Recognition List Number: 027 Publication Date: 08/02/2011
Part B: SUPPLEMENTARY INFORMATION
Recognition Number 10-64: ANSI Z80.7-2002, Ophthalmics - Intraocular Lenses. (Ophthalmic)
Date of Standard: 2002. |
| Address of Standards Organization: |
| American National Standards Institute (ANSI) |
| 25 West 43rd Street |
| 4th Floor |
| New York, NY 10036 |
| |
|
| CDRH Office and Division Associated with Recognized Standards: |
| OFFICE OF DEVICE EVALUATION (ODE) |
|
|
| Devices Affected: |
|
| Processes Affected: |
| PMA, IDE, PDP, HDE, Design Control Input |
|
| Type of Standard: |
| Vertical, National |
| Extent of Recognition: |
Complete standard and any annexes, with the following modification to 8.2.3:
The endotoxin level shall be 0.2 endotoxin units or less per intraocular lens.
NOTE: In using this maximum endotoxin level, the uncertainty of the test method should be taken into account. |
|
| Related CFR Citations and Product Codes: |
Regulation Number |
Device Name |
Device Class |
Product Code |
| §886.3600 |
Intraocular Lens |
Class 3 |
HQL |
| |
|
| Relevant Guidance: |
| There is no relevant guidance developed at this time. |
|
| FDA Technical Contact: |
| Don Calogero |
| FDA/CDRH/ODE |
| 10903 New Hampshire Avenue Building 66, Room 2500 |
| Silver Spring MD 20993 |
| 301/796-6483 |
| Email: don.calogero@fda.hhs.gov |
|
|