|
Class 2 Device Recall Brilliance 64 CT XRay System |
|
Date Initiated by Firm |
March 07, 2011 |
Date Posted |
May 19, 2011 |
Recall Status1 |
Terminated 3 on February 22, 2013 |
Recall Number |
Z-2206-2011 |
Recall Event ID |
58510 |
510(K)Number |
K033326
|
Product Classification |
Computed Tomography X-Ray System - Product Code JAK
|
Product |
Brilliance 64, Computed Tomography X-Ray System, Model Numbers: 728231 and 728326, including the Ingenuity CT. The recalled units are equipped with software versions: 3.5.21028 and 3.5.25028.
Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. |
Code Information |
The following Serial Numbers of the Brilliance 64 (including Ingenuity CT) are subject to recall/software correction: 4003, 9745, 9875, 90135, 90169, 95414, 300003, 300004, 300005, and 300010. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
For Additional Information Contact |
440-483-7000
|
Manufacturer Reason for Recall |
A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
|
FDA Determined Cause 2 |
Other |
Action |
A 'Field Safety Notice - Urgent - Medical Device Correction' notification was sent to the firm's customers via Certified mail on 3/8/2011. The notification letter explains the nature of the Metal Artifact Reduction/Metal Artifact Correction software error; the Brilliance CT X-Ray systems subject to correction and how to identify them; and the circumstances under which this error can occur. It also asks that the user stop running the MAR application on any of the affected Brilliance systems identified with the recalled serial numbers and software versions alluded to in the medical device correction notification. Lastly, the notification states that Philips Medical will provide a solution that will fix the issue and the notification lists contact names; telephone numbers; and E-mail addresses which the customer can contact for further information concerning this device issue, or customer support assistance. |
Quantity in Commerce |
10 units |
Distribution |
Nationwide Distribution -- DC, IL, IN, MI, MO, NY, OH, OR, PA, TX, and VT. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = PHILIPS MEDICAL SYSTEMS, INC.
|
|
|
|