Recognition List Number: 026 Publication Date: 03/18/2011
Part B: SUPPLEMENTARY INFORMATION
Recognition Number 9-66: AAMI / ANSI / ISO 8638:2010, Cardiovascular implants and Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and hemofilters. (ObGyn/Gastroenterology)
Date of Standard: 2010. |
| Addresses of Standards Organizations: |
| Association for the Advancement of Medical Instrumentation (AAMI) |
| 4301 North Fairfax Drive |
| Suite 301 |
| Arlington, VA 22203 |
| |
| American National Standards Institute (ANSI) |
| 25 West 43rd Street |
| 4th Floor |
| New York, NY 10036 |
| |
| International Organization for Standardization (ISO)* |
| 1, Rue de Varembe |
| Case Postale 56 |
| CH 1211 Geneva 20, 0 |
|
SWITZERLAND
|
|
| CDRH Office and Division Associated with Recognized Standards: |
| OFFICE OF DEVICE EVALUATION (ODE) | | |
|
|
| Devices Affected: |
| Hemodialyzer blood tubing, hemodialysis systems |
|
| Processes Affected: |
| 510(k), PMA, PDP, IDE, HDE, Quality System Regulations |
|
| Type of Standard: |
| Vertical, International |
| Extent of Recognition: |
ANSI/AAMI/ISO 8638:2010 Cardiovascular implants and artificial organs-Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and Hemofilters; With the following qualifications:
Additional testing not included in this standard may be required for marketing clearance or approval.
For 510(k) devices, testing and design requirements specified in Sections 4 and 5 should be applied in terms of comparing predicate and proposed devices and determining substantial equivalence.
For Section 4.2 Sterility, all fluid-contacting surfaces of the device, and the mating surfaces of all connectors integral to the device, should be sterile.
For Section 4.4.1 Structural integrity, additionally, device should be capable of enduring its maximum blood flow rate for a period of time representative of expected clinical use without tubing collapse or other signs of lack of performance.
For Section 4.4.9.1 Integral transducer protectors, the transducer protector should be capable of maintaining secure and leak-free connection to the hemodialysis machine when subjected to pressures of 2 times the manufacturer's recommended pressure for the device.
For Section 4.4.9.2 Non-integral transducer protectors, the transducer protector should be capable of maintaining secure and leak-free connection to the hemodialysis machine when subjected to pressures of 2 times the manufacturer's recommended pressure for the device.
For Section 4.4.11 Pump segment performance, performance characteristics of the pump segment should be evaluated at maximum specified flow rate. Also, endurance testing should be performed, looking for tubing collapse or other signs of lack of performance, for a period of time representative of expected clinical use.
For Section 5.2 Biological safety, testing shall be carried out in accordance with FDA Blue Book Memorandum #G95-1: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."
For Section 5.5.1.1 Positive-pressure test, additionally, endurance testing at maximum blood flow rates should be performed, looking for tubing collapse or other signs of lack of performance, for a period of time representative of expected clinical use.
For Section 5.5.6.1 Needle access ports, testing should be performed with the largest recommended gauge needle and for a period of time representative of expected clinical use. Testing pressure should be 1.5 times the maximum specified pressure, even if it exceeds 700 mm Hg.
For Section 5.5.6.2 Needleless access ports, testing should be performed for a period of time representative of expected clinical use. Testing pressure should be 1.5 times the maximum specified pressure, even if it exceeds 700 mm Hg.
For Section 5.5.9 Transducer protectors, compliance with 4.4.9 and 4.4.9.1 should be verified by testing to withstand pressures of 2 times the manufacturer's recommended pressure for the device.
For Section 5.5.10 Blood pathway flow dynamics, compliance with 4.4.10 should be verified by mechanical hemolysis testing for new or significantly altered hemodialysis tubing designs that affect the pattern of blood flow.
For Section 5.5.11 Pump segment performance, testing should be performed for a period of time representative of expected clinical use and using the maximum recommended blood flow rate.
For Section 6 Labeling, the notes stating "in all cases above, symbols from ISO 15223 may be used where appropriate," does not apply. Symbols alone should not be used in lieu of their meaning.
For Section 6.2 Labeling on the unit containers, e) should provide a statement of sterility and nonpyrogenicity, and whether the entire contents of the container or only the fluid pathways and mating surfaces of connectors integral to the device are sterile. Additionally, the unit container labeling should identify the specific dialysis delivery system(s) with which it has been successfully tested and validated. |
|
| Related CFR Citations and Product Codes: |
Regulation Number |
Device Name |
Device Class |
Product Code |
| §876.5820 |
Accessories, Blood Circuit, Hemodialysis |
Class 2 |
KOC |
| §876.5820 |
Adaptor, Fistula |
Class 2 |
FKM |
| §876.5820 |
Adaptor, Y |
Class 2 |
FJP |
| §876.5820 |
Connector, Blood Tubing, Infusion T |
Class 2 |
FKB |
| §876.5820 |
Connector, Tubing, Dialysate |
Class 2 |
FKY |
| §876.5820 |
Declotting Tray, Kit (Including Contents) |
Class 2 |
FJZ |
| §876.5820 |
Filter, Blood, Dialysis |
Class 2 |
FKJ |
| §876.5820 |
Protector, Transducer, Dialysis |
Class 2 |
FIB |
| §876.5820 |
Set, Dialysis, Single Needle With Uni-Directional Pump |
Class 2 |
FIF |
| §876.5820 |
Set, Dialysis, Single Needle With Uni-Directional Pump, Reprocessed |
Class 2 |
NNG |
| §876.5820 |
Set, Tubing, Blood, With And Without Anti-Regurgitation Valve |
Class 2 |
FJK |
| §876.5820 |
System Accessories, Extracorporeal |
Class 2 |
KXM |
| §876.5820 |
System, Blood, Extracorporeal And Accessories |
Class 2 |
LLB |
| §876.5820 |
Tray, Start/Stop (Including Contents), Dialysis |
Class 1 |
FKG |
| §876.5820 |
Tubing, Dialysate |
Class 2 |
FID |
| §876.5820 |
Tubing, Dialysate (And Connector) |
Class 2 |
KQQ |
| |
|
| Relevant Guidance: |
| Guidance for Industry and FDA Staff, Hemodialysis Blood Tubing Sets - Premarket Notification [510(k)] Submissions. Issued April 23, 2008 |
|
| FDA Technical Contacts: |
| Carolyn Neuland |
| FDA/CDRH/ODE |
| 10903 New Hampshire Avenue Building 66, Room G226 |
| Silver Spring MD 20993 |
| 301/796-6523 |
| Email: carolyn.neuland@fda.hhs.gov |
|
| Joshua Nipper |
| FDA/CDRH/ODE |
| 10903 New Hampshire Avenue Building 66, Room G230 |
| Silver Spring MD 20993 |
| 301/796-6524 |
| Email: joshua.nipper@fda.hhs.gov |
|
|
|
| * In the United States, copies of this standard can be obtained from: |
| American National Standards Institute (ANSI)* |
| 25 West 43rd Street |
| 4th Floor |
| New York, NY 10036 |
|
|