MedSun: Newsletter #75, August 2012
Newsletter #75, August 2012
Articles
Arrow International, Inc. Multi-Lumen Venous Catheterization Set with Blue FlexTip ARROWg+ard Catheter: Class I Recall - Contains Unlabeled Drug Ingredient |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of Arrow International, Inc. Multi-Lumen Venous Catheterization Set with Blue FlexTip ARROWg+ard Catheter. The device's labeling erroneously states that the product "contains no medication," however; the device contains chlorhexidine and silver sulfadiazine. Additionally, the product’s label is missing the appropriate chlorhexidine contraindication. This product may cause serious adverse health consequences, including death, if used in a patient who is allergic to either chlorhexidine or silver sulfadiazine.
Additional Information:
FDA MedWatch Safety Alert. Arrow International, Inc. Multi-Lumen Venous Catheterization Set with Blue FlexTip ARROWg+ard Catheter: Class I Recall – Contains Unlabeled Drug Ingredient. July 30, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm313721.htm
CareFusion EnVe Ventilators: Class I Recall - Leak In Patient Breathing Circuit or System |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of CareFusion EnVe Ventilators. A leak may occur in the patient breathing circuit or the system, resulting in the ventilator not holding the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously. The device will activate both audible and visual alarms to notify the health care professional that ventilation delivery to the patient may be compromised.
Additional Information:
FDA MedWatch Safety Alert. CareFusion EnVe Ventilators: Class I Recall - Leak In Patient Breathing Circuit or System. July 20, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm312966.htm
Hospira Injectable Drug Products: Recall - Visible Particulates from Defective Glass Vials |
Print Item E–mail Item |
FDA MedWatch Safety Alert
Hospira and FDA notified healthcare professionals of a nationwide recall of certain injectable drug products, due to visible particles embedded in the glass located at the neck of the vial. There may be potential for product to come into contact with the embedded particles and the particles may become dislodged into the solution. In the event in which particulate matter could be injected into a patient, there may be the potential for patient injury where medical intervention may be required.
Additional Information:
FDA MedWatch Safety Alert. Hospira Injectable Drug Products: Recall - Visible Particulates from Defective Glass Vials. July 16, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm312048.htm
Maquet Medical Systems USA FLOW-i Anesthesia System: Class I Recall - Device Field Correction |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of the Maquet Medical Systems USA FLOW-i Anesthesia System (software). A technical alarm may be generated on the FLOW-i system when using the Manual/Automatic (MAN/AUTO) switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in between the two positions. Use of this product may cause serious adverse health consequences, including death.
Additional Information:
FDA MedWatch Safety Alert. Maquet Medical Systems USA FLOW-i Anesthesia System: Class I Recall - Device Field Correction. July 11, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm311471.htm
Alere Triage Diagnostic Tests (Multiple Tests): Class I Recall - Potential for Significantly Decreased Precision |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of Alere Triage products. Identified lots may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing. There have been reports of patients receiving inappropriate clinical management which may have been due to such erroneous results.
Additional Information:
FDA MedWatch Safety Alert. Alere Triage Diagnostic Tests (Multiple Tests): Class I Recall - Potential for Significantly Decreased Precision. July 11, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm311405.htm
GE Healthcare Aestiva/5 7900 Ventilator - Class I Recall - Potential for Unrecognized Overdose |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of the GE Healthcare Aestiva/5 7900 Ventilator. There is a potential for two vaporizers to deliver each agent at the same time. This could result in over-delivery of a single agent if both vaporizers contain the same agent, or in delivery of more than one agent. Unrecognized overdose with simultaneous inhaled anesthetics is manifested by hypotension, arrhythmias, and bradycardia that, if persists and remains unrecognized, may lead to circulatory shock.
Additional Information:
FDA MedWatch Safety Alert. GE Healthcare Aestiva/5 7900 Ventilator - Class I Recall - Potential for Unrecognized Overdose. July 11, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm311361.htm
FDA Safety Communication: UPDATE on Advanced Sterilization Products (ASP) Sterrad Cyclesure 24 Biological Indicators Revised Expiration Dates |
Print Item E–mail Item |
FDA Medical Device Safety
The FDA is issuing this communication to inform personnel at facilities that utilize Sterrad sterilizers of the following:
*ASP has issued a voluntary recall of Sterrad Cyclesure 24 Biological Indicators due to the revised expiration time (shelf life) from 15 months to 6 months. This recall followed an FDA review of ASP data, which showed that the Sterrad Cyclesure 24 Biological Indicators cannot effectively monitor the sterilization process throughout a 15-month shelf life.
*Use of Sterrad Cyclesure 24 Biological Indicators beyond their 6-month shelf life may prevent verification of proper sterilization. However, the FDA believes that the risk of infection for an individual patient is very low, given that this biological indicator is only one of three sterilization system monitors and the window of possible exposure to the malfunctioning Cyclesure 24 devices is relatively short.
*Supplies of Sterrad Cyclesure 24 Biological Indicators may be temporarily reduced. The FDA is working with ASP to make suitable product available and to assure that health care facilities experience little or no disruption.
Only the lot numbers of Sterrad Cyclesure 24 Biological Indicators that are listed in ASP’s Recall Alert are affected by ASP’s voluntary recall. In order to determine if the Sterrad Cyclesure 24 Biological Indicators you have in inventory are subject to this recall, please refer to the list of affected lots found in ASP’s Recall Alert.
Additional Information:
FDA Medical Device Safety. FDA Safety Communication: UPDATE on Advanced Sterilization Products (ASP) Sterrad Cyclesure 24 Biological Indicators Revised Expiration Dates . July 7, 2012.
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm310549.htm
Stryker Initiates Voluntary Product Recall of Modular-Neck Stems |
Print Item E–mail Item |
FDA Medical Device Safety
Stryker has voluntarily recalled its Rejuvenate and ABG II modular-neck stems. Stryker has notified healthcare professionals and regulatory bodies of this voluntary recall. Patients who received a Rejuvenate Modular or ABG II modular-neck stem are encouraged to contact their surgeon. Patients uncertain if they have one of these products implanted should contact their surgeon or consult their medical records.
Additional Information:
FDA Medical Device Safety. Stryker Initiates Voluntary Product Recall of Modular-Neck Stems Action Specific to Rejuvenate and ABG II Modular-Neck Stems. July 6, 2012.
http://www.fda.gov/Safety/Recalls/ucm311043.htm
CareFusion Airlife Infant Breathing Circuit: Class I Recall - Potential for Leak in Closed Ventilation System |
Print Item E–mail Item |
FDA MedWatch Safety Alert
FDA notified healthcare professionals of the Class I recall of the Airlife infant breathing circuit due to the risk that the Y-adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient. These products may cause serious adverse health consequences, including death. The recalled products were distributed from July 1, 2010 through May 7, 2012. See the Recall Notice for a listing of affected lot numbers.
Additional Information:
FDA MedWatch Safety Alert. CareFusion Airlife Infant Breathing Circuit: Class I Recall - Potential for Leak in Closed Ventilation System. July 6, 2012.
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm310871.htm
FDA Website - Tubing and Luer Misconnections: Preventing Dangerous Medical Errors |
Print Item E–mail Item |
FDA Website
This recently updated FDA website explains how misconnections occur, provides real-life examples of misconnections, and offers tips and additional resources for healthcare professionals on how to prevent them. Guidance for manufacturers of enteral feeding systems is also included.
Additional Information:
FDA Website. Tubing and Luer Misconnections: Preventing Dangerous Medical Errors. July 26, 2012.
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/TubingandLuerMisconnections/default.htm
First Rapid Home-Use HIV Kit Approved for Self-Testing |
Print Item E–mail Item |
FDA Consumer Update
The FDA has approved the OraQuick In-Home HIV Test, a rapid home-use HIV test kit that does not require sending a sample to a laboratory for analysis. The kit provides a test result in 20-40 minutes, and consumers can test themselves in their own home. The kit, which tests a sample of fluid from your mouth, is approved for sale in stores and online to anyone age 17 and older. FDA also wants consumers to know that positive test results using the OraQuick test must be confirmed by follow-up laboratory-based testing.
Additional Information:
FDA Consumer Update. First Rapid Home-Use HIV Kit Approved for Self-Testing. July 18, 2012.
http://www.fda.gov/ForConsumers/ConsumerUpdates/ucm310545.htm
Unique Device Identification Proposed Rule |
Print Item E–mail Item |
FDA - Medical Devices
The Food and Drug Administration (FDA) has released a proposed rule that most medical devices distributed in the United States carry a unique device identifier, or UDI. A UDI system has the potential to improve the quality of information in medical device adverse event reports, which will help the FDA identify product problems more quickly, better target recalls and improve patient safety.
Additional Information:
FDA - Medical Devices. Unique Device Identification Proposed Rule. July 13, 2012.
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/UniqueDeviceIdentification/default.htm
Highlighted MedSun Reports
Highlighted Reports |
Print Item E–mail Item |
This section contains a sample of reports from all the MedSun reports received during a particular period. The reports were submitted by MedSun Representatives. In some instances the reports have been summarized and/or edited for clarity. The entries that follow represent a cross section of device-related events submitted by MedSun reporters during the period May 1 through May 31, 2012. All other reports can be searched under the ‘MedSun reports’ menu pane. Note: the two month delay is due to quality control and follow-up.
CARDIOVASCULAR
Device 1:
Type: Pacemaker, External, Temporary
Manufacturer: Medtronic, Inc.
Brand: 5348 Single-chamber
Model#: 5348
Device 2:
Type: Cable, Pacemaker, Disposable
Manufacturer: Remington Medical, Inc.
Brand: Adap-2000
Model#: ADAP-2000
Problem:
The nurse was attempting to insert a pacemaker adapter cable into the pacemaker. Adapter had obvious malformation that prohibited cable from fully inserting into the pacemaker.
Device 1:
Type: Cable, Pacemaker, External, Single Use
Manufacturer: Medtronic Cardiac Rhythm Disease Management
Model#: 5487
Lot #: 5015301
Device 2:
Type: Cable, Pacemaker, External, Single Use
Manufacturer: Medtronic Cardiac Rhythm Disease Management
Model#: 5487
Lot #: 5015401
Problem:
Patient was brought to Cath Lab dependent on pacing, for placement of a semi-permanent pacer for complete heart block. Right ventricular (RV) pacing lead (Medtronic Novus) was placed in the RV apex. Pacing threshold through analyzer was less than 1v at 0.4 ms. Lead was attached to the Medtronic cable to the pacing box, but no capture, no spikes. Second pacing box was tried, no capture, no spikes. Medtronic cable was replaced with a second Medtronic cable, again with no capture and no spikes. A third Medtronic cable was then opened and attached, which worked. Nothing different was done when attaching the third cable than when attaching the previous two. After these events a similar sequence of events with another patient was described anecdotally, however at this time we do not have specifics on the second case.
Device:
Type: Cardiovascular Procedure Kit
Manufacturer: Navilyst Medical, Inc.
Brand: Namic Left Heart Kit
Model#: 197002002
Lot #: GFWB1443
Other #: Product packed inside L Heart Kit
Problem:
At the end of the case the guidewire was being wiped with a sponge moistened with heparinized saline. When the sponge was observed after wiping the guidewire, a green colored residue was noted on the sponge. The process of wiping the guidewire with a sponge was repeated and more green colored residue was observed on the sponge. The guidewire and sponges were saved. Packaging for this guidewire was not saved. Customer Service at the company was notified and guidewire/sponges will be returned to the company for investigation.
Device:
Type: Cardiovascular Procedure Kit
Manufacturer: Navilyst Medical Inc
Brand: Namic Left Heart Kit
Lot #: GFWB1443
Cat #: 19700202
Problem:
A staff member opened a left heart kit and opened the packaging and used a 3 mm J angiographic guidewire, Navilyst brand. When wiping the guidewire with a 4x4 that was moistened with heparinized saline, a green colored film was noted on the 4x4. The guidewire was looked at closely and staff noted that the guidewire appeared green in color. Customer Service at Navilyst was contacted and this report was added to the previous report. The guidewire and 4x4 will be sent to Navilyst for further inspection.
Device:
Type: Catheter, Continuous Flush
Manufacturer: Micro Therapeutics, Inc. Dba Ev3 Neurovascular
Brand: Marathon
Model#: 105-5055
Cat #: 105-5055
Problem:
Patient with a complicated history including a ruptured posterior fossa arteriovenous malformation status post (s/p) posterior fossa craniotomy and intraparyenchymal hemorrhage evacuation. Eleven days earlier the patient went to the for an endovascular treatment of an arteriovenous malfunction (AVM). Per dictation, " ...following termination of the Onynx-18 infusion, the physicians prepared to withdraw the microcatheter from the patient. Using gentle traction, while aspirating on the catheter, physician attempted to withdraw the microcatheter from the patient. As they exerted additional force on the microcatheter, the superior cerebellar artery appeared to deflect from its baseline position. Under constant fluoroscopic guidance, traction was maintained on the microcatheter. It was then noted that there was a sudden decrease in tension of the microcatheter, which had fractured in the proximal vertebral artery. The proximal portion of the microcatheter was thus free and removed from the patient. The retained portion of the microcatheter remained in place in the left superior cerebellar artery, adherent to the Onyx. The 6-French groin sheath was left in place with the plan for future intra-operative angiogram."
Device:
Type: Catheter, Intra-aortic Balloon
Manufacturer: Datascope Corp.
Brand: Sensation
Model#: Sensation
Cat #: 0684-00-0469-01
Problem:
Surgeon inserted the IAB catheter with some difficulty. When Respiratory Therapy hooked the device up to the monitoring console no readings were obtained and the device would not calibrate. Fluoroscopy was used to assist in visualizing the placement. This did not yield satisfactory results and the catheter was removed from the patient. No visible signs of damage to the catheter were noted post removal.
Device:
Type: Catheter, Intra-aortic Balloon
Manufacturer: Arrow International, Inc.
Brand: Ultraflex
Model#: IAB-06840-U
Lot #: KF2032806
Cat #: IAB-06840-U
Problem:
The intra-aortic balloon catheter (IAB) ruptured when initial gas filling was occurring. A visual of red foamy substance in the gas port on the balloon was noted. The defective IAB and sheath removed as one unit. New sheath and IAB were inserted after the defective IAB and sheath were removed. Unable to remove the ruptured balloon out of the sheath so the whole unit was removed altogether.
Device:
Type: Device, Percutaneous Retrieval
Manufacturer: Cook Medical
Brand: Günther Tulip
Model#: GTRS-200-RB
Lot #: 3098415
Cat #: G13287
Problem:
Doctor was doing a vena cava filter removal. The end of the retrieval device broke off which is the wire. The wire is in the filter. Doctor could not retrieve the filter or wire. The package and the rest of the retrieval device were given to the Supervisor.
Device:
Type: Snare, Percutaneous Retrieval
Manufacturer: ev3
Brand: Amplatz Gooseneck
Model#: GN2500
Lot #: 9557678
Problem:
Patient undergoing Mediport revision in Interventional Radiology. During the procedure when inserting the Amplatz microwire snare, it was noted that the radiopaque marker at the tip of the wire was no longer seen by fluoroscopy. The wire tip was identified in the right atrium then right ventricle and will likely lodge in the patient's lung.
======================
Manufacturer response for Goose neck snare, 6.0 Fr Snare Catheter 120cm x 25cm x 102cm (per site reporter)
======================
Not known at this time.
Device:
Type: System, Laser, Transmyocardial Revascularization
Manufacturer: Cardiogenesis Corporation
Brand: Sologrip Iii
Lot #: TA-03733
Other #: 705-37020
Problem:
Physician using Cardiogenesis Sologrip III Handpiece Delivery System. Wire adjacent to handle burned two holes in physician's gown at his chest area. Substituted device with another Cardiogenesis Sologrip III Handpiece. Procedure completed without further incidence. No untoward patient effects. Physician not injured. The calibration for this device is 47%.
Device:
Type: Transducer, Blood-pressure, Extravascular
Manufacturer: ICU Medical, Inc
Brand: Neonatal Disposable Transducer Kit
Cat #: 42634-05
Problem:
While performing hourly patient temperature/assessment at 0200, RN noticed blood backing up from blue lumen of UVC (umbilical vessel catheter) into tubing. RN assessed all parts of the line to make sure that all connections were tight, when RN noticed TPN/lipids leaking from where the stopcock met the transducer. RN changed the stopcock that immediately preceded the transducer thinking that the stopcock had a crack. RN then flushed the line to find that the IV fluid was still leaking from where the stopcock and transducer met. RN changed the entire transducer line and the problem was resolved.
======================
Manufacturer response for IV Kit, Transpac IV Monitoring Kit Neonatal, 12", 3 Stopcocks, Disposable Transducer, 30 Ml Squeeze Flush (per site reporter)
======================
awaiting response
Device:
Type: Tubing Set, Contrast Injection
Manufacturer: Medrad, Inc.
Brand: Avanta Fluid Management System
Model#: AVA 500 MPAT
Lot #: 113801
Problem:
During set up of the multi-patient disposable set for the Medrad injector, the vent cap fell off, which could have caused the contrast to leak had it happened during a case. Unit was taken down and a new one from the same lot was used and worked fine.
EAR, NOSE, & THROAT
Device:
Type: System, Vocal Cord Medialization
Manufacturer: Bioform Medical Inc.
Brand: Radiesse Injection Needle
Lot #: LOG247665
Cat #: 8044MO
Problem:
Physician was attempting to inject and medialize the true vocal cords. The injection needle malfunctioned/became clogged and was not useable.
GASTROENTEROLOGY & UROLOGY
Device 1:
Type: Light Cable
Manufacturer: Lumitex Medical Devices, Inc.
Brand: Light Mat Ultra-thin Flexible
Lot #: I01802
Other #: Ref# UA2550
Device 2:
Type: Light Sources, Xenon
Manufacturer: Integra LifeSciences Corp.
Brand: Luxtex Mlx
Model#: MLX
Problem:
During a surgical procedure a surgical illuminator was connected to a Luxtex light source. The surgical illuminator was a Light Mat, mfg, LumitexMD Inc. Ref# UA2550 Lot# I01802. The circulating nurse noticed an electrical burning smell that continued to get worse. She discovered the cord from the surgical illuminator was smoking. She went to remove the cord and as she grabbed the cord to remove it from the light source, she burned her fingers. The metal connection of the surgical illuminator was very hot.
======================
Manufacturer response for Light Sources, Xenon, Luxtex MLX (per site reporter)
======================
They told us we are not using their illuminator so it is our problem.
Device:
Type: Ng Feeding Tube
Manufacturer: Neomed
Brand: Tube, Feeding, Neomed
Model#: FTL8.0S-EO
Lot #: 20111104
Problem:
During a routine x-ray viewing that the nasogastric (NG) feeding tubes were not showing up on the X-rays. Hospital contacted Neomed at the time and Neomed indicated they had heard of the issue from another institution. They asked facility to remove several lot numbers from use. Facility removed all lots and was sent new materials. However, as of yesterday, there are still issues viewing the radio opaque stripe on the x-ray film.
Device:
Type: Suturing Device
Manufacturer: Covidien, Formerly US Surgical, A Division of Tyco Healthcare
Brand: Endo Stitch
Model#: 173016
Lot #: N9A64
Cat #: 173016
Problem:
Endo Stitch was being used during a Laparoscopic Hysterectomy. During the cuff closure, an Endo Stitch needle was lost during removal at the trocar and had obviously become detached from the instrument. Needle was not located during search of abdomen or external sterile field. Staff had concerns that the needle would not show on an X-ray due to the size of the needle.
Device:
Type: Tube, Feeding, Syringe
Manufacturer: Ameritus Medical Products
Brand: Syringe, Ameritus, Oral, Enteral
Model#: AEG30-KMS 30 ml
Lot #: 20120210 201501
Problem:
A new enteral Ameritus feeding tubing system was put in place in the NICU. The tubing was attached to the syringe securely and attached to connecting tubing and the feeding was started. However, feeding was noted to have leaked out of an apparent hole or crack in the syringe. After investigating, the nurse noticed that there was a leak in the feeding syringe. The nurse felt a rough edge on the rim of the syringe where the (male) pointed infusion hole lies. There may have also been a pin hole or crack in the syringe, but it could not be clearly seen. The male female connection from the syringe to the feeding tubing was secure and not backing out through the connection as if it was blocked. In fact, the feeding tubing was flushed without difficulty. Again, it appeared to be a problem with the actual syringe leaking. As a result, a new syringe and extension set was hung and infused without difficulty. There was no harm to the patient.
GENERAL & PLASTIC SURGERY
Device:
Type: Clip Applier
Manufacturer: Covidien
Brand: Endo Clip Auto Suture Clip Applier
Model#: REF 176620
Lot #: N1H0889X
Problem:
During a laparoscopic lobectomy, a 5mm endoclip clip applier made by Covidien with reference #176620, and lot# N1H0889X was used. The clip applier was placed on a larger artery, which fed the lobe of the lung which was being removed. The clip from the clip applier dislodged from the applier halfway. The clip clamped on the artery but would not completely come free from the applier.
Device:
Type: Sealing Device
Manufacturer: Ethicon Endo-Surgery, Inc.
Brand: Enseal Trio
Model#: ETRIO335H
Lot #: H43Z60E
Cat #: ETRIO335H
Problem:
During the procedure, the surgeon noted that after he deployed the EnSeal, the blade between the jaws snapped. The surgeon noted that there were dense adhesion's midline; however it is unknown whether this contributed to the device breaking. A new device was obtained and utilized without incident. It is unknown whether the new device was of the same, or a different, lot. There were no adverse effects to the patient.
Device:
Type: Sealing Device
Manufacturer: Covidien, Formerly Valleylab, A Division of Tyco Healthcare
Brand: Ligasure Atlas
Model#: LS1037
Lot #: SIC0019X
Cat #: LS1037
Problem:
LigaSure device was being used when the unit's jaw would close, but the unit would not cauterize. Unit was functioning normally outside of patient's body. A new LigaSure was opened and functioned as designed by the manufacturer.
Device:
Type: Sponge, Gauze
Manufacturer: Medline Industries, Inc.
Brand: Medline Accu-sort Gauze Sponge Usp Type Vii Gauze
Other #: Reorder NON21428
Problem:
There has been a packaging change for the Medline gauze sponge Type VII (4X4) from a soft container to a hard container (see pictures). This packaging change is similar to the packaging of the Raytec safety sponge from another manufacturer. The Medline is not radiopaque and the Raytec sponge is radiopaque. Staff could inadvertently utilize the wrong sponge when radiopaque is needed. Health professional comment: The non-radiopaque sponge material packaging is now similar to the radiopaque sponge packaging. If the two products are mixed up on the sterile field we could have a non-counted sponge used in a wound and perhaps left behind because it is a dressing material, not counted and it does not have a radiopaque strip on it for easy identification in a post-op x-ray.
======================
Manufacturer response for gauze sponge, Medline Accu-Sort gauze sponge USP Type VII Gauze (per site reporter)
======================
No response from manufacturer.
See device images:


Device:
Type: Stapler, Reload, Surgical
Manufacturer: Covidien, Formerly US Surgical, A Division of Tyco Healthcare
Brand: Endo Gia Universal Roticulator
Model#: 030450
Lot #: N0H0131
Cat #: 030450
Problem:
The surgeon was performing a chest exploration. During the procedure, he was able to confirm the need for a lobectomy. He was using staples for dissection and separation of vasculature. Several loads were used during the procedure. However, while he was using the gun on the pulmonary artery, the surgeon reports the gun misfired. The surgeon discovered the misfire when he was pulling the gun back and the pulmonary artery tore. The patient lost a significant amount of blood and became hypotensive. The artery was repaired, and the patient was transferred to the ICU for recovery.
GENERAL HOSPITAL
Device:
Type: Catheter, Intravascular
Manufacturer: Teleflex Inc,
Brand: Arrowgard Blue Plus Quad-lumen Cvc
Model#: AK-45854-SK
Lot #: RF1117405 120711 1154
Cat #: AK-45854-SK
Problem:
Physician was placing central line. After placing the line it was noticed that one of the ports was leaking. It was the white proximal port below the hub. The port was clamped and labeled not to use and was saved once the device was discontinued.
Device:
Type: Catheter, Intravascular
Manufacturer: Becton Dickinson
Brand: Saf-t-intima
Lot #: 2060135
Cat #: 383323
Problem:
While the nurse attempted IV insertion with BD Intima catheter, the safety did not engage and the needle came out disconnected from the catheter. Blood flowed from the catheter.
Device:
Type: Catheter, Picc
Manufacturer: Bard Access Systems, Inc.
Brand: Groshong Picc
Model#: 7717307
Lot #: REVK1172
Problem:
Called to evaluate patient's PICC for Alteplase administration due to sluggish flush and absent blood return. The nurse reported that the PICU nurse had reported lack of blood return on transfer two days prior, but no action had been taken. On my assessment, I was unable to flush the catheter at all, but was able to obtain blood return. I spoke with Physician who recommended hub replacement, stating that he has had multiple instances (>10) of 3 Fr PICC's that will not flush, but aspiration is possible. Replacement of the hub restores patency. Another nurse also experienced this and has contacted the manufacturer. Awaiting feedback once details are sent. Two hubs have been saved for return to Bard.
Device:
Type: Set, Blood Transfusion
Manufacturer: B.Braun
Brand: Infusomat Space Pump Y-type Blood Set
Cat #: 490043
Problem:
The nurse reported that a B.Braun Infusomat space pump Y-type blood set leaked above the drip chamber where the two tubings meet. The nurse was priming the tubing with normal saline for blood transfusion when she noticed that fluid was leaking just above the drip chamber. The tubing was sequestered and given to the Risk Manager. Unfortunately, the packaging was not saved.
NEUROLOGY
Device:
Type: Shunt, Central Nervous System
Manufacturer: Integra Lifesciences Corporation
Brand: Accudrain
Cat #: INS-8400
Problem:
The patient had a PICC line and an EVD (external ventricular drain) line in place. The patient was in the MRI suite for an exam with contrast. The Radiology tech intended to inject the contrast dye into the PICC line. Instead, the tech accessed the EVD port and injected the dye into the EVD system. The EVD system was set up to drainage. Contrast dye was inadvertently injected into both the brain and the drainage bag. The event was discovered immediately. The physician was notified. MRI images indicate that some dye was present in the brain ventricles. The patient was monitored closely and had no adverse reactions. The plan of care did not change and there was no patient harm.
This event is an unintended Luer misconnection. The tubing on the EVD system near the port is clearly marked and highlighted yellow to bring attention to the port and its intended use. However, the syringe Luer has no engineered properties that physically prevent the Luer connection. In this case, the radiology tech did not trace the line to its origin. Because the patient was on the MRI table, all lines, both the PICC line and the EVD line, were positioned up toward the patient's head. The injection ports on the PICC line are visually similar to the port on the EVD system except that the EVD system has bright yellow marking on the tubing leading up to and coming away from the EVD port.
All staff involved in this event is experienced in performing an MRI with contrast. It is more common to have a patient with an IV line or PICC line in the radiology department than it is to have a patient with an EVD. There was a false sense of reassurance that the port was correct. We have changed our practice to require all techs to trace the line to its source of origin before injecting dye. We have not had an event like this in the past.
ORTHOPEDIC
Device:
Type: Drill Bit
Manufacturer: Arthrex, Inc.
Lot #: 466474A
Cat #: AR-1360B-CP
Problem:
Surgeon was using drill bit that was labeled at 6cm intending to drill a hole in that diameter. The hole appeared bigger and the 6cm screw would not fit. The drill bit was measured and was actually 9cm. Surgeon proceeded to use screw to fit the 9cm hole.
Device:
Type: Spinal Vertebral Body
Manufacturer: NuVasive, Inc.
Brand: Cordent Xl
Lot #: 6908850
Problem:
The cage was to be implanted in spine post spinal tumor removal. Two cages broke during implantation. One cage retrieved entirely but in two pieces; the second cage retrieved in multiple pieces with one piece irretrievable and remaining in patient. Third cage (from different vendor) implanted.
RADIOLOGY
Device 1:
Type: Mri Scanner
Manufacturer: GE Healthcare
Brand: Signa Hdxt
Model#: Signa HDxt 1.5T
Cat #: 2416758
Other #: last upgrade Sept-Nov/2011
Device 2:
Type: Electrode, Ecg, Mri
Manufacturer: Covidien, Formerly Ludlow, A Division of Tyco Healthcare
Brand: Medi-trace
Model#: 850 Series
Lot #: 026718
Cat #: 22853
Other #: 900400 (Medi-Trace)
Device 3:
Type: Monitor, Physiological, Mri
Manufacturer: MedRad, Inc.
Brand: Veris
Other #: 10434
Problem:
Patient had outpatient MRI and was discharged. Patient returned as an inpatient 7 days later. Burns were reported by the patient at that time. Burns were on chest in areas where electrodes were applied.
======================
Manufacturer response for MRI Scanner, Signa HDxt (per site reporter)
======================
On site evaluation of equipment by GE. GE reports no problems when doing test and calibrations.
======================
Manufacturer response for Monitor, Physiological, MRI, Veris (per site reporter)
======================
An investigation questionnaire was completed. Follow up recommendations were received.
======================
Manufacturer response for Electrodes, ECG, MRI (per site reporter)
======================
Manufacturer requested sample product from the same lot numbers to be returned to them for evaluation.
Device:
Type: Dosimeter, Radiation, Diode Detector
Manufacturer: Sun Nuclear Corporation
Brand: Qed
Model#: QED Scatter 11300-2
Problem:
Upon removal of dosing diodes from patient's skin, a blister were noted at the site of one of the diodes.
Device:
Type: Mri Scanner
Manufacturer: GE Healthcare
Brand: Signa Hdxt
Model#: 2416758
Other #: last upgrade Sept-Nov 2011
Problem:
Patient was receiving MRI of shoulder. Shoulder coil was in use. Patient was unable to be still for duration of procedure. Procedure time was lengthened as a result. Patient began to complain of a burning sensation in shoulder area. Procedure was terminated at that time.
Onsite evaluation of equipment by GE - GE reports no problems when doing testing and calibrations. Unable to reproduce effects. On site visit by applications person helpful in identifying possible related causes.
Device 1:
Type: X-ray Shield, Protective, Personal
Manufacturer: RF Design, Inc.
Brand: Rf Design
Model#: VS11702
Cat #: VS11702
Other #: 0407113
Device 2:
Type: X-ray Shield, Protective, Personal
Manufacturer: RF Design, Inc.
Brand: Rf Design
Model#: VS11701
Cat #: VS11701
Other #: 0407114
Device 3:
Type: X-ray Shield, Protective, Personal
Manufacturer: RF Design, Inc.
Brand: Rf Design
Model#: VS11701
Cat #: VS11701
Other #: 0407112
Device 4:
Type: X-ray Shield, Protective, Personal
Manufacturer: RF Design, Inc.
Brand: Rf Design
Model#: EB601
Cat #: EB601
Other #: 0801111
Problem:
During scheduled annual maintenance (done one month earlier than scheduled) on lead aprons/vests, four lead aprons/vests were found to have significant damage. All four of the aprons were less than one year old. Both the tech and the vendor reported said that in over twenty years they have never seen damage like this before. There were multiple gaping tears in odd places (i.e. thyroid & groin area). The users of the lead aprons/vests cared for them per manufacturer instructions. Each of the four aprons/vests has a different number based on color and size. Our Radiation Safety Officer will be conducting dosimetry radiation badge reviews for each person involved. The lead aprons/vests were returned to the vendor for evaluation and have since been replaced. Of note: Older apron/vests have not shown this level of damage.
The on-site rep indicated that he had not seen damage such as this before and replaced all four lead devices.
Device:
Type: X-ray System, Cath/angio
Manufacturer: Philips Medical Systems
Brand: Allura Fd 20
Model#: Allura FD 20
Problem:
Poor image quality: Bolus chase/run-off images are poor. Unable to subtract bowel during abdomen/pelvis imaging. Images being transferred to PACS become very dark and unreadable.
Manufacturer has spent multiple hours with multiple technicians and specialists to resolve the issue. As of mid-April, 2012 the issue continues to be unresolved.
Device 1:
Type: X-ray System, Radiographic
Manufacturer: GE Medical Systems, LLC
Brand: Ge Revolution Xq/i With Cotello
Device 2:
Type: X-ray System, Radiographic
Manufacturer: GE Medical Systems, LLC
Brand: Ge Revolution Xr/d With Cotello
Problem:
The GE XQ/i and XR/d X-ray units are capable of dual-energy subtraction chest radiography (DES). The time necessary to perform the image processing for DES was long. In order to accelerate the speed at which the system processes DES images, GE retrofit the XQ/i and XR/d systems with a second, external computer called the "Cotello." The addition of the Cotello processor successfully accelerated the speed at which DES images could be generated. However, we recently discovered that the Cotello hampers the system for correctly processing radiographs produced with fixed radiographic technique, where the user specifies the mAs. Pediatric chest radiographs and other routine radiographic procedures are performed using fixed radiographic techniques. Any radiographs obtained using an XQ/i or XR/d system when using a fixed radiographic technique produce images that are very poor quality. Because of poor image quality, these images may not be usable for diagnostic purposes. This could result in the patient having to repeat the examination, which leads to increased radiation exposure dose(s) to the patient.
~~~~~~
Special Note: The lollipop icon distinguishes highlighted reports that describe medical device events involving neonatal or pediatric patients, or those events involving a medical device that is indicated for use in neonatal and pediatric patient populations. FDA defines pediatric patients as less than 21 years of age.
Updated August 1, 2012