Recognition List Number: 025 Publication Date: 10/04/2010 

Part B: SUPPLEMENTARY INFORMATION

Recognition Number 11-210: ASTM F 543-07e1, Standard Specification and Test Methods for Metallic Medical Bone Screws. (Orthopaedics)

Date of Standard: 2007. 
Address of Standards Organization:
  ASTM International (ASTM)
  100 Barr Harbor Drive
  West Conshohocken, PA 19428-2959
 
CDRH Office and Division Associated with Recognized Standards:
OFFICE OF DEVICE EVALUATION (ODE)
Devices Affected:
Metallic bone screws (non-spinal applications)
Processes Affected:
510(K), PMA, IDE, PDP, HDE, Design Controls
Type of Standard:
 International, Vertical, Test Methods
Extent of Recognition:
Complete Standard

The specification standard provides requirements for materials, finish and marking, care and handling, and the acceptable dimensions and tolerances for metallic bone screws implanted into bone tissue. As the specification standard states, the dimensions and tolerances are applicable only to metallic bone screws described in the specification itself. Additional information must be provided to document that the design of the product will provide adequate mechanical properties for the particular application.
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
§888.3030 Plate, Fixation, Bone Class 2 HRS
 
Regulation
Number
Device Name Device
Class
Product
Code
§888.3040 Pin, Fixation, Threaded Class 2 JDW
 
Relevant Guidance:
Reviews Guidance Checklist for Orthopedic External Fixation Devices Version #5, 2/21/97
FDA Technical Contact:
  Ronald Jean
  FDA/CDRH/ODE
  10903 New Hampshire Avenue Building 66, Room 1414
  Silver Spring MD 20993
  301/796-5650
  Email: ronald.jean@fda.hhs.gov
-
-