEC09420530GA SEC09430563GA SED10536329GA SED10536346GA SED10536348GA SEC09420535GA SED11046818GA SEC09430571GA SEC09430580GA SEC09430582GA SEC09420537GA SEC09430564GA SED10505674GA SED10505675GA SED10505678GA SED10505696GA SED10515754GA SED10515767GA SED10515769GA SED10515770GA SED10515771GA SED11036614GA SED11057061GA SED11067129GA SED11077147GA SED11077156GA SED11077180GA SED11077188GA SED11087313GA SED11087384GA SED11087388GA SED11087406GA SEC09430574GA SEC09430579GA SEC09420533GA SEC09420512GA SED10454786GA SED10454787GA SED10454788GA SED10454789GA SED10454790GA SED10454791GA SEC09430561GA SEC09420525GA SEC09420536GA SED10516046GA SED10516068GA SED10516083GA SED10526158GA SED10526159GA SED10526160GA SED10526165GA SED10526167GA SED10526169GA SED10526170GA SED10526171GA SED10526172GA SED10526173GA SED10526174GA SED10526181GA SED10526184GA SED10526185GA SED10526186GA SED10526187GA SED10526191GA SED10526192GA SED10526193GA SED10526197GA SED10526199GA SED10526201GA SED10526202GA SED10526203GA SED10526204GA SED10526208GA SED10526209GA SED10526211GA SED10526221GA SED10526224GA SED10526227GA SED10526228GA SEC09420495GA SEC09430581GA SEC09430569GA SEC09420527GA SEC09430565GA SED10526198GA SED10526200GA SED10526207GA SED10526210GA SED11056962GA SED11056964GA SED11056965GA SED11056966GA SED11056967GA SED11056968GA SED11056969GA SED11056977GA SED11056978GA SED11056979GA SED11056980GA SED11056984GA SED11056986GA SED11056987GA SED11066997GA SED11066999GA SED11067000GA SED11067002GA SED11067003GA SED11067004GA SED11067005GA SED11067006GA SED11067007GA SED11067008GA SED11067009GA SED11067010GA SED11067011GA SED11067012GA SED11067013GA SED11067014GA SED11067015GA SED11067016GA SED11067017GA SED11067018GA SED11067019GA SED11067020GA SED11067021GA SED11067022GA SED11067023GA SED11067024GA SED11067025GA SED11067026GA SED11067027GA SED11067028GA SED11067029GA SED11067030GA SED11067031GA SED11067032GA SED11067033GA SED11067034GA SED11067035GA SED11067039GA SED11067040GA SED11067042GA SED11067043GA SED11067097GA SED11067100GA SED11067110GA SED11056941GA SED11056951GA SED11087303GA SEC09340242GA SEC09340245GA SEC09340246GA SEC09340247GA SEC09340248GA SEC09340249GA SEC09340251GA SEC09340254GA SEC09340257GA SEC09340259GA SEC09340260GA SEC09340261GA SEC09340263GA SEC09340265GA SEC09340266GA SEC09340269GA SEC09340271GA SEC09340285GA SEC09340303GA SEC09340305GA SEC09340307GA SEC09340315GA SEC09340319GA SEC09340333GA SEC09340334GA SEC09420502GA SEC09420538GA SEC09430566GA SED10515995GA SED10516005GA SED10516025GA SEC09420526GA SED10536395GA SED10536400GA SED10536419GA SED10536422GA SED10363830GA SED10363833GA SED10363836GA SED10363837GA SED10363871GA SED10363876GA SED10363878GA SED10363879GA SED10363881GA SED10363883GA SED10363884GA SED10363886GA SED10363888GA SED10363890GA SED10363891GA SED10363893GA SED10363894GA SED10363895GA SED10363898GA SED10363903GA SED10363904GA SED10363910GA SED10363911GA SED10373928GA SED10373930GA SED10373932GA SED09310475GA SED10152647GA SED10526116GA SED10526117GA SED10526118GA SED10526119GA SED10526120GA SED10536373GA SED10536374GA SED10536375GA SED10252890GA SED10252893GA SED10252896GA SED10252898GA SED10252909GA SED10252916GA SED10252923GA SED10252928GA SED10262946GA SED10262947GA SED10313515GA SED10313516GA SED10313517GA SED10313518GA SED10313519GA SED10323520GA SED10293145GA SED09330594GA SED10313500GA SED10363868GA SED10373915GA SED10373917GA SED10373918GA SED10373920GA SED10374004GA SED10384026GA SED10384035GA SED10384039GA SED10384040GA SED10384043GA SED10384044GA SED10384045GA SED10384068GA SED10384075GA SED10384078GA SED10485143GA SED10273049GA
SED10283073GA SED10262951GA SED10262953GA SED10404234GA SED10404236GA SED10404237GA SED10333617GA SED10333621GA SED10363823GA SED10363824GA SED10363825GA SED10363826GA SED10363827GA SED10363828GA SED10363829GA SED10363831GA SED10363832GA SED10363834GA SED10363835GA SED10363870GA SED10363874GA SED10363877GA SED10363880GA SED10363882GA SED10373927GA SED10373929GA SED10373933GA SED10384027GA SED10384028GA SED10384030GA SED10384031GA SED10384038GA SED10384041GA SED10384076GA SED10384157GA SED10384158GA SED10384159GA SED10384160GA SED10384161GA SED10384162GA SED10384163GA SED10384164GA SED10384165GA SED10384166GA SED10384167GA SED10384168GA SED10384169GA SED10384170GA SED10394171GA SED10394172GA SED09310521GA SED09441082GA SED10152630GA SED10152631GA SED10142521GA SED10142523GA SED10142527GA SED10142529GA SED10142557GA SED10142559GA SED10142561GA SED10142563GA SED10142571GA SED10142576GA SED10142578GA SED10232876GA SED10232877GA SED10232878GA SED10232879GA SED10232880GA SED10232881GA SED10232882GA SED10232883GA SED10232884GA SED10232886GA SED10232887GA SED10363901GA SED10021717GA SED10021718GA SED10021722GA SED10031770GA SED10031772GA SED10031776GA SED10505599GA SED09451341GA SED09471459GA SED09471460GA SED09471461GA SED09471463GA SED09471464GA SED09471467GA SED09471468GA SED09471469GA SED09471470GA SED09471472GA SED09471474GA SED09471476GA SED09471480GA SED09471481GA SED09471486GA SED09471488GA SED09471491GA SED09471493GA SED09471496GA SED10495319GA SED10495327GA SED10495328GA SED10495333GA SED10495334GA SED10505352GA SED10505368GA SED10505398GA SED10505399GA SED10505426GA SED10505433GA SED10505437GA SED10505438GA SED10505439GA SED10505440GA SED10505441GA SED10505442GA SED10505461GA SED10505462GA SED10505469GA SED10505470GA SED10505474GA SED10505478GA SED10505485GA SED10515829GA SED10515876GA SED09310468GA SED10112151GA SED10112161GA SED10112175GA SED10112182GA SED10112195GA SED10112196GA SED10112197GA SED10132449GA SED10132450GA SED10132452GA SED10132453GA SED10132456GA SED10132458GA SED10132462GA SED10132464GA SED09390835GA SED09330591GA SED09310488GA SED09320555GA SED09390847GA SED10021708GA SED09310467GA SED09420994GA SED09420980GA SED09310484GA SED09370820GA SED09330570GA SED09390883GA SED09451339GA SED09451340GA SED10061886GA SED10102053GA SED10323578GA SED10526217GA SED11056945GA SED10505434GA SED10021730GA SED10031735GA SED10031744GA SED10031749GA SED10071894GA SED10071896GA SED10071901GA SED10071902GA SED10071903GA SED10071904GA SED10071907GA SED10071913GA SED10071914GA SED10071916GA SED10071920GA SED10071921GA SED10071931GA SED10071937GA SED10071939GA SED10081947GA SED10081955GA SED10081956GA SED10081959GA SED10081961GA SED10081962GA SED10081964GA SED10081965GA SED10081966GA SED10081969GA SED10081973GA SED10081976GA SED10081981GA SED10081984GA SED10081985GA SED10081986GA SED10081988GA SED10081989GA SED10081990GA SED10081994GA SED10081995GA SED10081997GA SED10082004GA SED10082007GA SED10082008GA SED10082009GA SED10082011GA SED10082012GA SED10082015GA SED10082016GA SED10082018GA SED10082019GA SED10082020GA SED10082021GA SED10082022GA SED10092030GA SED10092032GA SED10102082GA SED10102083GA SED10102107GA SED10102112GA SED10333625GA SED10333626GA SED10333628GA SED10434559GA SED10434560GA SED10434561GA SED10434562GA SED10434563GA SED10434564GA SED10434565GA SED10434641GA SED10434642GA SED10434643GA SED10434644GA SED10434645GA SED10434646GA SED10434647GA SED10444648GA SED10444649GA SED10444650GA SED10444651GA SED10444652GA SED10444653GA SED10444655GA SED10444656GA SED10444657GA SED10444658GA SED10444659GA SED10444660GA SED10444661GA SED10444662GA SED10444663GA SED10444664GA SED10444665GA SED10444666GA SED10444667GA SED10444668GA SED10444669GA SED10444670GA SED10444671GA SED10444672GA SED10444673GA SED10444674GA SED10444694GA SED10444695GA SED10444696GA SED10444697GA SED10444698GA SED10444699GA SED10444700GA SED10444701GA SED10444702GA SED10444703GA SED10444704GA SED10444705GA SED10444706GA SED10444707GA SED10444708GA SED10444709GA SED10444710GA SED10444711GA SED10444712GA SED10444713GA SED10444714GA SED10526179GA SED10526190GA SED10526213GA SED10526214GA SED10526216GA SED10526218GA SED10526219GA SED10526220GA SED10526222GA SED10526223GA SED10526225GA SED10526226GA SED10526229GA SED10526231GA SED10536401GA SED10536402GA SED10536403GA SED10536408GA SED10536414GA SED10536416GA SED10536417GA SED10536423GA SED10536426GA SED09320528GA SED10263012GA SED10273035GA SED10273047GA SED10273048GA SED10273061GA SED10273062GA SED10273063GA SED10283064GA SED10283065GA SED10283066GA SED10283067GA SED10283068GA SED10283069GA SED10283070GA SED10283071GA SED10283072GA SED10303401GA SED10303402GA SED10303421GA SED09421001GA SED10263009GA SED10263013GA SED10526299GA SED10526304GA SED10526309GA SED10526311GA SED10526312GA SED10536321GA SED10536335GA SED10536336GA SED10536337GA SED10536341GA SED10262962GA SED09320552GA SED09451369GA SED09390857GA SED10212790GA SED10212791GA SED10212792GA SED10212793GA SED10212794GA SED10212795GA SED09320558GA SED09330572GA SED09370832GA SED09310523GA SED09370811GA SED09310487GA SED10252913GA SED10252917GA SED10293118GA SED10333649GA SED10333650GA SED10333651GA SED10333652GA SED10333653GA SED10333654GA SED10333655GA SED10333656GA SED10343726GA SED10343727GA SED10343728GA SED10343729GA SED10363840GA SED10363861GA SED10363862GA SED10373924GA SED10384135GA SED10384136GA SED10384137GA SED10384138GA SED10384139GA SED10384140GA SED10404268GA SED10404269GA SED10404270GA SED10404271GA SED10424492GA SED10424493GA SED10424494GA SED10424495GA SED10424496GA SED10424497GA SED10434508GA SED10434509GA SED10434510GA SED10434536GA SED10454780GA SED10454781GA SED10454782GA SED10454783GA SED10454784GA SED10454785GA SED10464833GA SED10464835GA SED10464836GA SED10464837GA SED10464838GA SED10464839GA SED10464840GA SED10464841GA SED10464843GA SED10464950GA SED10464951GA SED10475033GA SED10475034GA SED10475035GA SED10475036GA SED10485082GA SED10485107GA SED10485109GA SED10485110GA SED10505506GA SED10505556GA SED10515979GA SED10515981GA SED10515983GA SED10515986GA SED10515988GA SED10516019GA SED10516020GA SED10516101GA SED10516102GA SED10516103GA SED10516104GA SED10526109GA SED10526110GA SED10526114GA SED10526162GA SED10526166GA SED10536463GA SED10536467GA SED10536522GA SED11036596GA SED11036633GA SED11036646GA SED11036655GA SED11046817GA SED11046819GA SED11046820GA SED11046841GA SED11046845GA SED11046850GA SED11046857GA SED11046858GA SED11046880GA SED11046881GA SED11046882GA SED11046883GA SED11056915GA SED11056916GA SED11056917GA SED11056918GA SED11056934GA SED11056944GA SED11056953GA SED11056954GA SED11056957GA SED11077230GA SED11077256GA SED09320527GA SED10182700GA SED10252924GA SED10262941GA SED10262942GA SED10262961GA SED10262968GA SED10262991GA SED10262993GA SED10262995GA SED10273051GA SED10283077GA SED10293120GA SED10303330GA SED10303331GA SED10303332GA SED10303333GA SED10303346GA SED10303347GA SED10303348GA SED10303349GA SED10303350GA SED10303351GA SED10303353GA SED10303366GA SED09370819GA SED09370826GA SED09390834GA SED09320532GA SED09390856GA SED09431072GA SED09330574GA SED09310471GA SED09390877GA SED09390860GA SED09310520GA SED09330596GA SED10222849GA SED10252899GA SED10252900GA SED10252911GA SED10262939GA SED09400896GA SED09320560GA SED09410954GA SED10142579GA SED10142583GA SED10142585GA SED10142586GA SED10142598GA SED10536458GA SED10536509GA SED10536525GA SED10536526GA SED10536531GA SED10536532GA SED11026533GA SED11026535GA SED11026537GA SED11026539GA SED11026545GA SED11026552GA SED11026553GA SED11026554GA SED11026555GA SED11026558GA SED10262940GA SED10262943GA SED10222857GA SED09090249GP SED08060043GP SED08060044GP SED07510001GP SED07510002GP SED07510003GP SED07510004GP SED07510005GP SED07510006GP SED07510007GP SED07510008GP SED07510009GP SED07510010GP SED07510011GP SED07510012GP SED07510013GP SED07510014GP SED07510015GP SED07510016GP SED07510017GP SED07510018GP SED07510019GP SED07510020GP SED07510021GP SED07510022GP SED07510023GP SED07510024GP SED07510025GP SED07510026GP SED07510027GP SED07510028GP SED07510029GP SED07510030GP SED07510031GP SED07510032GP SED07510033GP SED07510034GP SED07510035GP SED07510036GP SED07510037GP SED07510038GP SED07510039GP SED07510040GP SED07510041GP SED07510042GP SED08060045GP SED08060046GP SED08060047GP SED08060048GP SED08060049GP SED08060050GP SED08060051GP SED08060052GP SED08060053GP SED08060054GP SED08060055GP SED08060056GP SED08060057GP SED08060058GP SED08060059GP SED08060060GP SED08060061GP SED08060062GP SED08060063GP SED08060064GP SED08060065GP SED08060066GP SED08060067GP SED08060068GP SED08060069GP SED08060070GP SED08060071GP SED08060072GP SED08070073GP SED08070074GP SED08070075GP SED08070076GP SED08070077GP SED08080078GP SED08080079GP SED08080080GP SED08080081GP SED08080082GP SED08080083GP SED08080084GP SED08080085GP SED08080086GP SED08080087GP SED08080088GP SED08080089GP SED08080090GP SED08080091GP SED08080092GP SED08080093GP SED08080094GP SED08080095GP SED08080096GP SED08080097GP SED08080098GP SED08080099GP SED08080100GP SED08080101GP SED08080102GP SED08080103GP SED08080104GP SED08080105GP SED08080106GP SED08080107GP SED08080108GP SED08080109GP SED08080110GP SED08080111GP SED08080112GP SED08140113GP SED08140114GP SED08140115GP SED08140116GP SED08140117GP SED08140118GP SED08140119GP SED08140120GP SED08150123GP SED08150124GP SED08200125GP SED08200126GP SED08200127GP SED08200128GP SED08200129GP SED08200130GP SED08200131GP SED08210132GP SED08210133GP SED08210134GP SED08210135GP SED08210136GP SED08210137GP SED08210138GP SED08250153GP SED08250154GP SED08250155GP SED08250156GP SED08250157GP SED08250158GP SED08250159GP SED08250160GP SED08250161GP SED08250162GP SED08250163GP SED08250164GP SED08250165GP SED08250166GP SED08290167GP SED08290168GP SED08290169GP SED08290170GP SED08290171GP SED08290172GP SED08290173GP SED08320174GP SED08320175GP SED08370176GP SED08370177GP SED08370178GP SED08370179GP SED08370180GP SED08370181GP SED08370182GP SED08370184GP SED08370185GP SED08370186GP SED08400187GP SED08400188GP SED08400189GP SED08400190GP SED08400191GP SED08400192GP SED08400193GP SED08400194GP SED08400195GP SED08400196GP SED08400197GP SED08400198GP SED08400199GP SED08400201GP SED08400202GP SED08400203GP SED08400204GP SED08490205GP SED08490206GP SED08490207GP SED08490209GP SED08490210GP SED08490211GP SED08490212GP SED08490213GP SED08490214GP SED08490215GP SED08510216GP SED08510217GP SED08510218GP SED08510219GP SED08510220GP SED08510221GP SED08510222GP SED08510223GP SED08510224GP SED08510225GP SED08510226GP SED08510227GP SED08510228GP SED08510229GP SED08510230GP SED08510231GP SED09030233GP SED09030234GP SED09030235GP SED09030236GP SED09030237GP SED09030238GP SED09030239GP SED09030240GP SED09050242GP SED09050243GP SED09050244GP SED09090250GP SED09090251GP SED09090252GP SED09130285GP SED09130300GP SED09330580GA SED09330590GA SED09340610GA SED09340626GP SED09390859GA SED09421003GA SED09421008GA SED11046724GA SED10454732GA SED10454733GA SED10454734GA SED10495311GA SED10495316GA SED10495318GA SED10495321GA SED10495323GA SED10495324GA SED10495331GA SED10495332GA SED10505378GA SED10505382GA SED10505383GA SED10505387GA SED10505415GA SED10505418GA SED10505420GA SED10505422GA
SED10505424GA SED10505430GA SED10505451GA SED10505453GA SED10505455GA SED10505456GA SED10505457GA SED10505458GA SED10505459GA SED10505473GA SED10505475GA SED10505476GA SED10505491GA SED10505492GA SED10505493GA SED10505496GA SED10505502GA SED10505504GA SED10102105GA SED10102116GA SED10263001GA SED10263003GA SED10263004GA SED10263005GA SED10263007GA SED10263008GA SED10263010GA SED10263011GA SED10263014GA SED10263015GA SED10263016GA SED10273030GA SED10273031GA SED10273032GA SED10273037GA SED10273038GA SED10273039GA SED10273040GA SED10273043GA SED10273044GA SED10273045GA SED10273046GA SED10273050GA SED09370822GA SED10071918GA SED10082013GA SED10092038GA SED11046851GA SED11046853GA SED11046859GA SED11046865GA SED11046870GA SED11046871GA SED10102065GA SED10102075GA SED10102093GA SED10102096GA SED10102097GA SED10102098GA SED10102099GA SED10102120GA SED10102121GA SED10102122GA SED10102126GA SED10102127GA SED10102128GA SED10102129GA SED10102130GA SED10102133GA SED10102135GA SED10102136GA SED10102139GA SED10112142GA SED10112143GA SED10112146GA SED10112147GA SED10112152GA SED10112154GA SED10112157GA SED10112158GA SED10112160GA SED10112162GA SED10112164GA SED10112166GA SED10132465GA SED10132466GA SED10142519GA SED10142520GA SED10142522GA SED10152617GA SED10152627GA SED10152645GA SED10152654GA SED10152662GA SED10162673GA SED09330578GA SED09370828GA SED09390874GA SED09320554GA SED10041836GA SED10434511GA SED09130286GP SED09130290GP SED09130291GP SED09130298GP SED09130302GP SED09180333GP SED09180338GP SED09130305GP SED09130312GP SED09180314GP SED09180334GP SED09180341GP SED09220344GP SED09220345GP SED09220346GP SED09220347GP SED09220348GP SED09220350GP SED09220351GP SED09220355GP SED09220357GP SED09220358GP SED09220362GP SED09220363GP SED09220365GP SED09220367GP SED09220369GP SED09220370GP SED09220371GP SED09220372GP SED09220373GP SED09220374GP SED09220375GP SED09220377GP SED09220378GP SED09220380GP SED09220381GP SED09220382GP SED09220383GP SED09230384GP SED09230385GP SED09230386GP SED09230388GP SED09230389GP SED09230390GP SED09240395GP SED09240399GP SED09240401GP SED09280405GA SED09280415GA SED09280421GA SED09280443GA SED09280402GA SED09280403GA SED09280406GA SED09280416GA SED09280424GA SED09280436GA SED09280437GA SED09310450GA SED09320556GA SED09441075GA SED09471478GA SED09491583GA SED09491587GA SED09491595GA SED09491596GA SED09491597GA SED09491599GA SED09511674GA SED09511676GA SED10021713GA SED09130288GR SED09130296GP SED09130307GP SED09370802GA SED10212796GA SED10212797GA SED10212798GA SED09280417GA SED10384072GA SED10373978GA SED10373984GA SED10373986GA SED10373991GA SED10373992GA SED10373993GA SED10373998GA SED10374000GA SED10374001GA SED10374002GA SED10374005GA SED10384006GA SED10384008GA SED10384009GA SED10384010GA SED10384011GA SED10384012GA SED10384014GA SED10384015GA SED10384016GA SED10384017GA SED10384018GA SED10384019GA SED10384020GA SED10384021GA SED10384022GA SED10384023GA SED10384033GA SED10384046GA SED10384069GA SED10384073GA SED10384077GA SED10536396GA SED09280404GA SED09310457GA SED09370833GA SED10424505GA SED10434512GA SED10434513GA SED10434514GA SED09330598GA SED11046790GA SED10132485GA SED10162685GA SED10222852GA SED10222853GA SED10222854GA SED10222855GA SED10222856GA SED10222858GA SED10222859GA SED10222860GA SED10222861GA SED10222862GA SED10222863GA SED10222864GA SED10222865GA SED10222866GA SED10222867GA SED10222869GA SED10222870GA SED10222871GA SED10222872GA SED10222873GA SED10222874GA SED10222875GA SED10252901GA SED10252902GA SED10252903GA SED10252904GA SED10252908GA SED10252914GA SED10252915GA SED10252927GA SED10262929GA SED10262930GA SED10262931GA SED10262932GA SED09471432GA SED10021683GA SED10021684GA SED10021707GA SED09531679GA SED10092034GA SED10112168GA SED10112169GA SED10112171GA SED10112172GA SED10112174GA SED10112176GA SED10112177GA SED10112206GA SED10112207GA SED10293177GA SED09370809GA SED10424498GA SED10424499GA SED10424500GA SED10424501GA SED10424502GA SED09390879GA SED10384067GA SED10262933GA SED10262934GA SED10262935GA SED10262936GA SED10262937GA SED10262938GA SED10262955GA SED10262956GA SED10262957GA SED10262967GA SED10262970GA SED10262971GA SED09310489GA SED09441080GA SED09370825GA SED09390841GA SED09330595GA SED09370827GA SED09310490GA SED09310500GA SED09310495GA SED09310512GA SED09310456GA SED09310511GA SED09310501GA SED09310493GA SED09320545GA SED09420990GA SED09310452GA SED09320547GA SED09320540GA SED09320559GA SED10102115GA SED09310508GA SED09280410GA SED09310507GA SED09320529GA SED09310454GA SED09320550GA SED10536438GA SED10536493GA SED10536508GA SED09280419GA SED09310476GA SED09310451GA SED09420993GA SED10132461GA SED10132469GA SED09330568GA SED09330597GA SED09320564GA SED10384032GA SED10384070GA SED10515985GA SED09441077GA SED09310524GA SED09310449GA SED09320526GA SED09330566GA SED09310455GA SED09320546GA SED10293147GA SED10293148GA SED10293152GA SED10293163GA SED10293175GA SED09471484GA SED10122345GA SED10394177GA SED10404238GA SED10132476GA SED10132481GA SED10132482GA SED10132489GA SED10132494GA SED10132497GA SED10132498GA SED10132510GA SED09451348GA SED09471462GA SED10414286GA SED10414287GA SED10414288GA SED10424397GA SED10424398GA SED10424399GA SED10424400GA SED10424401GA SED10424402GA SED10424403GA SED10424404GA SED10424405GA SED10424406GA SED10424407GA SED10424408GA SED10424409GA SED10424410GA SED10424412GA SED10424413GA SED10424414GA SED10424415GA SED10424416GA SED10424417GA SED10424418GA SED10424419GA SED10424420GA SED10424421GA SED10424422GA SED10424423GA SED10424424GA SED10424425GA SED10424426GA SED09330576GA SED10132382GA SED10132386GA SED10132390GA SED10132391GA SED10132397GA SED10132400GA SED10132404GA SED10132405GA SED10132406GA SED10132411GA SED10132416GA SED10132417GA SED10132419GA SED10132431GA SED10495312GA SED10536354GA SED10536357GA SED10536358GA SED10536359GA SED10536366GA SED10536367GA SED10536404GA SED10536405GA SED10536406GA SED10536407GA SED10536409GA SED10536410GA SED10536411GA SED10536412GA SED10536415GA SED10536421GA SED10536425GA SED10536427GA SED10293116GA SED10293117GA SED10293136GA SED10293140GA SED10293144GA SED10293156GA SED10293157GA SED10293159GA SED10303377GA SED10303380GA SED10303381GA SED10303385GA SED10303386GA SED10303387GA SED10303390GA SED10303391GA SED10303393GA SED10303395GA SED10303408GA SED10303409GA SED10303410GA SED10303411GA SED10303412GA SED10303413GA SED10373922GA SED10373931GA SED10373934GA SED10373949GA SED10373951GA SED10373952GA SED10373953GA SED10373954GA SED10373956GA SED10378924GA SED10061891GA SED10092035GA SED10092036GA SED10092040GA SED10092043GA SED10092044GA SED10333657GA SED10333658GA SED10333659GA SED10333660GA SED10333661GA SED10333662GA SED10333663GA SED10343664GA SED10343665GA SED10343666GA SED10343667GA SED10343668GA SED10343669GA SED10343670GA SED10343671GA SED10343672GA SED10343673GA SED10343674GA SED10343675GA SED10343676GA SED10343677GA SED10343678GA SED10343679GA SED10343680GA SED10343681GA SED10343682GA SED10343683GA SED10343684GA SED10343685GA SED10343686GA SED10343687GA SED10343688GA SED10343689GA SED10343690GA SED10343691GA SED10343692GA SED10343693GA SED10343694GA SED10343695GA SED10343696GA SED10343697GA SED10343698GA SED10343699GA SED10343700GA SED10343701GA SED10343702GA SED10343703GA SED10343704GA SED10343705GA SED10343706GA SED09320544GA SED09330577GA SED09421005GA SED09441076GA SED09441088GA SED10031741GA SED10031745GA SED10061884GA SED10343707GA SED10343708GA SED10343709GA SED10343710GA SED10343711GA SED10343712GA SED10343713GA SED10384029GA SED10464952GA SED09310480GA SED09340663GA SED09330588GA SED09310481GA SED09320549GA SED09441086GA SED09310464GA SED09320533GA SED09310516GA SED10293132GA SED10293133GA SED10293138GA SED10293141GA SED10293146GA SED10293149GA SED10293150GA SED10293151GA SED10293153GA SED10293154GA SED10293155GA SED10293160GA SED10293162GA SED10293165GA SED10293166GA SED10293167GA SED10293168GA SED10293169GA SED10293170GA SED10293171GA SED10293173GA SED10293174GA SED10293179GA SED10293180GA SED10293181GA SED10293183GA SED10293191GA SED10293192GA SED10293214GA SED10293215GA SED10293218GA SED10293222GA SED10353766GA SED11036580GA SED11036593GA SED09280411GA SED09280431GA SED09310472GA SED09310478GA SED09310492GA SED09310499GA SED09310504GA SED09310505GA SED09310509GA SED09310513GA SED09310518GA SED09320553GA SED09320557GA SED09320562GA SED09320563GA SED09330567GA SED09330575GA SED09330589GA SED09410978GA SED09420985GA SED09420997GA SED09441087GA SED09441089GA SED09130294P SED10122322GA SED10182691GA SED10182692GA SED10293119GA SED101323367GA SED10132385GA SED10132389GA SED10132394GA SED10132395GA SED10132396GA SED10132398GA SED10132399GA SED101324036A SED10132407GA SED10132387GA SED10132403GA SED10132506GA SED10152633GA SED10071938GA SED10082000GA SED10102102GA SED10454803GA SED10454810GA SED10132420GA SED10222834GA SED10222836GA SED09310491GA SED09310496GA SED09320551GA SED09320543GA SED09420992GA SED09310473GA SED09441085GA SED09370815GA SED10041834GA SED10071942GA SED10081998GA SED10102106GA SED10102108GA SED10102114GA SED10102118GA SED09330592GA SED10536456GA SED10536462GA SED10536504GA SED10536506GA SED09030232GP SED09130284GP SED09130287GP SED09130289GP SED09130293GP SED09130294GP SED09130295GP SED09130297GP SED09130303GP SED09130311GP SED09180325GP SED09180326GP SED09180337GP SED09220354GP SED09220364GP SED09220368GP SED09220376GP SED09230387GA SED09230393GA SED09240392GP SED09280408GA SED09280438GA SED09280440GA SED09280441GA SED09280442GR SED09310510GA SED09310522GR SED09320538GA SED09340617GA SED09420998GA SED09431061GA SED09441090GA SED09441091GA SED09441092GA SED09441093GA SED09441094GA SED09441095GA SED09441096GA SED09441097GA SED09441098GA SED09441099GA SED09441100GA SED09441101GA SED09441102GA SED09441103GA SED09441104GA SED09441105GA SED09441106GA SED09441107GA SED09441108GA SED09441109GA SED09441110GA SED09441111GA SED09441112GA SED09441113GA SED09441114GA SED09451361GA SED09471466GA SED09491602GA SED09491608GA SED10263000GA SED10263002GA SED10263006GA SED10273036GA SED10273052GA SED10273053GA SED10273055GA SED10273057GA SED10273058GA SED10273059GA SED10283078GA SED10283079GA SED10283080GA SED10283081GA SED10283082GA SED10283083GA SED10283084GA SED10283085GA SED10283086GA SED10283087GA SED10373936GA SED10373937GA SED10373941GA SED10373959GA SED10373965GA SED10373967GA SED10373968GA SED10373969GA SED10373979GA SED10373989GA SED10373996GA SED10373997GA SED10373999GA SED10384007GA SED10384013GA SED10384024GA SED10384042GA SED10394173GA SED09320541GA SED10373925GA SED10373926GA SED09320531GA SED09451345GA SED09491601GA SED10122318GA SED10122352GA SED10122361GA SED10122363GA SED10122365GA SED10122371GA SED10122372GA SED10122373GA SED10122375GA SED10122376GA SED10122377GA SED10363863GA SED10404272GA SED10434574GA SED10434575GA SED10434576GA SED10434577GA SED10434578GA SED10434580GA SED10434581GA SED10434582GA SED10434583GA SED10434584GA SED10434585GA SED10434586GA SED10434587GA SED10434588GA SED10434589GA SED10434590GA SED10434591GA SED09330582GA SED09310483GA SED09310446GA SED10112141GA SED10293121GA SED10293122GA SED09330586GA SED09310525GA SED09310462GA SED09310515GA SED09320530GA SED10081974GA SED10112179GA SED10112246GA SED10112249GA
SED10112270GA SED10122299GA SED10122304GA SED10122306GA SED10122320GA SED10122329GA SED10122380GA SED10132381GA SED10132383GA SED10132384GA SED09330593GA SED09310503GA SED09310461GA SED09320565GA SED09320536GA SED10102064GA SED10102074GA SED10102077GA SED10102078GA SED10102087GA SED10102088GA SED10102089GA SED10102094GA SED09330585GA SED09441083GA SED09451344GA SED09451346GA SED09471494GA SED09511667GA SED09451342GA SED09491600GA SED09491603GA SED10464947GA SED10464948GA SED10293158GA SED10293161GA SED09471487GA SED10112258GA SED10343722GA SED10343723GA SED10343724GA SED10343725GA SED10363838GA SED10363839GA SED10363841GA SED10363843GA SED10373914GA SED10454745GA SED10454746GA SED10454804GA SED10454805GA SED10454806GA SED10454807GA SED10454808GA SED10454811GA SED10212799GA SED10212800GA SED10212801GA SED10212802GA SED10212803GA SED10212804GA SED10212805GA SED10212806GA SED10212807GA SED10212808GA SED10212809GA SED10212810GA SED09471477GA SED09491589GA SED09491592GA SED09511675GA SED09320542GA SED09320539GA SED09370799GA SED09310502GA SED09310469GA SED09370806GA SED09310448GA SED09310474GA SED10081999GA SED10102103GA SED10414280GA SED10414281GA SED09320535GA SED09310498GA SED10526271GA SED10526276GA SED10526283GA SED10526289GA SED10526291GA SED10232885GA SED10414283GA SED10414284GA SED10414285GA SED10444675GA SED10444676GA SED10444677GA SED10444678GA SED10444679GA SED10444680GA SED10444681GA SED10444682GA SED10444683GA SED10444684GA SED10444685GA SED10444686GA SED10444687GA SED10444688GA SED10444689GA SED10444690GA SED10444691GA SED10444692GA SED10444693GA SED09320561GA SED09310477GA SED09310453GA SED09330587GA SED09441084GA SED09310458GA SED09340662GA SED09330571GA SED09420981GA SED09310514GA SED10112193GA SED10112194GA SED10112247GA SED10112248GA SED10112261GA SED10112264GA SED10122298GA SED10122300GA SED10122308GA SED10122336GA SED09330584GA SED11056919GA SED11056920GA SED11056921GA SED11056922GA SED11036573GA SED11036576GA SED11036578GA SED11036579GA SED11036581GA SED11036590GA SED11036591GA SED11036592GA SED11036594GA SED11036636GA SED11036637GA SED11036638GA SED11036639GA SED11036640GA SED11036642GA SED11036644GA SED11036645GA SED10283076GA SED09370830GA SED09390846GA SED09390850GA SED09390852GA SED09390862GA SED09390863GA SED09390864GA SED09390865GA SED09390866GA SED09390869GA SED09390870GA SED09390873GA SED09390875GA SED09390878GA SED09390880GA SED09390881GA SED09390884GA SED09390885GA SED09390886GA SED09390887GA SED09390888GA SED09390889GA SED09390893GA SED09400897GA SED09400898GA SED09400899GA SED09400900GA SED09400901GA SED09400902GA SED09400903GA SED09400904GA SED09400905GA SED09400907GA SED09400909GA SED09400910GA SED09400911GA SED09400912GA SED09400913GA SED09400914GA SED09400915GA SED09400916GA SED09400917GA SED09400919GA SED09400920GA SED09400921GA SED09400922GA SED09400925GA SED09400943GA SED09400946GA SED09400949GA SED09400951GA SED09471490GA SED10192741GA SED10212778GA SED10212812GA SED10212813GA SED10212814GA SED10212815GA SED10222816GA SED10222817GA SED10222818GA SED10222819GA SED10222820GA SED10222821GA SED10222822GA SED10222823GA SED10222824GA SED10222825GA SED10222826GA SED10222827GA SED10222828GA SED10222829GA SED10222830GA SED10222831GA SED10222835GA SED10222843GA SED10222844GA SED10222845GA SED10222846GA SED10222847GA SED10222848GA SED10222850GA SED10222851GA SED10262992GA SED10293270GA SED10293271GA SED10293272GA SED10293273GA SED10293274GA SED10303296GA SED10303297GA SED10303298GA SED10303300GA SED10303301GA SED10303302GA SED10303303GA SED10303304GA SED10303305GA SED10303306GA SED10303307GA SED10303308GA SED10303309GA SED10303310GA SED10303311GA SED10303312GA SED10303313GA SED10303314GA SED10303315GA SED10303316GA SED10303317GA SED10303318GA SED10303319GA SED10303382GA SED10303396GA SED10303397GA SED10303398GA SED10303399GA SED10303400GA SED10303403GA SED10303407GA SED10303414GA SED10303416GA SED10313503GA SED10313504GA SED10313505GA SED10313506GA SED10313507GA SED10313508GA SED10313509GA SED10313510GA SED10313511GA SED10313512GA SED10313513GA SED10313514GA SED10323579GA SED10454735GA SED10454736GA SED10454737GA SED10454741GA SED10454742GA SED10454743GA SED10454744GA SED10454747GA SED10454748GA SED10454750GA SED10454751GA SED10454752GA SED10454753GA SED10454755GA SED10454757GA SED10454758GA SED10454763GA SED10454765GA SED10454766GA SED10132388GA SED10132409GA SED09310470GA SED09320534GA SED09370803GA SED09370804GA SED09370805GA SED09370812GA SED09370813GA SED09370814GA SED09370817GA SED09370821GA SED09370829GA SED09390836GA SED09390838GA SED09390839GA SED09390840GA SED09390842GA SED09390843GA SED09390876GA SED09390882GA SED09390890GA SED09400894GA SED10526233GA SED10526238GA SED10526240GA SED10526241GA SED10526243GA SED10526246GA SED10526247GA SED10526248GA SED10526249GA SED10526252GA SED10526254GA SED10526261GA SED10526264GA SED10526267GA SED10526268GA SED10526269GA SED10526270GA SED10526272GA SED10526273GA SED10526274GA SED10526275GA SED10526277GA SED10526278GA SED10526279GA SED10526281GA SED10526282GA SED10526284GA SED10526285GA SED10526287GA SED10526288GA SED10526290GA SED10526292GA SED09400941GA SED09400945GA SED09400948GA SED09330573GA SED09451303GA SED09461399GA SED09461402GA SED09461406GA SED09461411GA SED09491519GA SED09491520GA SED09491522GA SED09491523GA SED09491524GA SED09491526GA SED09491527GA SED09491528GA SED09491534GA SED09491536GA SED09491537GA SED09491540GA SED09491543GA SED09491544GA SED09491545GA SED09491547GA SED09491548GA SED09491564GA SED09491565GA SED09491567GA SED09491569GA SED09491572GA SED09491575GA SED09491576GA SED09491577GA SED09491578GA SED10071897GA SED10092028GA SED10212811GA SED10323588GA SED09330581GA SED10132516GA SED10142524GA SED10142526GA SED10142528GA SED10252889GA SED10252891GA SED10252895GA SED10252920GA SED10252921GA SED10252922GA SED10252926GA SED09320537GA SED09320548GA SED09310447GA SED09340609GA SED09310494GA SED10373921GA SED10373923GA SED10373945GA SED10333607GA SED10333608GA SED10333609GA SED10333632GA SED10333633GA SED10333634GA SED10333635GA SED10333636GA SED10021697GA SED10021714GA SED09451319GA SED09461407GA SED09461410GA SED09461412GA SED09461416GA SED09471431GA SED09471433GA SED09471440GA SED09471441GA SED09471442GA SED10313480GA SED10313481GA SED10313482GA SED10313483GA SED10313484GA SED10313485GA SED10313488GA SED10313489GA SED10313490GA SED10313491GA SED10313492GA SED10313493GA SED10313494GA SED10313495GA SED10061883GA SED10071905GA SED10082014GA SED10092024GA SED10092031GA SED10092033GA SED10092041GA SED10092042GA SED10092047GA SED09451322GA SED09471479GA SED09471483GA SED09491579GA SED09491590GA SED09491594GA SED11056971GA SED11056972GA SED11056973GA SED11056974GA SED11056975GA SED11056976GA SED11056985GA SED11056988GA SED11056989GA SED11056990GA SED11056991GA SED11056992GA SED11056993GA SED11056994GA SED11056995GA SED11056996GA SED11057068GA SED11066998GA SED11067001GA SED11067041GA SED11067048GA SED11067053GA SED11067075GA SED11067076GA SED11067079GA SED10424458GA SED10424459GA SED10424460GA SED10424461GA SED10424462GA SED10424463GA SED10424464GA SED10434515GA SED10434516GA SED10434517GA SED10434519GA SED10434520GA SED10434521GA SED10434522GA SED10434523GA SED10434524GA SED10434525GA SED10434526GA SED10434527GA SED10434528GA SED10434529GA SED10434530GA SED10434531GA SED10464972GA SED10475021GA SED10475022GA SED10495188GA SED10495189GA SED10495205GA SED10495224GA SED10495262GA SED10495293GA SED10495303GA SED10495304GA SED10495306GA SED10495308GA SED10495313GA SED10495314GA SED10495315GA SED10495320GA SED10505359GA SED10505360GA SED10505361GA SED10505362GA SED10505363GA SED10505364GA SED10505365GA SED10505369GA SED10505374GA SED10505388GA SED10505404GA SED10505514GA SED10505534GA SED10505597GA SED10505598GA SED10536460GA SED10536489GA SED11036582GA SED11036653GA SED11046715GA SED11046721GA SED11046732GA SED11046735GA SED09370744GA SED09491529GA SED10021721GA SED10021723GA SED10021724GA SED10021733GA SED10031736GA SED10031737GA SED10031738GA SED10031739GA SED10031740GA SED10031742GA SED10031743GA SED10031746GA SED10031748GA SED10031755GA SED10031758GA SED10031762GA SED10031778GA SED10041839GA SED10041843GA SED10041844GA SED10041845GA SED10041848GA SED10041853GA SED10041854GA SED10041856GA SED10041857GA SED10041868GA SED10041869GA SED10041870GA SED10041872GA SED10061890GA SED10142551GA SED10142552GA SED10142553GA SED10384090GA SED10394175GA SED10394176GA SED10404203GA SED10404204GA SED10404209GA SED10404210GA SED10404257GA SED10424456GA SED10424457GA SED10536511GA SED11057063GA SED11067047GA SED11067050GA SED11067052GA SED11067071GA SED11067126GA SED11067127GA SED11067138GA SED11077146GA SED11077167GA SED11077170GA SED11077173GA SED11077174GA SED11077187GA SED11087309GA SED11087310GA SED11087311GA SED11087314GA SED11087316GA SED11087318GA SED11087349GA SED11087350GA SED11087354GA SED11087357GA SED11087363GA SED11087390GA SED11087391GA SED11087408GA SED11087410GA SED09310466GA SED09310479GA SED09310482GA SED09310517GA SED09340628GA SED09340637GA SED09340647GA SED09370723GA SED09370725GA SED09370726GA SED09370727GA SED09370729GA SED09370730GA SED09370738GA SED09370739GA SED09370741GA SED09370746GA SED09370747GA SED09370748GA SED09370749GA SED09370750GA SED09370752GA SED09370753GA SED09370798GA SED10041822GA SED09280420GA SED09280422GA SED09280423GA SED09280432GA SED09461381GA SED09461385GA SED09461389GA SED09461392GA SED09461398GA SED10071893GA SED10071895GA SED10071898GA SED10071906GA SED10071909GA SED10071912GA SED10333595GA SED10333598GA SED10333599GA SED10333600GA SED10333601GA SED10333602GA SED10333603GA SED10333604GA SED10333605GA SED10333606GA SED10102132GA SED09431059GA SED09441180GA SED09441286GA SED09451311GA SED09451312GA SED09451313GA SED09451314GA SED09451315GA SED09451316GA SED09451318GA SED09461379GA SED09461380GA SED09461382GA SED09461383GA SED09461384GA SED09461386GA SED09461387GA SED09461388GA SED09461390GA SED09461391GA SED09461393GA SED09461394GA SED09461396GA SED09461397GA SED09461400GA SED09461401GA SED09461403GA SED09461404GA SED09461405GA SED09461408GA SED09461409GA SED09461414GA SED09471448GA SED09471450GA SED09471452GA SED09471456GA SED09471457GA SED09491614GA SED09491615GA SED09491617GA SED09491619GA SED10132441GA SED10132442GA SED10132443GA SED10132445GA SED10132446GA SED10132447GA SED10132479GA SED10132483GA SED10041840GA SED10041865GA SED10041866GA SED10041867GA SED10041871GA SED10071925GA SED10071926GA SED10071927GA SED10071928GA SED10071929GA SED10071930GA SED10071932GA SED10071933GA SED10071934GA SED10071935GA SED10071936GA SED10081943GA SED10081944GA SED10081945GA SED10081946GA SED10081948GA SED10081949GA SED10081950GA SED10081951GA SED10081952GA SED10081953GA SED10081954GA SED10081957GA SED10081958GA SED10081960GA SED10081963GA SED10081967GA SED10081968GA SED10081970GA SED10081971GA SED10081972GA SED10081975GA SED10081977GA SED10081979GA SED10081980GA SED10081983GA SED10081987GA SED10081992GA SED10112242GA SED10112148GA SED10041863GA SED10071899GA SED10071911GA SED10071917GA SED10071919GA SED10071923GA SED10112184GA SED10112215GA SED10112216GA SED10112220GA SED10112233GA SED10112235GA SED10112236GA SED10112240GA SED09421004GA SED09421006GA SED09421007GA SED09421009GA
SED09421010GA SED09421011GA SED09431063GA SED09431064GA SED09431065GA SED09431067GA SED09431068GA SED09431070GA SED09431071GA SED09431074GA SED09441078GA SED09441157GA SED09441158GA SED09441159GA SED09441160GA SED09441161GA SED09441162GA SED09441165GA SED09441168GA SED09441172GA SED09441173GA SED09441175GA SED09441177GA SED09441178GA SED09441186GA SED09441189GA SED09441195GA SED09441198GA SED09441199GA SED09441200GA SED09441201GA SED09441202GA SED09441203GA SED09441204GA SED09441206GA SED09441207GA SED09441208GA SED09441209GA SED09441210GA SED09441211GA SED09441212GA SED09441277GA SED09451351GA SED09451366GA SED09451367GA SED09451368GA SED09451370GA SED09451371GA SED09451372GA SED09451373GA SED09451374GA SED09451375GA SED09451377GA SED09451378GA SED10132472GA SED10132480GA SED10132496GA SED10142541GA SED10142580GA SED10142590GA SED10152636GA SED10152638GA SED10152639GA SED10152640GA SED10152643GA SED10152658GA SED10152660GA SED10162675GA SED10162678GA SED10162683GA SED10182707GA SED10182708GA SED10182710GA SED10182711GA SED10252918GA SED10081978GA SED10081982GA SED10081991GA SED10081993GA SED10081996GA SED10082005GA SED10152601GA SED10424504GA SED09451353GA SED10384116GA SED10071941GA SED10122366GA SED09390837GA SED09390851GA SED09390855GA SED09390858GA SED09451317GA SED09471465GA SED09511639GA SED10021687GA SED10021726GA SED10021732GA SED10041809GA SED10041816GA SED10041819GA SED10041821GA SED10041826GA SED10041827GA SED10041828GA SED10041829GA SED10041830GA SED10041831GA SED10041837GA SED10041838GA SED10041841GA SED10041842GA SED10041846GA SED10041847GA SED10041849GA SED10041850GA SED10041851GA SED10041852GA SED10041855GA SED10041858GA SED10041860GA SED10041861GA SED10041864GA SED10061874GA SED10071922GA SED10071924GA SED10112190GA SED10112198GA SED10112208GA SED10112209GA SED10112210GA SED10112214GA SED10112217GA SED10112218GA SED10112219GA SED10112221GA SED10112222GA SED10112223GA SED10112224GA SED10112225GA SED10112227GA SED10112229GA SED10112230GA SED10112231GA SED10112232GA SED10112234GA SED10112237GA SED10112238GA SED10112239GA SED10112241GA SED10112243GA SED10112244GA SED10112245GA SED10112250GA SED10112251GA SED10112266GA SED10112279GA SED10122301GA SED10122325GA SED10122326GA SED10122356GA SED10122362GA SED10313486GA SED10313487GA SED10343715GA SED10343716GA SED10343717GA SED10343718GA SED10343719GA SED10343720GA SED10343721GA SED10353767GA SED10353768GA SED10353769GA SED10353770GA SED10353771GA SED10353772GA SED10424454GA SED10424455GA SED10333648GA SED10505427GA SED10505452GA SED10505497GA SED10505517GA SED10505518GA SED10505527GA SED10505541GA SED10505547GA SED10505549GA SED10505570GA SED10505600GA SED10505603GA SED10505611GA SED09310497GA SED10142539GA SED10142545GA SED10142584GA SED10152608GA SED10182693GA SED10182694GA SED10182695GA SED10182696GA SED10303388GA SED10303405GA SED10303415GA SED10303422GA SED10434567GA SED10434608GA SED10485163GA SED10515970GA SED10515972GA SED10516006GA SED10516008GA SED10516037GA SED10516047GA SED10516059GA SED11056935GA SED11056936GA SED11056956GA SED11056958GA SED09280413GA SED09340645GA SED09340648GA SED09441081GA SED10333641GA SED10333642GA SED10333643GA SED10333644GA SED10333645GA SED10333646GA SED10333647GA SED09390849GA SED09390868GA SED10152616GA SED10192737GA SED10192738GA SED10192739GA SED10192740GA SED10192743GA SED10192744GA SED10192745GA SED10192746GA SED10192747GA SED10192748GA SED10192749GA SED10192750GA SED10192751GA SED10212753GA SED10212754GA SED10212755GA SED10212756GA SED10212757GA SED10212758GA SED10212760GA SED10212761GA SED10212762GA SED10212763GA SED10212764GA SED10212765GA SED10212766GA SED10212767GA SED10212768GA SED10212769GA SED10212770GA SED10212771GA SED10212772GA SED10212773GA SED10212774GA SED10212775GA SED10212776GA SED10212777GA SED10212779GA SED10212780GA SED10212781GA SED10212782GA SED10212783GA SED10212784GA SED10212785GA SED09390871GA SED09400895GA SED09410956GA SED09410967GA SED09410975GA SED09410977GA SED09421014GA SED10252907GA SED10252910GA SED10424411GA SED10434552GA SED10434553GA SED10434554GA SED10434566GA SED10434568GA SED10434569GA SED10434570GA SED10434571GA SED10434572GA SED10434573GA SED10434579GA SED10434626GA SED10434627GA SED10434628GA SED10434630GA SED10434631GA SED10434632GA SED10434633GA SED10434634GA SED10434635GA SED10434636GA SED10434637GA SED10434638GA SED10434639GA SED10434640GA SED10464970GA SED10515967GA SED10515968GA SED10515969GA SED10516030GA SED10516032GA SED10516033GA SED10516051GA SED10516052GA SED10516056GA SED10516057GA SED10516070GA SED10526180GA SED10526212GA SED10526234GA SED10526235GA SED10526236GA SED10526237GA SED10526239GA SED10526242GA SED10526250GA SED10526251GA SED10526257GA SED10526258GA SED10526259GA SED10526260GA SED10526262GA SED10536451GA SED10536461GA SED10536510GA SED11036634GA SED11036643GA SED11046741GA SED11046744GA SED10192742GA SED10212752GA SED10434592GA SED10434594GA SED10434595GA SED10434596GA SED10434597GA SED10434598GA SED10434599GA SED10434600GA SED10434601GA SED10434602GA SED10434603GA SED10434604GA SED10434605GA SED10434606GA SED10434607GA SED10434609GA SED10434610GA SED10434611GA SED10434612GA SED10434613GA SED10434614GA SED10434615GA SED10434616GA SED10434617GA SED10434618GA SED10434619GA SED10434620GA SED10434621GA SED10434622GA SED10434623GA SED10434624GA SED10434625GA SED10464914GA SED10464915GA SED10464921GA SED10464922GA SED10464923GA SED10464925GA SED10464926GA SED10464927GA SED10464928GA SED10464929GA SED10464930GA SED10464931GA SED10464932GA SED10464933GA SED10464934GA SED10464935GA SED10464936GA SED10464937GA SED10464938GA SED10464939GA SED10464940GA SED10464942GA SED10464943GA SED10464944GA SED10464946GA SED10475008GA SED10475027GA SED10475028GA SED10475029GA SED10475030GA SED10475031GA SED10475032GA SED10485037GA SED10485038GA SED10485039GA SED10485040GA SED10485041GA SED10485042GA SED10485043GA SED10485044GA SED10485045GA SED10485046GA SED10485053GA SED10526300GA SED10526302GA SED10526306GA SED10536313GA SED10536314GA SED10536315GA SED10536316GA SED10536317GA SED10536318GA SED10536319GA SED10536320GA SED10536322GA SED10536323GA SED10536324GA SED10536333GA SED10536334GA SED10536338GA SED10536339GA SED10536340GA SED10536342GA SED10536343GA SED10536344GA SED10536345GA SED10536347GA SED10536353GA SED10536355GA SED10536356GA SED10536433GA SED10536523GA SED11067112GA SED11077152GA SED11087312GA SED11087315GA SED11087353GA SED11087358GA SED11087407GA SED11087411GA SED11087412GA SED10152628GA SED10162670GA SED10162672GA SED10162674GA SED10162677GA SED10162681GA SED10162684GA SED10162690GA SED09280426GA SED09280428GA SED10132427GA SED10283074GA SED10283075GA SED09370824GA SED09390845GA SED09280412GA SED09280445GA SED09420986GA SED10152602GA SED10152605GA SED10373935GA SED10373938GA SED10373939GA SED10384050GA SED10384065GA SED10384066GA SED10384071GA SED10384079GA SED10384091GA SED10384098GA SED10404207GA SED10404264GA SED10404265GA SED10404266GA SED10404267GA SED10414276GA SED10434532GA SED10434533GA SED10434534GA SED10434535GA SED10434537GA SED10434538GA SED10434539GA SED10434540GA SED10434541GA SED10434542GA SED10434555GA SED10434556GA SED10434557GA SED10434558GA SED10444654GA SED10444715GA SED10444716GA SED10444717GA SED10444718GA SED10444719GA SED10444720GA SED10444721GA SED10444722GA SED10444723GA SED10444724GA SED10444725GA SED10444727GA SED10454728GA SED10454729GA SED10454730GA SED10454731GA SED09420987GA SED10102123GA SED09400926GA SED09400935GA SED09400936GA SED09400940GA SED09421012GA SED09421019GA SED09431025GA SED09431034GA SED09431036GA SED09431038GA SED09431039GA SED09431040GA SED09431042GA SED09431043GA SED09431044GA SED09431056GA SED09431057GA SED10041818GA SED10041823GA SED10041825GA SED10082003GA SED10082017GA SED10092023GA SED10092026GA SED10092027GA SED10102050GA SED10102054GA SED10102055GA SED10102057GA SED10102059GA SED10102060GA SED10102066GA SED10102069GA SED10102076GA SED10102084GA SED10102090GA SED10102091GA SED10102061GA SED10102063GA SED10102067GA SED10102068GA SED10102070GA SED10102071GA SED10102072GA SED10102073GA SED10102081GA SED10102092GA SED10071908GA SED10262959GA SED10262974GA SED10262975GA SED10262977GA SED10262978GA SED10262981GA SED10303320GA SED10303321GA SED10303322GA SED10303323GA SED10303324GA SED10303325GA SED10303326GA SED10303327GA SED10303328GA SED10303329GA SED10303334GA SED10303336GA SED10303337GA SED10303338GA SED10303339GA SED10303340GA SED10303341GA SED10303342GA SED10303343GA SED10303344GA SED10303345GA SED10303352GA SED10303354GA SED10303355GA SED10303356GA SED10303357GA SED10303358GA SED10303359GA SED10303360GA SED10303363GA SED10303364GA SED10303365GA SED10303367GA SED10303368GA SED10303369GA SED10303370GA SED10303371GA SED10303372GA SED10303373GA SED10303374GA SED10303375GA SED10303376GA SED10303378GA SED10303379GA SED10303383GA SED10303384GA SED10303389GA SED10303394GA SED10303404GA SED10303406GA SED10303417GA SED10303418GA SED10303424GA SED10303426GA SED10303427GA SED10303428GA SED10303429GA SED10303430GA SED10303431GA SED10303432GA SED10303433GA SED10303434GA SED10303435GA SED10303436GA SED10303437GA SED10303438GA SED10303442GA SED10303445GA SED10303448GA SED10303456GA SED10303457GA SED10373994GA SED10414289GA SED10414290GA SED10414291GA SED10414292GA SED10414293GA SED10414294GA SED10414295GA SED10414296GA SED10414297GA SED10414298GA SED10414299GA SED10414300GA SED10414301GA SED10414302GA SED10414303GA SED10414304GA SED10414305GA SED10414306GA SED10414307GA SED10414308GA SED10414309GA SED10414310GA SED10414311GA SED10414312GA SED10414318GA SED10414319GA SED10414320GA SED10414321GA SED10414322GA SED10414323GA SED10414324GA SED10414325GA SED10414326GA SED10414327GA SED10414328GA SED10414329GA SED10414330GA SED10414331GA SED10414332GA SED10414333GA SED10414334GA SED10414335GA SED10414336GA SED10414337GA SED10414338GA SED10414339GA SED10414340GA SED10414341GA SED10414342GA SED10414343GA SED10414344GA SED10414345GA SED10414346GA SED10414347GA SED10414348GA SED10414349GA SED10414350GA SED10414351GA SED10414352GA SED10424356GA SED10424358GA SED10424359GA SED10424360GA SED10424361GA SED10424362GA SED10424363GA SED10424364GA SED10424365GA SED10424366GA SED10424367GA SED10424368GA SED10424369GA SED10424370GA SED10424371GA SED10424372GA SED10516028GA SED10516031GA SED10516049GA SED10516058GA SED11056952GA SED10102085GA SED10102086GA SED10102137GA SED10404206GA SED10404254GA SED10404255GA SED10404256GA SED10424471GA SED10424472GA SED10424473GA SED10424474GA SED10526125GA SED10526130GA SED09280409GA SED09280430GA SED10303392GA SED10021700GA SED10021701GA SED10021704GA SED10021705GA SED10021706GA SED10021710GA SED10031747GA SED10031750GA SED10031751GA SED10031752GA SED10031753GA SED10031754GA SED10031756GA SED10031757GA SED10031759GA SED10031760GA SED10031761GA SED10031763GA SED10031764GA SED10031765GA SED10031766GA SED10031767GA SED10031768GA SED10031769GA SED10031771GA SED10031773GA SED10031774GA SED10031775GA SED10031777GA SED10031779GA SED10031780GA SED10031781GA SED10031782GA SED10031783GA SED10031784GA SED10031785GA SED10031786GA SED10031787GA SED10031788GA SED10031789GA SED10031790GA SED10031791GA SED10031792GA SED10031793GA SED10031794GA SED10041795GA SED10041796GA SED10041797GA
SED10041798GA SED10041799GA SED10041800GA SED10041801GA SED10041802GA SED10041803GA SED10041804GA SED10041805GA SED10041807GA SED10041808GA SED10041810GA SED10041811GA SED10041813GA SED10041814GA SED10132392GA SED10132393GA SED10132401GA SED10132402GA SED09531680GA SED10293164GA SED10293172GA SED10293176GA SED10293182GA SED10293184GA SED10293185GA SED10293186GA SED10293187GA SED10293194GA SED10293195GA SED10293196GA SED10293197GA SED10293198GA SED10293199GA SED10293200GA SED10293201GA SED10293202GA SED10293203GA SED10293204GA SED10293205GA SED10293206GA SED10293207GA SED10293208GA SED10293209GA SED10293210GA SED10293211GA SED10293212GA SED10293213GA SED10293216GA SED10293217GA SED10293219GA SED10293220GA SED10293221GA SED10293223GA SED10293224GA SED10293225GA SED10293226GA SED10293227GA SED10293228GA SED10293229GA SED10293230GA SED10293231GA SED10293232GA SED10293233GA SED10293234GA SED10293235GA SED10293236GA SED10293237GA SED10293238GA SED10293239GA SED10293240GA SED10293241GA SED10293242GA SED10293243GA SED10293244GA SED10293245GA SED10293246GA SED10293247GA SED10293248GA SED10293249GA SED10293250GA SED10293251GA SED10293252GA SED10293253GA SED10293254GA SED10293255GA SED10293256GA SED10293257GA SED10293258GA SED10293259GA SED10293260GA SED10293261GA SED10293262GA SED10293263GA SED10293264GA SED10293265GA SED10293266GA SED10293267GA SED10293268GA SED10293269GA SED10293275GA SED10293276GA SED10293277GA SED10293278GA SED10293279GA SED10293280GA SED10293281GA SED10293282GA SED10293283GA SED10293284GA SED10293285GA SED10303286GA SED10303287GA SED10303288GA SED10303289GA SED10303290GA SED10303291GA SED10303292GA SED10303293GA SED10303294GA SED10303295GA SED10061873GA SED10061878GA SED10303361GA SED10303362GA SED10323521GA SED10323522GA SED10323523GA SED10323524GA SED10323525GA SED10323526GA SED10323527GA SED10323528GA SED10323529GA SED10323530GA SED10323531GA SED10323532GA SED10323533GA SED10323534GA SED10323535GA SED10323536GA SED10323537GA SED10323538GA SED10323539GA SED10323540GA SED10323541GA SED10323542GA SED10323543GA SED10323544GA SED10323545GA SED10323546GA SED10323547GA SED10323548GA SED10323549GA SED10323550GA SED10323551GA SED10323552GA SED10323553GA SED10323554GA SED10323555GA SED10323556GA SED10323557GA SED10323558GA SED10323559GA SED10323560GA SED10323561GA SED10323562GA SED10323563GA SED10323564GA SED10323565GA SED10323566GA SED10323567GA SED10323568GA SED10323569GA SED10323570GA SED10323571GA SED10323572GA SED10323573GA SED10323574GA SED10323575GA SED10323576GA SED10323577GA SED09451347GA SED10343730GA SED10343731GA SED10343732GA SED10343733GA SED10343734GA SED10343735GA SED10353736GA SED10353737GA SED10353738GA SED10353739GA SED10353740GA SED10353741GA SED10434518GA SED10495326GA SED10505356GA SED10505372GA SED10505377GA SED10505381GA SED10505385GA SED10505421GA SED10505423GA SED10505432GA SED10505435GA SED10505436GA SED10505444GA SED10505445GA SED10505463GA SED10505468GA SED10505471GA SED10505477GA SED10505712GA SED09340653GA SED09340661GA SED10495280GA SED10495281GA SED10495295GA SED10495299GA SED10495302GA SED10495347GA SED10505349GA SED10505376GA SED10505392GA SED10505393GA SED10505394GA SED10505417GA SED10505446GA SED10515818GA SED10515871GA SED10515884GA SED11067111GA SED11067116GA SED11067122GA SED11077231GA SED10132448GA SED10132451GA SED10132455GA SED10132457GA SED10132460GA SED10132463GA SED10404190GA SED10404220GA SED10404227GA SED10404229GA SED10404233GA SED10404235GA SED10404240GA SED10404242GA SED10536326GA SED10536327GA SED10536332GA SED10536483GA SED10536485GA SED10536502GA SED11036612GA SED11036615GA SED11036616GA SED11036659GA SED11036661GA SED11036663GA SED11036664GA SED11036665GA SED11036676GA SED11036677GA SED11036678GA SED11036679GA SED11036681GA SED11036690GA SED11036691GA SED11036693GA SED11046708GA SED11046714GA SED09410958GA SED10404185GA SED10424475GA SED10424476GA SED10424477GA SED10424478GA SED10424479GA SED10424480GA SED10424481GA SED10424482GA SED10424483GA SED10424484GA SED10424486GA SED10424487GA SED10424488GA SED10454816GA SED10454817GA SED10454818GA SED10454819GA SED10454820GA SED10454821GA SED10454822GA SED10454823GA SED10454824GA SED10454825GA SED10454826GA SED10454827GA SED10464828GA SED10464829GA SED10464830GA SED10464831GA SED10464832GA SED10464834GA SED10536441GA SED10536446GA SED10536452GA SED10536465GA SED10536468GA SED10536478GA SED10536530GA SED11026540GA SED11026542GA SED11026548GA SED11026566GA SED11026569GA SED11026571GA SED11026572GA SED10112180GA SED10112186GA SED10112188GA SED10112200GA SED10404193GA SED10404228GA SED10404231GA SED10404232GA SED10404239GA SED10404241GA SED10404243GA SED10404244GA SED10384052GA SED10384062GA SED10384063GA SED10384089GA SED10404183GA SED10404184GA SED09390844GA SED09390853GA SED09390854GA SED09390891GA SED09420995GA SED09491609GA SED09491610GA SED09491613GA SED10122349GA SED10142554GA SED10152615GA SED10152620GA SED10152629GA SED10152635GA SED10152642GA SED10162665GA SED10162666GA SED10162671GA SED10162680GA SED10162686GA SED10495322GA SED10505454GA SED10505460GA SED10505483GA SED10505484GA SED10505489GA SED10505490GA SED10505494GA SED10505495GA SED10505501GA SED10505505GA SED10505522GA SED10505525GA SED10505526GA SED11056950GA SED09491535GA SED09491538GA SED09431054GA SED09431055GA SED09431058GA SED09431060GA SED09441255GA SED09441257GA SED09441259GA SED09441261GA SED09441264GA SED09441265GA SED09441267GA SED09441268GA SED09441272GA SED09441273GA SED09441275GA SED09441280GA SED09441282GA SED09441288GA SED09441291GA SED09441294GA SED09451302GA SED10082006GA SED10092037GA SED10071910GA SED10071915GA SED10092025GA SED10092029GA SED10021696GA SED10021699GA SED10021702GA SED10021703GA SED10021709GA SED10021711GA SED10182697GA SED10182698GA SED10182699GA SED09280444GA SED10475000GA SED10475001GA SED10475006GA SED10475007GA SED10475010GA SED10475011GA SED10475012GA SED10475013GA SED10475014GA SED10475015GA SED10475016GA SED10475017GA SED10475018GA SED10475019GA SED10475020GA SED10475023GA SED10485087GA SED10485113GA SED10485120GA SED10021727GA SED10424491GA SED09531682GA SED10353753GA SED10353773GA SED10353774GA SED10353775GA SED10353776GA SED10353777GA SED10353778GA SED10353779GA SED10353780GA SED10353781GA SED10353782GA SED10353783GA SED10353784GA SED10353785GA SED10353786GA SED10353787GA SED10353788GA SED10353789GA SED10353790GA SED10353791GA SED10353792GA SED10353793GA SED10353794GA SED10353795GA SED10363796GA SED10363797GA SED10363798GA SED10363799GA SED10363800GA SED10363801GA SED10363802GA SED10363803GA SED10363804GA SED10363805GA SED10363806GA SED10363807GA SED10363808GA SED10363809GA SED10363810GA SED10363811GA SED10363812GA SED10363813GA SED10363814GA SED10363815GA SED10363816GA SED10363817GA SED10363818GA SED10363819GA SED10363820GA SED10363821GA SED10363822GA SED10373940GA SED10373942GA SED10373943GA SED10373944GA SED10373946GA SED10373947GA SED10373948GA SED10373950GA SED10373955GA SED10373957GA SED10373958GA SED10373960GA SED10373961GA SED10373962GA SED10373963GA SED10373964GA SED10373976GA SED10373982GA SED10373985GA SED10373987GA SED10373990GA SED10384025GA SED10384036GA SED10384037GA SED10384047GA SED10384048GA SED10384051GA SED10384080GA SED10384081GA SED10384082GA SED10384083GA SED10384084GA SED10384097GA SED10384101GA SED10384110GA SED10384113GA SED10384114GA SED10384145GA SED10384146GA SED10384150GA SED10384151GA SED10112267GA SED10132500GA SED10152655GA SED10252919GA SED09431033GA SED09451323GA SED09451326GA SED10102049GA SED09390861GA SED09400937GA SED11036574GA SED11036575GA SED11036666GA SED11036667GA SED11036670GA SED11036671GA SED11036687GA SED11036699GA SED11036700GA SED11036707GA SED11046725GA SED11046727GA SED11046729GA SED11046730GA SED11046731GA SED11046739GA SED11046743GA SED11046746GA SED11067038GA SED11067078GA SED11077159GA SED11077162GA SED11117951GA SED09400933GA SED09400953GA SED10061879GA SED10112187GA SED10112202GA SED10152626GA SED10152632GA SED10152634GA SED10162667GA SED10162668GA SED10353762GA SED10353763GA SED10353764GA SED10353765GA SED10454761GA SED10454764GA SED10454772GA SED10464850GA SED10464851GA SED10464852GA SED10464873GA SED10464874GA SED10464875GA SED10464884GA SED10464885GA SED10464886GA SED10464887GA SED10464888GA SED10464889GA SED10464890GA SED10464891GA SED10464892GA SED10464893GA SED10464894GA SED10464895GA SED10464896GA SED10464897GA SED10464898GA SED10464899GA SED10464900GA SED10464901GA SED10464902GA SED10464903GA SED10464905GA SED10464906GA SED10464908GA SED10464909GA SED10464910GA SED10464911GA SED10464912GA SED10464913GA SED10464916GA SED10464917GA SED10464918GA SED10464919GA SED10464920GA SED10464953GA SED10464954GA SED10464955GA SED10464956GA SED10464957GA SED10464958GA SED10485047GA SED10485048GA SED10485049GA SED10485050GA SED10485051GA SED10485052GA SED10485054GA SED10485055GA SED10485056GA SED10485057GA SED10485058GA SED10485060GA SED10485061GA SED10485062GA SED10485063GA SED10485064GA SED10485065GA SED10485066GA SED10485069GA SED10485070GA SED10485059GA SED10485142GA SED10485144GA SED10485145GA SED10485146GA SED10485147GA SED10485148GA SED10485149GA SED10485150GA SED10485151GA SED10485152GA SED10485153GA SED10485154GA SED10485155GA SED10485156GA SED10485157GA SED10485158GA SED10485159GA SED10485160GA SED10485161GA SED10485162GA SED10485164GA SED10485165GA SED10485166GA SED10485167GA SED10495171GA SED10495172GA SED10495173GA SED10495174GA SED10495175GA SED10495176GA SED10495177GA SED10495178GA SED10495179GA SED10495180GA SED10495181GA SED10495182GA SED10495183GA SED10495184GA SED10495185GA SED10495187GA SED10495190GA SED10495191GA SED10495192GA SED10495193GA SED10495194GA SED10495195GA SED10495196GA SED10495197GA SED10495198GA SED10495199GA SED10495201GA SED10495203GA SED10495204GA SED10495209GA SED10495211GA SED10495212GA SED10495213GA SED10495214GA SED10495215GA SED10495216GA SED10495217GA SED10495218GA SED10495219GA SED10495220GA SED10495221GA SED10495222GA SED10495223GA SED10495225GA SED10495226GA SED10495227GA SED10495228GA SED10495229GA SED10495230GA SED10495231GA SED10495232GA SED10495233GA SED10495234GA SED10495235GA SED10495236GA SED10495237GA SED10495238GA SED10495239GA SED10495240GA SED10495241GA SED10495242GA SED10495243GA SED10495244GA SED10495245GA SED10495246GA SED10495247GA SED10495248GA SED10495249GA SED10495250GA SED10495251GA SED10495252GA SED10495253GA SED10495254GA SED10495255GA SED10495256GA SED10495257GA SED10495258GA SED10495259GA SED10495260GA SED10495261GA SED10495264GA SED10495265GA SED10495266GA SED10495267GA SED10495268GA SED10495269GA SED10495270GA SED10495271GA SED10495272GA SED10495273GA SED10495274GA SED10495277GA SED10495278GA SED10495279GA SED10495282GA SED10495284GA SED10495286GA SED10495287GA SED10495288GA SED10495289GA SED10495290GA SED10495291GA SED10495292GA SED10495294GA SED10495296GA SED10495297GA SED10495298GA SED10495301GA SED10495305GA SED10495307GA SED10495310GA SED10495317GA SED10495340GA SED10505348GA SED10505351GA SED10505353GA SED10505354GA SED10505355GA SED10505357GA SED10505380GA SED10505386GA SED10505391GA SED10505395GA SED10505396GA SED10505397GA SED10505402GA SED10505403GA SED10505429GA SED10505464GA SED10505465GA SED10505466GA SED11046864GA SED11046869GA SED09451330GA
SED09451331GA SED09451336GA SED09451338GA SED10353745GA SED10353746GA SED10353747GA SED10353748GA SED10353749GA SED10353750GA SED10353751GA SED10353752GA SED10132470GA SED10132475GA SED10132478GA SED10132484GA SED10132486GA SED10132487GA SED10132488GA SED10132492GA SED10132493GA SED10132495GA SED10132499GA SED10132503GA SED10132509GA SED10132512GA SED10132514GA SED10132517GA SED10142533GA SED10142544GA SED10142546GA SED10142547GA SED10142548GA SED10142549GA SED10142550GA SED10142556GA SED10142558GA SED10142560GA SED10142564GA SED10142565GA SED10142566GA SED10142567GA SED10142568GA SED10142572GA SED10142573GA SED10142574GA SED10142575GA SED10142577GA SED10142588GA SED10142589GA SED10142591GA SED10142592GA SED10142593GA SED10142594GA SED10142595GA SED10142596GA SED09400931GA SED10262944GA SED09340664GA SED09451333GA SED10536466GA SED11056913GA SED10536399GA SED10536435GA SED10536436GA SED10536453GA SED10536454GA SED10536469GA SED10536471GA SED10536472GA SED10536479GA SED10536487GA SED10536500GA SED10536512GA SED11046736GA SED11046750GA SED11046764GA SED11046766GA SED11046767GA SED11046769GA SED11046770GA SED11046773GA SED11046776GA SED11046778GA SED11046779GA SED11046785GA SED11046786GA SED11046787GA SED11046788GA SED11046789GA SED11046792GA SED11046793GA SED11046794GA SED11046795GA SED11046796GA SED11046797GA SED11046798GA SED11046799GA SED11046800GA SED11046803GA SED11046805GA SED11046806GA SED11046807GA SED11046808GA SED11046809GA SED11046810GA SED11046813GA SED11046814GA SED11046816GA SED11046842GA SED11046848GA SED11046854GA SED11046856GA SED11046868GA SED11046875GA SED11046876GA SED11046877GA SED11046878GA SED11046879GA SED11046884GA SED11046885GA SED11046886GA SED11046887GA SED11056888GA SED11056902GA SED10333637GA SED10333638GA SED10262945GA SED09280425GA SED09280433GA SED09280435GA SED10495283GA SED10495285GA SED10505722GA SED10516085GA SED10516086GA SED11036577GA SED11036583GA SED11036584GA SED11036599GA SED11046855GA SED11056889GA SED11056895GA SED11056896GA SED11056897GA SED11056899GA SED11056903GA SED11056904GA SED11056905GA SED11056906GA SED11056907GA SED11056908GA SED11056909GA SED11056911GA SED10536490GA SED11097479GA SED11097480GA SED11107727GA SED11107729GA SED11107731GA SED11107732GA SED11107733GA SED11118009GA SED09280427GA SED09280429GA SED09310463GA SED09340606GA SED09340643GA SED09340652GA SED09340656GA SED09410955GA SED09410959GA SED09410960GA SED09410961GA SED09410962GA SED09410963GA SED09410964GA SED09410965GA SED09410966GA SED09410968GA SED09410969GA SED09410970GA SED09410971GA SED09410972GA SED09410973GA SED09410974GA SED09410976GA SED09410979GA SED10192722GA SED09431041GA SED09471471GA SED09471475GA SED09471492GA SED09471495GA SED09471498GA SED10021694GA SED10021712GA SED10021719GA SED10021728GA SED10021729GA SED10021731GA SED10021734GA SED10041812GA SED10041820GA SED10061876GA SED10061877GA SED10061885GA SED10061887GA SED10061888GA SED10061892GA SED10092045GA SED10102052GA SED10102056GA SED10112252GA SED10112253GA SED10112254GA SED10112255GA SED10112256GA SED10112262GA SED10112263GA SED10112265GA SED10112269GA SED10112272GA SED10112273GA SED10112274GA SED10112275GA SED10112276GA SED10112277GA SED10112278GA SED10112280GA SED10122282GA SED10122284GA SED10122290GA SED10122292GA SED10122295GA SED10122296GA SED10122305GA SED10122307GA SED10122309GA SED10122311GA SED10122313GA SED10122315GA SED10122317GA SED10122319GA SED10122333GA SED10122338GA SED10122339GA SED10122343GA SED10122344GA SED10122346GA SED10122350GA SED10414353GA SED10414354GA SED10414355GA SED10424357GA SED10424373GA SED10424374GA SED10424375GA SED10424376GA SED10424377GA SED10424378GA SED10424379GA SED10424380GA SED10424381GA SED10424382GA SED10424383GA SED10424384GA SED10424387GA SED09340627GA SED09340631GA SED09340650GA SED09340657GA SED09340658GA SED09420982GA SED09420983GA SED09420984GA SED09420991GA SED09420996GA SED09420999GA SED09421000GA SED09421002GA SED09431066GA SED09431069GA SED09441179GA SED09451357GA SED09451360GA SED09451362GA SED09451363GA SED09491611GA SED09491612GA SED09491616GA SED09491618GA SED09491622GA SED09491623GA SED09491625GA SED09491627GA SED09491629GA SED09491630GA SED09491631GA SED09491633GA SED09491637GA SED09511641GA SED09511651GA SED09400928GA SED10061875GA SED10444726GA SED10505639GA SED10505711GA SED10515913GA SED10516054GA SED09400918GA SED09441260GA SED09451364GA SED09461395GA SED09461417GA SED09461418GA SED09471420GA SED09471421GA SED09471423GA SED09471426GA SED09471430GA SED09471435GA SED09471443GA SED09471444GA SED09471445GA SED09471446GA SED09471447GA SED10516066GA SED10516069GA SED11046874GA SED11056890GA SED11056891GA SED11056892GA SED11056893GA SED11056898GA SED11056900GA SED11056901GA SED10495202GA SED10516084GA SED10516100GA SED09441155GA SED09441156GA SED09441163GA SED09441164GA SED09441166GA SED09441167GA SED09441169GA SED09441170GA SED09441171GA SED09441174GA SED09441176GA SED09441184GA SED09441185GA SED09441187GA SED09441188GA SED09441190GA SED09441191GA SED09491521GA SED09491525GA SED09491539GA SED09491566GA SED09491568GA SED09491570GA SED09491571GA SED09491573GA SED09491574GA SED10505479GA SED10515756GA SED10515768GA SED10515815GA SED10515881GA SED10485127GA SED10132413GA SED10132414GA SED10132415GA SED10132418GA SED10132421GA SED10132422GA SED10132423GA SED10132424GA SED10132425GA SED10132426GA SED10132429GA SED10132430GA SED10132432GA SED10132433GA SED10132434GA SED10132435GA SED10132436GA SED10132438GA SED10132439GA SED10132440GA SED10132477GA SED10495335GA SED10505379GA SED10505411GA SED10505413GA SED09280439GA SED09310485GA SED09310506GA SED09330579GA SED09330583GA SED09330599GA SED09330600GA SED09330602GA SED09330603GA SED09330604GA SED09340607GA SED09340625GA SED09340626GA SED09350666GA SED09350667GA SED09350674GA SED09350675GA SED09350693GA SED09350694GA SED09350696GA SED09350699GA SED09350700GA SED09350715GA SED09350717GA SED09350721GA SED09370732GA SED09370733GA SED09370740GA SED09400929GA SED09400930GA SED09400932GA SED09400939GA SED09400950GA SED10152648GA SED10152653GA SED10162669GA SED10162682GA SED10162687GA SED10162689GA SED10122359GA SED10122360GA SED10122367GA SED10122369GA SED09441256GA SED09441258GA SED09441287GA SED09441292GA SED09451308GA SED09451320GA SED09451325GA SED09451327GA SED09451334GA SED09421013GA SED09421017GA SED09441269GA SED09441289GA SED09451301GA SED09451307GA SED09451328GA SED09451329GA SED09451332GA SED09451335GA SED09451337GA SED10212759GA SED10222833GA SED10222837GA SED10222838GA SED10222839GA SED10222840GA SED10222841GA SED10222842GA SED10132471GA SED10132505GA SED10132507GA SED10132508GA SED10132511GA SED10132513GA SED10132515GA SED10132518GA SED10142530GA SED10142534GA SED10142535GA SED10142538GA SED10333618GA SED10333619GA SED10333620GA SED10333622GA SED10333627GA SED10333629GA SED10333630GA SED10333631GA SED09370743GA SED09370823GA SED09370831GA SED09390848GA SED09400938GA SED09441262GA SED09441266GA SED09441271GA SED09441276GA SED09441278GA SED09441279GA SED09441284GA SED09441290GA SED09441293GA SED09451299GA SED09451306GA SED09451321GA SED09451343GA SED09451376GA SED10061880GA SED10061882GA SED10082010GA SED10092039GA SED10092046GA SED10092048GA SED10112259GA SED10122281GA SED10122285GA SED10122289GA SED10122293GA SED10122294GA SED10122310GA SED10122312GA SED10122316GA SED10122321GA SED10122323GA SED10122324GA SED10122327GA SED10122328GA SED10122330GA SED10122340GA SED10122341GA SED10122342GA SED10122347GA SED10122348GA SED10122354GA SED10122355GA SED10122357GA SED10122358GA SED10122364GA SED10122368GA SED10122370GA SED10122374GA SED10122378GA SED10122379GA SED10252905GA SED10252906GA SED10262948GA SED10262952GA SED10262954GA SED10262960GA SED10262965GA SED10262972GA SED10262973GA SED10262976GA SED10262979GA SED10262980GA SED10262982GA SED10262983GA SED10262984GA SED10262985GA SED10262986GA SED10262987GA SED10262988GA SED10262989GA SED10262990GA SED10182709GA SED10182712GA SED10182716GA SED10182717GA SED10182718GA SED10182719GA SED10182720GA SED10182721GA SED10192733GA SED10192734GA SED09441270GA SED09441281GA SED09441283GA SED09471434GA SED10021685GA SED10021686GA SED10021689GA SED10021693GA SED10021695GA SED09310460GA SED09310465GA SED09340612GA SED09340613GA SED09340615GA SED09340624GA SED09340630GA SED09340633GA SED09340634GA SED09340636GA SED09340639GA SED09350716GA SED09370724GA SED09370728GA SED09370731GA SED09370735GA SED09370737GA SED09370745GA SED09370751GA SED09370794GA SED09370796GA SED09370797GA SED09370800GA SED09370801GA SED09471424GA SED09471427GA SED09471428GA SED09471436GA SED09471438GA SED09471439GA SED09471485GA SED09471497GA SED09491586GA SED09511672GA SED10102095GA SED10102124GA SED10102125GA SED10102131GA SED10102134GA SED10102138GA SED10112140GA SED10112144GA SED10112145GA SED10112149GA SED10112153GA SED10112159GA SED10112163GA SED10112167GA SED10112170GA SED10112181GA SED10112183GA SED10112185GA SED10112189GA SED10112260GA SED10303425GA SED10303439GA SED09441263GA SED09441285GA SED09471419GA SED10132408GA SED10132410GA SED10132412GA SED10132428GA SED10373970GA SED10373971GA SED10373972GA SED10373973GA SED10373974GA SED10373975GA SED10373977GA SED10373980GA SED10373981GA SED10373983GA SED10373988GA SED10374003GA SED10384049GA SED10384054GA SED10384058GA SED10384059GA SED10384085GA SED10384099GA SED10384100GA SED10384102GA SED10384103GA SED10384107GA SED10384108GA SED10384109GA SED10384111GA SED10384112GA SED10384115GA SED10384117GA SED10384118GA SED10384119GA SED10384120GA SED10384121GA SED10384122GA SED10384123GA SED10384124GA SED10384125GA SED10384126GA SED10384127GA SED10384128GA SED10384129GA SED10384130GA SED10384131GA SED10384132GA SED10384133GA SED10384134GA SED10384141GA SED10384142GA SED10384143GA SED10384144GA SED10384147GA SED10384148GA SED10384149GA SED10384152GA SED10384153GA SED10384154GA SED10384155GA SED10384156GA SED10424427GA SED10424431GA SED10424432GA SED10424433GA SED10424434GA SED10424435GA SED10424438GA SED10424439GA SED10424440GA SED10424441GA SED10424442GA SED10424443GA SED10424444GA SED10424445GA SED10424446GA SED10424447GA SED10424448GA SED10424449GA SED10424450GA SED10424451GA SED10424452GA SED10424453GA SED10424465GA SED10424466GA SED10424467GA SED10424468GA SED10424469GA SED10424470GA SED10485133GA SED10485134GA SED10495300GA SED10495336GA SED10495337GA SED10495338GA SED10495339GA SED10505358GA SED10505370GA SED10505375GA SED10505405GA SED10505406GA SED10505407GA SED10505419GA SED10505443GA SED10505512GA SED10505544GA SED10505578GA SED10505612GA SED10505615GA SED10505617GA SED10505625GA SED10505626GA SED10505655GA SED10505660GA SED10505661GA SED10505676GA SED10505683GA SED10505694GA SED10505695GA SED10505707GA SED10505710GA SED10505714GA SED10505715GA SED10505716GA SED10505717GA SED10505719GA SED10505721GA SED10515726GA SED10515731GA SED10515735GA SED10515737GA SED10526280GA SED10526286GA SED11026570GA SED11036588GA SED11036589GA SED11036609GA SED11036610GA SED11036611GA SED11036618GA SED11036619GA SED11036620GA SED11036621GA SED11036622GA SED11036627GA SED11036658GA SED11036660GA SED11036662GA SED11036668GA SED11036675GA SED11036686GA SED11036688GA SED11036706GA SED09310459GA SED09310519GA SED09330569GA SED09330601GA SED09340608GA SED09350691GA
SED09350695GA SED09350718GA SED09370722GA SED09370734GA SED09370795GA SED09400908GA SED09400924GA SED09400952GA SED09461415GA SED09471429GA SED09471437GA SED09280418GA SED09280434GA SED10162676GA SED10182713GA SED10182714GA SED10182715GA SED10192735GA SED10333639GA SED10384053GA SED10384060GA SED10384061GA SED10384064GA SED10384086GA SED10384087GA SED10384088GA SED10384092GA SED10384093GA SED10394174GA SED10394178GA SED10394179GA SED10394181GA SED10394182GA SED10404186GA SED10404187GA SED10404188GA SED10404189GA SED10404191GA SED10404192GA SED10404194GA SED10404195GA SED10404212GA SED10404213GA SED10404214GA SED10404217GA SED10404218GA SED10404219GA SED10404223GA SED10404224GA SED10414314GA SED10414315GA SED10485067GA SED10485068GA SED10485071GA SED10485072GA SED10485073GA SED10485074GA SED10485075GA SED10485076GA SED10485077GA SED10485078GA SED10485079GA SED10485081GA SED10485083GA SED10495325GA SED10495329GA SED10495330GA SED10495343GA SED10495344GA SED10495345GA SED10495346GA SED10505350GA SED10505366GA SED10505367GA SED10505400GA SED10505401GA SED10505428GA SED10505558GA SED10505572GA SED10505584GA SED10505591GA SED10505618GA SED10505624GA SED10505629GA SED10505630GA SED10505631GA SED10505653GA SED10515807GA SED10515809GA SED10515810GA SED10515811GA SED10515812GA SED10515856GA SED10515857GA SED10515858GA SED10515874GA SED10515877GA SED10536393GA SED10536394GA SED10536429GA SED10536431GA SED10536434GA SED10536437GA SED10536439GA SED10536440GA SED10536442GA SED10536443GA SED10536444GA SED10536445GA SED10536447GA SED10536448GA SED10536449GA SED10536450GA SED10536455GA SED10536457GA SED10536459GA SED10536464GA SED10536470GA SED10536473GA SED10536474GA SED10536475GA SED10536476GA SED10536477GA SED10536480GA SED10536481GA SED10536482GA SED10536484GA SED10536486GA SED10536488GA SED10536491GA SED10536492GA SED10536494GA SED10536495GA SED10536497GA SED10536498GA SED10536499GA SED10536501GA SED10536503GA SED10536505GA SED10536517GA SED10536519GA SED10536520GA SED10536521GA SED10536527GA SED11026534GA SED11026536GA SED11026544GA SED11026546GA SED11026549GA SED11026550GA SED11026559GA SED11026562GA SED11026563GA SED11026564GA SED11026567GA SED11026568GA SED11036585GA SED11036586GA SED11036587GA SED11046722GA SED11046734GA SED11046740GA SED11046753GA SED11046760GA SED11046761GA SED11046762GA SED11056940GA SED11056947GA SED11056948GA SED11056949GA SED11057064GA SED11057067GA SED11067080GA SED11067081GA SED11067082GA SED11067085GA SED11067086GA SED11067087GA SED11067101GA SED11067102GA SED11067104GA SED11067105GA SED11067106GA SED11067107GA SED11067108GA SED11067113GA SED11077151GA SED11077171GA SED09441274GA SED09461413GA SED09471422GA SED09471425GA SED10323589GA SED10333590GA SED10333623GA SED10333624GA SED10343714GA SED10353754GA SED10353755GA SED10353756GA SED10353757GA SED10353758GA SED10353759GA SED10353760GA SED10353761GA SED09421015GA SED09421016GA SED09421021GA SED09431023GA SED09431037GA SED09431045GA SED09431046GA SED09431047GA SED09431048GA SED09431049GA SED09431050GA SED09431051GA SED09431052GA SED09431053GA SED09431062GA SED09431022GA SED09431024GA SED09431027GA SED09431028GA SED09431029GA SED09431031GA SED10132437GA SED09370818GA SED10293137GA SED10293143GA SED09340642GA SED09340660GA SED09340665GA SED09330605GA SED09340614GA SED09340623GA SED09340632GA SED09340635GA SED09340638GA SED09340641GA SED09340644GA SED10414282GA SED08400200GP SED09180316GP SED09180317GP SED09180319GP SED09180328GP SED09180331GP SED09180343GP SED10112191GA SED10112212GA SED10333591GA SED10333592GA SED10333593GA SED10333594GA SED10333596GA SED10333597GA SED10363842GA SED10363860GA SED10363864GA SED10363865GA SED10363866GA SED10363867GA SED10363872GA SED10363873GA SED10363885GA SED10363887GA SED10363892GA SED10363896GA SED10363899GA SED10363900GA SED10363902GA SED10363905GA SED10363906GA SED10363907GA SED10363909GA SED10363913GA SED10373916GA SED10404221GA SED10404273GA SED10404274GA SED10404275GA SED10414277GA SED10414278GA SED10414279GA SED09180315GP SED09180318GP SED09180320GP SED09180321GP SED09180322GP SED09180323GP SED09180324GP SED09180329GP SED09180330GP SED09180332GP SED09180335GP SED09180336GP SED09180340GP SED09180342GP SED10384096GA SED10414313GA SED10424428GA SED10424429GA SED10424430GA SED10424436GA SED10424437GA SED10485080GA SED10485084GA SED10485086GA SED10485088GA SED10485089GA SED10485090GA SED10485091GA SED10485092GA SED10485093GA SED10485094GA SED10485095GA SED10485096GA SED10485097GA SED10485098GA SED10485099GA SED10485100GA SED10485101GA SED10485102GA SED10485103GA SED10485104GA SED10485105GA SED10485106GA SED10485108GA SED10485112GA SED10485121GA SED10485138GA SED10485139GA SED10485140GA SED10485141GA SED10495186GA SED10495200GA SED10495206GA SED10495207GA SED10495208GA SED10495210GA SED10495263GA SED10495276GA SED10495341GA SED10495342GA SED10505384GA SED10505409GA SED10505410GA SED10505414GA SED10505486GA SED10505487GA SED10505634GA SED10515750GA SED10515751GA SED10515753GA SED10515757GA SED10515758GA SED10515759GA SED10515760GA SED10515762GA SED10515763GA SED10515766GA SED10515775GA SED10515780GA SED10515787GA SED10515788GA SED10515789GA SED10515800GA SED10515801GA SED10515802GA SED10515803GA SED10515804GA SED10515805GA SED10515976GA SED10515999GA SED10516026GA SED10516027GA SED10516029GA SED10516034GA SED10516035GA SED10516036GA SED10516038GA SED10516039GA SED10516040GA SED10516041GA SED10516048GA SED10516050GA SED10516053GA SED10516055GA SED10516060GA SED10516061GA SED10516062GA SED10516063GA SED10516065GA SED10516067GA SED10516071GA SED10516072GA SED10516080GA SED10516081GA SED10516087GA SED10516088GA SED10516090GA SED10516105GA SED10516106GA SED10516107GA SED10526108GA SED10526111GA SED10526113GA SED10526129GA SED10526135GA SED10526138GA SED10526139GA SED10526147GA SED10526148GA SED10526149GA SED10526150GA SED10526151GA SED10526152GA SED10526153GA SED10526154GA SED10526155GA SED10526156GA SED10526157GA SED10526161GA SED10536496GA SED11026538GA SED11026547GA SED11026556GA SED11026557GA SED11026560GA SED11026565GA SED11036595GA SED11036624GA SED11036625GA SED11036626GA SED11036628GA SED11036629GA SED11036630GA SED11036631GA SED11036632GA SED11036647GA SED11036648GA SED11036649GA SED11036650GA SED11036651GA SED11036652GA SED11036654GA SED11046812GA SED11046872GA SED11056961GA SED11057065GA SED11067045GA SED11067049GA SED11067096GA SED11067115GA SED11067118GA SED11067119GA SED11067120GA SED11067121GA SED11067128GA SED11077240GA SED11077250GA SED10404196GA SED10404197GA SED10404198GA SED10404199GA SED10404200GA SED10404201GA SED10404202GA SED10404205GA SED10404208GA SED10404211GA SED10404258GA SED10404259GA SED10404260GA SED10404261GA SED10404262GA SED10404263GA SED10454767GA SED10454768GA SED10454769GA SED10454771GA SED10454773GA SED10454774GA SED10454775GA SED10454776GA SED10454777GA SED10454778GA SED10454779GA SED10454792GA SED10454793GA SED10454794GA SED10454795GA SED10454796GA SED10454797GA SED10454798GA SED10454799GA SED10454800GA SED10454801GA SED10454802GA SED10041832GA SED10041833GA SED10071940GA SED09511664GA SED09511665GA SED10152599GA SED10152604GA SED10152651GA SED10262994GA SED10262996GA SED10262997GA SED10262998GA SED10262999GA SED09451300GA SED09451309GA SED09451324GA SED10102051GA SED10122353GA SED10142543GA SED10152612GA SED10273042GA SED10283094GA SED10283095GA SED10293125GA SED10293126GA SED10293127GA SED10293128GA SED10293129GA SED10293130GA SED10293131GA SED10293134GA SED10293139GA SED10273033GA SED10273034GA SED10273041GA SED10273054GA SED10273056GA SED10273060GA SED10283088GA SED10283089GA SED10283090GA SED10283091GA SED10283092GA SED10283093GA SED10293096GA SED10293097GA SED10293098GA SED10293099GA SED10293100GA SED10293101GA SED10293102GA SED10293103GA SED10293104GA SED10293105GA SED10293106GA SED10293107GA SED10293108GA SED10293109GA SED10293110GA SED10293111GA SED10293112GA SED10293113GA SED10293114GA SED10293115GA SED10303423GA SED10313496GA SED10313499GA SED10313501GA SED10313502GA SED10363889GA SED10313497GA SED10313498GA SED10414316GA SED10414317GA SED10122283GA SED10424490GA SED10021698GA SED10021716GA SED10021725GA SED10041815GA SED10061881GA SED10515820GA SED10515855GA SED10515859GA SED10515861GA SED10515882GA SED10515897GA SED10515902GA SED10515903GA SED10515921GA SED10515923GA SED10515925GA SED10515926GA SED10515927GA SED09340651GA SED09400927GA SED09400934GA SED10071900GA SED10102058GA SED10112257GA SED10112268GA SED10112271GA SED10122302GA SED10122303GA SED09451310GA SED09471449GA SED09471454GA SED09471455GA SED09491530GA SED09491531GA SED09491532GA SED09491620GA SED10132474GA SED10142540GA SED10505412GA SED10515755GA SED10515761GA SED10515764GA SED10515765GA SED10515772GA SED10515773GA SED10515774GA SED10515776GA SED10515777GA SED10515778GA SED10515779GA SED10515781GA SED10515782GA SED10515784GA SED10515785GA SED10515786GA SED10515790GA SED10515791GA SED10515792GA SED10515793GA SED10515794GA SED10515795GA SED10515796GA SED10515797GA SED10515798GA SED10515799GA SED10515806GA SED10515827GA SED10515839GA SED10515846GA SED10515847GA SED10515849GA SED10515850GA SED10515883GA SED10515966GA SED10515977GA SED10516001GA SED10516003GA SED10516010GA SED10516012GA SED10516013GA SED10516014GA SED10516015GA SED10516016GA SED10516017GA SED10516018GA SED10516021GA SED10516022GA SED10516023GA SED10516024GA SED10516042GA SED10516043GA SED10516044GA SED10516045GA SED10454738GA SED10454739GA SED10526168GA SED10526175GA SED10526176GA SED10526177GA SED10526178GA SED10526182GA SED10526183GA SED10526189GA SED10526194GA SED10526196GA SED10526215GA SED10526232GA SED10353744GA SED09340649GA SED09431073GA SED09441079GA SED09441214GA SED09451365GA SED10102109GA SED10102111GA SED10102113GA SED10102117GA SED10102119GA SED10333610GA SED10333611GA SED10333612GA SED10333613GA SED10333614GA SED10333615GA SED10333616GA SED10384055GA SED10384056GA SED10384057GA SED10384104GA SED10384105GA SED10384106GA SED10424385GA SED10424386GA SED10424388GA SED10424389GA SED10424390GA SED10424391GA SED10424392GA SED10424393GA SED10424394GA SED10424395GA SED10424396GA SED10424485GA SED10424506GA SED10434593GA SED10434629GA SED10515987GA SED10516073GA SED10516096GA SED10516099GA SED10526127GA SED10526131GA SED10526132GA SED10526163GA SED10526164GA SED10615816GA SEG10477461HX0 SED10363875GA SED10363897GA SED10363908GA SED10363912GA SED10373919GA SED11067084GA SED11067088GA SED11067090GA SED11067091GA SED11067092GA SED11067094GA SED11067095GA SED11067098GA SED11067109GA SED11077148GA SED11077149GA SED11077157GA SED11077168GA SED11077178GA SED11077208GA SED11077214GA SED11077216GA SED11077217GA SED010464924GA SED010485111GA SED010485115GA SED010485117GA SED010485118GA SED010485119GA SED010485122GA SED010485123GA SED010485124GA SED010485125GA SED010485126GA SED010485130GA SED010485131GA SED10516097GA SED10526145GA SED10293142GA SED10293178GA SED10293188GA SED10293189GA SED10293190GA SED10293193GA SCD09336481PA SED10516091GA SED10516092GA SED10516093GA SED10516094GA SED10516095GA SED10526121GA SED10526122GA SED10526123GA SED10526124GA SED10526126GA SED10526128GA SED10526137GA SED10526140GA SED10323580GA SED10323581GA SED10323582GA SED10323583GA SED10323584GA SED10323585GA SED10323586GA SED10323587GA SED10212789GA SED09280414GA SED10454740GA SED10515816GA SED11036597GA SED11036598GA SED11036600GA SED11036601GA SED11036602GA SED11036613GA SED11046860GA SED11046873GA SED11056894GA SED10212786GA SED10212787GA SED10212788GA SED10252912GA SED10262949GA SED10262950GA SED10262958GA SED10262963GA SED10262964GA SED10262966GA SED10262969GA SED10424503GA SED10464924GA SED10485111GA SED10485115GA SED10485117GA SED10485118GA SED10485119GA SED10485122GA SED10485123GA SED10485124GA SED10485125GA SED10485126GA SED10485128GA SED10485129GA SED10485130GA SED10485131GA SED10485132GA SED10485135GA SED10485136GA SED10485137GA SED10505408GA SED10505467GA SED10515745GA SED10515819GA SED10515822GA SED10515823GA SED10515830GA SED10515835GA SED10515854GA SED10515862GA SED10515863GA SED10515864GA SED10515865GA SED10515890GA SED10515891GA SED10515892GA SED10515893GA SED10515895GA SED10515896GA SED10515899GA SED10515900GA SED10515901GA SED10515904GA SED10515906GA SED10515908GA SED10515909GA SED10515910GA SED10515911GA SED10515914GA SED10515916GA SED10515918GA SED10515920GA SED10515971GA SED10515973GA SED10515974GA SED10515978GA SED10516000GA SED10516004GA SED10516064GA SED10516074GA SED10516075GA SED10516076GA SED10516077GA SED10516078GA SED10516079GA SED10516082GA SED10526188GA SED10526195GA SED10526205GA SED10526206GA SED10526230GA SED11036623GA SED11046723GA SED11046751GA SED11046772GA SED11046801GA SED11046802GA SED11046804GA SED11046811GA SED11046815GA SED11046843GA SED11046844GA SED11046846GA SED11046847GA SED11046849GA SED11046861GA SED11046862GA SED11046863GA SED11046866GA SED11046867GA SED11056910GA SED11056914GA SED11056937GA SED11056938GA SED11056946GA SED11056955GA SED11097476GA SED10041835GA SED10082001GA SED10082002GA SED10102100GA SED10102101GA SED10102104GA SED10142531GA SED10192723GA SED10192724GA SED10192725GA SED10192726GA SED10192727GA SED10192728GA SED10192729GA SED10192730GA SED10192731GA SED10192732GA SED10132501GA SED10142562GA SED10142569GA SED10142570GA SED10222868GA SED10363869GA SED103638869GA SED10394180GA SED10526144GA SED010485135GA SED010485137GA SED010485136GA SED10424507GA SED10516011GA SED10536507GA SED10536518GA SED10536524GA SED10536528GA SED10536529GA SED11026541GA SED11026543GA SED11026551GA SED11026561GA SED10102080GA SED09340659GA SED09350668GA SED09350669GA SED09350670GA SED09350671GA SED09350672GA SED09350673GA SED09350676GA SED09350677GA SED09350690GA SED09350692GA SED09350697GA SED09350698GA SED09350701GA SED09350712GA SED09350713GA SED09350714GA SED09350719GA SED09350720GA SED09370736GA SED0937073GA SED09420988GA SED09471451GA SED09471453GA SED09471458GA SED09471482GA SED09471489GA SED09491580GA SED09491581GA SED09491588GA SED09491591GA SED09511670GA SED09511671GA SED09511673GA SED10122351GA SED10132490GA SED10132491GA SED10132502GA SED10132504GA SED10142597GA SED10152646GA SED10152656GA SED10152657GA SED10152663GA SED10182706GA SED10192736GA SED10252892GA SED10293135GA SED10424489GA SED11036617GA SED11036656GA SED11036657GA SED11036672GA SED11036673GA SED11036674GA SED10112178GA SED10112192GA SED10112201GA SED10112203GA SED10112204GA SED10112205GA SED10112211GA SED10112213GA SED10112226GA SED10112228GA SED10122286GA SED10122337GA SED10404215GA SED10404216GA SED10404222GA SED10404225GA SED10404226GA SED09491632GA SED09511648GA SED09511659GA SED09421018GA SED09421020GA SED09431026GA SED09431030GA SED09431032GA SED09431035GA SED09451304GA SED09451305GA SED09441196GA SED09511658GA SED09511660GA SED09511661GA SED09511662GA SED09511663GA SED09511666GA SED09511668GA SED09370742GA SED09370807GA SED09370808GA SED09370810GA SED09370816GA SED09390867GA SED09390872GA SED09390892GA SED09400906GA SED09491606GA SED09511669GA SED09511677GA SED09511678GA SED10021690GA SED10021692GA SED10182701GA SED10222832GA SED10252894GA SED10252897GA SED10252925GA SED10142542GA SED10152641GA SED10152644GA SED10152649GA SED10152652GA SED10152659GA SED10152661GA SED10152664GA SED10515814GA SED10515832GA SED10515852GA SED10515875GA SED10515889GA SED10132467GA SED10132468GA SED10142525GA SED10142532GA SED10142536GA SED10142537GA SED10303441GA SED10303444GA SED10303446GA SED10303447GA SED10303450GA SED10303453GA SED10303455GA SED10313459GA SED09280407GA SED09340611GA SED09340616GA SED09340618GA SED09340619GA SED09340620GA SED09340622GA SED09340646GA SED09410957GA SED09420989GA SED09491593GA SED09491604GA SED09491605GA SED09491607GA SED09491626GA SED09491634GA SED09491635GA SED09491636GA SED09491638GA SED09511642GA SED09511643GA SED09511644GA SED09511645GA SED09511649GA SED09511650GA SED09511653GA SED09511655GA SED09511656GA SED09511657GA SED10041859GA SED10041862GA SED10061889GA SED10102062GA SED10102079GA SED10132444GA SED10132454GA SED10132459GA SED10464945GA SED10464949GA SED10464959GA SED10464960GA SED10464961GA SED10464962GA SED10464963GA SED10464964GA SED10464965GA SED10464966GA SED10464967GA SED10464968GA SED10464969GA SED10464971GA SED10474973GA SED10474974GA SED10474975GA SED10474976GA SED10474977GA SED10474978GA SED10474979GA SED10474980GA SED10474981GA SED10474982GA SED10474983GA SED10474984GA SED10474985GA SED10474986GA SED10474987GA SED10474988GA SED10474989GA SED10474990GA SED10474991GA SED10474992GA SED10474993GA SED10474994GA SED10474995GA SED10474996GA SED10474997GA SED10474998GA SED10474999GA SED10475002GA SED10475003GA SED10475004GA SED10505447GA SED10505448GA SED10505449GA SED10505480GA SED10505481GA SED10505488GA SED10505498GA SED10505499GA SED10505510GA SED11087291GA SED11087301GA SED11087414GA SED10475005GA SED10475009GA SED10485114GA SED10485116GA SED10505371GA SED10505373GA SED10505389GA SED10505390GA SED10505416GA SED10505425GA SED10505431GA SED10505472GA SED10515783GA SED10515808GA SED10515825GA SED10515828GA SED10515831GA SED10515848GA SED10515851GA SED10515873GA SED11046791GA SED09340621GA SED09340629GA SED09340640GA SED09340654GA SED09340655GA SED10293123GA SED10293124GA SED09471473GA SED10021715GA SED10021720GA SED10041824GA SED10112150GA SED10112155GA SED10112156GA SED10112165GA SED10112173GA SED10112199GA SED10122287GA SED10122288GA SED10122291GA SED10122297GA SED10122314GA SED10122331GA SED10122332GA SED10122334GA SED10122335GA SED09090248GP SED09130299GP SED09130301GP SED09130304GP SED09130306GP SED09130308GP SED09130309GP SED09130310GP SED09130313GP SED10182702GA SED10182703GA SED10182704GA SED10182705GA SED10132473GA SED10303335GA SED10333640GA SED10353742GA SED10353743GA SED09400923GA SED09400942GA SED09400947GA SED09531681GA SED10021688GA SED10021691GA SED10041806GA SED10041817GA SED09491533GA SED09491541GA SED09491546GA SED09400944GA SED10142555GA SED10142581GA SED10142582GA SED10142587GA SED10152600GA SED10152603GA SED10152606GA SED10152607GA SED10152609GA SED10152610GA SED10152611GA SED10152613GA SED10152614GA SED10152618GA SED10152619GA SED10152621GA SED10152622GA SED10152623GA SED10152624GA SED10152625GA SED10152637GA SED10152650GA SED10162679GA SED10162688GA SED10303419GA SED10303420GA SED10303440GA SED10303443GA SED10303449GA SED10303451GA SED10303452GA SED10303454GA SED10313458GA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has become aware of a potential safety issue associated with the CARESCAPE Monitor B650 & B850 Cardiac Output measurement using Patient Data Module (PDM), and activation of ECG Leads Off alarm in specific cases.
1. When the CARESCAPE Monitor is used to measure Cardiac Output values with the Patient Data Module (PDM) or TRAM Module, Cardiac Output values may be calculated incorrectl
|
FDA Determined Cause 2 |
Software design |
Action |
The firm, GE Healthcare, sent "Urgent Medical Device Correction" letters dated March 23, 2011 to its consignees/customers. The letters were addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letters described the Safety Issues, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. The customers were instructed to follow the safety instructions documented in the letter. GE Healthcare is working on a software correction that will be provided to their customers once it is available and will contact the customer to schedule the software update.
If you have any questions regarding this medical device correction or the identification of affected items, please contact Technical Support at 1-800-558-7044, or your local Service Representative. |
Quantity in Commerce |
6902 |
Distribution |
AL, AZ, AR, CA, CO, CT, FL,GA, ID, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA,PR, SC, TN,TX, UT, VA, WA, WV, WI.
UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, THAILAND, SWITZERLAND, SWEDEN SLOVENIA, SLOVAKIA, SPAIN, SINGAPORE, SAUDI ARABIA, QATAR, PORTUGAL POLAND, NORWAY, NORTH KOREA, NEW ZEALAND, NETHERLANDS, MALTA, MALAYSIA, LITHUANIA, LEBANON, KUWAIT, ITALY, ISRAEL, IRELAND, INDIA, ICELAND, HONG KONG, GERMANY, FRANCE, FINLAND, DENMARK, ESTONIA, CZECH REPUBLIC, BELGIUM BAHRAIN, COLOMBIA, CHILE CANADA, BRAZIL, AUSTRIA, AUSTRALIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|