Date Initiated by Firm November 12, 2013
Create Date January 06, 2014
Recall Status1 Terminated 3
Recall Number Z-0629-2014
Recall Event ID 66622
Product Fresenius 2008 Series Hemodialysis Machines:
Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T

The 2008H Hemodialysis machine is indicated for acute and chronic dialysis therapy. Hemodialysis is indicated for patients with acute or chronic renal failure, when conservative therapy is judged to be inadequate. The dialysate delivery machine is used for hemodialysis in hospital, dialysis centers, and at home. The 2008K, 20081(2 and 2008T hemodialysis machines are designed to perform hemodialysis in hospitals and dialysis clinics. They can be used for patients suffering chronic or acute renal failure. The 2008K, 2008K2 and 2008T are designed to provide hemodialysis treatment by controlling and monitoring both the dialysate and extracorporeal blood circuits. The Fresenius 2008K~home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K~home is also indicated for hemodialysis in the home and must be observed by a trained and qualified person as prescribed by their physician.
Code Information All serial numbers
Recalling Firm/
Manufacturer
Fresenius Medical Care Holdings, Inc.
920 Winter St
Waltham MA 02451-1521
For Additional Information Contact
800-662-1237
Manufacturer Reason
for Recall
2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
Action Fresenius Medical issued a Product Notification letter dated November 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who encounter a saline bag that is inappropriately filling, should discontinue the setup and remove the machine from service until the drain line and drain-associated hydraulic components can be checed by a qualified service technician. Customers were also instructed to ensure that the hemodialysis machine is installed and operated in compliance with the Machine Specification section of their operator's manual, including the specification located in the Water/Drain Section. Customers with questions were instructed to call 1-800-227-2572. For questions regarding this recall call 800-662-1237.
Quantity in Commerce 111,504 units US and 3,638 Canada and Mexico-
Distribution Worldwide Distribution - USA (nationwide) and Internationally Mexico and Canada.

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-