• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Accolade TMZF Plus Hip Stem

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Accolade TMZF Plus Hip Stemsee related information
Date Initiated by FirmMarch 01, 2011
Date PostedJune 13, 2011
Recall Status1 Terminated 3 on October 10, 2012
Recall NumberZ-2557-2011
Recall Event ID 58455
510(K)NumberK032300 K994366 
Product Classification Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate - Product Code ISL
ProductAccolade TMZF Plus Hip Stem: Catalog # 3020-0537 Accolate d TMZF Hip Stem # 5 Catalog #6020-3535 Accolade 132 Size 3.5 Catalog # 6021-0230 Accolade Plus TMZF Hip Stem #2 Catalog # 6021-0335 Accolade Plus TMZF Hip Stem #3 Catalog #3021-0435 Accolade TMZF Hip Stem #4 Catalog # 6021-0435 Accolade Plus TMZF Hip Stem #4 Catalog #6021-2530 Accolade (127 deg) Size 2.5 Accolade (127 deg) Size 2.5 Catalog # 6021-4535 Accolade (127 DEG) Size 4.5 K994366; k023105; K032300 Manufactured by " Stryker Ireland, Carrigtwohill Industrial Estate Carrington County Cork, Ireland The subject hip stem is a single-use device intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty.
Code Information Catalog @ 3020-0537 Accolated TMZF Hip Stem # 5 Lot #31390203 8 units CAtalog #6020-3535 Accolade 132 Size 3.5 Lot #31446602R 0 units  CAtalog # 6021-0230 Accolade Plus TMZF Hip Stem #2 Lot #31515703 10 units Catalog # 6021-0335 Accolade Plus TMZF Hipe Stem #3 Lot 31586402 10 units Catalog #3021-0435 Accolade TMZF Hipe STem #4 31597205 1 unit CAtalog # 6021-0435 Accolade Plus TMZF Hipe STem #4 Lot #31597206 4 units Catalog #6021-2530 Accolade (127 deg) Size 2.5 Accolade (127 deg) Size 2.5 Lot #31486505 10 units Catalog # 6021-4535 Accolade (127 DEG) Size 4.5 Lot #31850001 1 unit
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Drive
Mahwah NJ 07430
For Additional Information ContactColleen O'Meara
201-831-5970
Manufacturer Reason
for Recall
Accolade TMZF Plus Hip Stems are being recalled due to grit blast media which was observed in the drive hole.
FDA Determined
Cause 2
Employee error
ActionStryker Orthopaedics sent an Urgent Product Recall letter dated March 1, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to fax back the attached Product Recall Acknowledgement Form within 5 days to 201-831-6069. Customers were also instructed to: Contact the hospitals in their territory that have the affected product to arrange return of the product. Return the affected product via and Inter-Org Transfer to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey 07430. Attach the fluorescent orange PRODUCT REMEDIATION sticker to their return, indicating the Product Remediation # RA 2009-459. Mark the outer box with the words "Product Recall." For any questions call (201) 972-2100.
Quantity in Commercetotal of all units is 44
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = ISL
-
-