Recognition List Number: 034 FR Publication Date: 01/30/2014 

Part B: Supplementary Information

Recognition Number 11-220: ASTM F2068-09, standard specification for femoral prostheses - metallic implants. (Orthopaedics)

Date of Standard: 2009. 
Address of Standards Development Organization:
 ASTM International (ASTM)
  100 Barr Harbor Drive
  West Conshohocken, PA 19428-2959
 
CDRH Office and Division associated with recognized standards:
OFFICE OF DEVICE EVALUATION (ODE)
Devices Affected:
Femoral Components of Total Hip Replacement (THR) and Hemi-Hip Replacements
Processes Affected:
510(K), IDE, PMA, PDP, Design Controls
Type of Standard:
 Vertical, National
Extent of Recognition:
Complete standard.

Regarding clauses 6.1.1 and 6.1.2 in ASTM F2068-09, for stems 250mm in length as defined in ISO 7206-4: 2010, one should reference ISO 7206-4: 2010 and its corresponding FDA Supplementary Information Sheet.

Standard provides femoral component fatigue test methods. Additional information must be provided to document that the design of the product will provide adequate mechanical properties for the particular application. In applications subjected to wear or articulation, additional data must be provided to document resistance to wear and abrasion.
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented Class 3 LPF
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented Class 3 MRA
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate Class 2 MEH
§888.3360 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Class 2 LWJ
§888.3300 Prosthesis, Hip, Constrained, Metal Class 3 KXD
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer Class 2 KWZ
§888.3320 Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) Class 3 JDL
§888.3330 Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) Class 3 KWA
§888.3340 Prosthesis, Hip, Semi-Constrained, Composite/Metal Class 2 KMC
§888.3350 Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented Class 2 OQH
§888.3350 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Class 2 JDI
§888.3353 Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented Class 2 OQI
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish Class 2 MAY
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate Class 2 MEH
§888.3358 Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented Class 2 OQG
§888.3358 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented Class 2 LPH
§888.3358 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous Class 2 MBL
§888.3360 Prosthesis, Hip, Hemi-, Femoral, Metal Class 2 KWL
§888.3360 Prosthesis, Hip, Femoral Component, Cemented, Metal Class 2 JDG
§888.3360 Prosthesis, Hip, Hemi-, Femoral, Metal Class 2 KWL
§888.3360 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Class 2 LWJ
§888.3360 Prosthesis, Upper Femoral Class 2 JDD
§888.3390 Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented Class 2 KWY
Relevant Guidance:
Guidance for Industry and FDA Staff: Non-clinical Information for Femoral Stem Prostheses, September 17, 2007
FDA Technical Contact:
 John Goode
  FDA/CDRH/ODE
  10903 New Hampshire Avenue Building 66, Room 1510
  Silver Spring MD 20993
  301/796-6407
  email: john.goode@fda.hhs.gov
-
-