This section contains a sample of reports from all the MedSun reports received during a particular period. The reports were submitted by MedSun Representatives. In some instances the reports have been summarized and/or edited for clarity. The entries that follow represent a cross section of device-related events submitted by MedSun reporters during the period May 1 through May 30. All other reports can be searched under the ‘MedSun reports’ menu pane. Note: the two month delay is due to quality control and follow-up.



ORTHOPEDIC

Device:
Type: Shaver, Arthroscopic
Manufacturer: Stryker Endoscopy
Brand: Formula 180 Shaver Handpiece
Lot #: 375-708-500
Cat #: 375-708-500

Problem:

Following manufacturer recommended cleaning and sterilization procedures, an internal scope examination of multiple Stryker shavers revealed human tissue debris within the body section of the devices.

Comment from FDA: Please see FDA communication, “Ongoing Safety Review of Arthroscopic Shavers” online available: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/medsun/news/newsletter.cfm?news=38#1


Device:
Type: Hip Femoral Head
Manufacturer: Wright Medical Technology, Inc
Brand: Conserve Total A-class Head W/bfh Technology Short Neck
Model #: label stated # 36mm OD short neck
Lot #: label stated # 029798201
Cat #: label stated # 38AM-3604

Problem:

This was a planned total hip implant procedure. When the femoral head from Wright Medical was opened, it was observed that the wrong implant was in the package. Surgeon had requested a 36mm short. The package label stated 36mm OD Short Neck, and the inside label said the same thing, 36mm OD Short Neck. When the surgeon looked at the device on the sterile field, it did not look correct. It was observed that the actual stamped information on the implant itself said 36mm Long.


GENERAL & PLASTIC SURGERY

Device:
Type: Fixation Device
Manufacturer: Covidien United States Surgical Corporation
Brand: Protack
Cat #: 174006

Problem:

The ProTack instrument from Autosuture should never be used in closing hernia defect in and about the diaphragm. The use of the tacks on the diaphragm can result in episodic cardiac injury leading to cardiac tamponade and premature demise of a patient. This injury can be challenging to find and confound attempts to alleviate progressive decline of a patient's condition. This injury can occur months or years after they are placed. A surgeon's recommendation is to prohibit the future use of the ProTack Autosuture instrument in repairing diaphragmatic hiatal hernias.


Device 1:
Type: Prep Solution
Manufacturer: 3M Healthcare
Brand: Duraprep
Lot #: 2012-02AJ1
Cat #: 8630
Other #: NDC: 17518-011-08

Device 2:
Type: Drape, Antimicrobial
Manufacturer: 3M Healthcare
Brand: Ioban 2
Lot #: 2011-02 CD
Cat #: 6640EZ
Other #: 34-8701-3137-1

Device 3:
Type: Drape, Antimicrobial
Manufacturer: 3M Health Care

Problem:

Patient with intact skin underwent surgical prep with Duraprep for foot surgery. Ioban steri-drape placed over prepped skin. When steri-drape attempted to be removed after the initial portion of the procedure, patient's skin was blistered and peeled off with the steri-drape. When the foot was examined, additional areas on the foot blistered where Duraprep was used. The procedure was terminated at that point. 3M Remover Lotion was used by the surgeon to gently remove the remaining Duraprep from the patient's skin. Dorsal foot with denuded area. Large toe and sole of foot with denuded areas.


Device:
Type: Wound Therapy Device
Manufacturer: KCI
Brand: Infovac
Model #: 60090

Problem:

RN reports Wound Vac error message "container full", but container was not full. RN reset, error message then reading "container not engaged", troubleshooting guide followed, still reading "not engaged", container changed, no results, machine changed completely and began working immediately. No impact to patient. Device is a rented machine so will be returned to vendor for evaluation.


Device 1:
Type: Endoscopic Vein Harvesting System
Manufacturer: Terumo Cardiovascular Systems Corp.
Brand: Virtuosaph
Model #: Ref # MCVS550
Lot #: 8YK
Other #: PK # 8056741

Device 2:
Type: Endoscopic Vein Harvesting System
Manufacturer: Terumo Cardiovascular Systems Corp.
Brand: Virtuosaph
Model #: Ref # MCVS550
Lot #: 8YK
Other #: PK # 8055467

Problem:

When performing endoscopic vein harvesting, the bipolar/cutting feature was not working. An additional device was opened and also did not work. Third device worked properly - no patient harm.


Device:
Type: Surgical Blade
Manufacturer: BD
Brand: Bd Bard Protected Blade System
Model #: 11
Lot #: 8162706
Cat #: 373911

Problem:

Patient with a history of chronic pelvic pain was undergoing laparoscopy using an open technique - lysis of extensive adhesions in the left upper quadrant and right upper quadrant and lysis of pelvic adhesions. Physician went through three Bard-Parker # 11 blades before she could find one sharp enough to make an incision.


Device:
Type: Hemostatic Matrix
Manufacturer: Baxter International
Brand: Floseal Hemostatic Matrix
Lot #: HA081229
Problem:

Floseal was needed to control bleeding during surgical procedure. The product box is labeled STERILE with notation, "Unless opened or damaged, contents are sterile." Box containing items to be used is not sealed in sterile packaging.

Product box opened by circulating nurse and contents removed by sterile-gloved scrub tech and placed on the sterile field. Contents of the box included gelatin matrix component that is in a closed sterile package, 2 vials; human Thrombin and calcium chloride, and a 5ml syringe inside a sealed package.

The non-sterile packages removed from the box were placed onto the sterile field. The sterile field was unknowingly contaminated. The labeling on the box that noted that the contents were sterile unless opened or damaged confused the staff opening the product for use.
======================
Manufacturer response for hemostatic Matrix, Floseal Hemostatic Matrix
======================
The manufacturer has acknowledged the information and will be investigating.


RADIOLOGY


Device:
Type: Transducer, Ultrasound, Transesophageal Echocardiograph
Manufacturer: Toshiba America Medical Systems, Inc.
Brand: Toshiba
Model #: PET- 511BTM

Problem:

The patient was undergoing a Trans-Esophageal Echo Cardiogram when the TEE scope buckled.


Device:
Type: Mri, 1. 5t
Manufacturer: Philips Medical Systems
Brand: Philips Achieva

Problem:

IV nurse entered MRI suite and brought her metal IV cart halfway in doorway of MRI suite. The force of the MRI magnet caused the IV cart to lift up and it flew through the air, on to the MRI machine. Patient was lying outside of the MRI, on the MRI table at the time. The cart did not hit him. Another Nurse was on right side of patient and was looking for venous access in his right arm at the time of the incident. She was not injured. No injury for any individual - lots of potential for injury!

The IV nurse had been called to access the patient located in MRI. She did not realize the power of the MRI magnet especially when the patient was not in the MRI. She intended to leave the cart at the door, but should not have entered the room.

This is being reported not because of a device malfunction but as an alert of an incident regarding an MRI and need for better vigilance and perhaps education regarding the fact that MRI magnets are ALWAYS on and the need for better safety.

IV Cart in MRI suite

IV Cart in MRI suite


Device:
Type: Catheter, Brachytherapy, Intracavity, Icrt
Manufacturer: Best Vascular, Inc.
Brand: Novoste Betacath System
Lot #: Beta-Cath 3.5R System
Problem:

Novoste BetaCath System: the ICRT catheter was kinked so the Sr-90 radioactive train could not be sent back to the device. After the treatment was finished, MD tied to push the Sr-90 radioactive source back to the device; the source left the treatment area and was stuck in the catheter junction outside of the patient's body. MD said the catheter had a kink at the junction. The catheter was pulled out from the patient's body and dumped in a shielded transportation box with whole device. The patient and room was surveyed using radiation survey meter. No radiation was found inside the patient body and in the treatment room of the cardiac cath lab.


Device:
Type: Ct Scanner
Manufacturer: Siemens Medical Solutions USA, Inc.
Brand: Somatom Definition As +

Problem:

Operator attempted to tilt by 2 degrees the CT gantry to perform a sinus scan study. A ticking sound followed by a banging noise was heard prior to the CT cover cowl separating from the machine, falling on the patient's back and arm, and then falling to the floor. This occurred within approximately 14 seconds.

The gantry cover was not secured properly and it opened. The cover cowl was forced off by the moving gantry mass and it fell onto the patient.


MICROBIOLOGY

Device:
Type: Lab Reagent Kit
Manufacturer: Becton Dickinson
Brand: B-d Directigen Ez Kit For Rsv Rapid Antigen
Lot #: 8296111, 8330466, 8296126
Cat #: B694410
Problem:

Respiratory Syncytial Virus (RSV) rapid antigen assays were ordered on patients on two consecutive days. After thirteen results were reported positive using the Becton Dickinson Directigen EZ kits, the director was notified as well as the Medical Director of Hospital Epidemiology & Infection Control. RSV antigen assays were followed up with PCR and culture for confirmation. BD Technical services were notified and a conference call followed three days after the reported positive results. We had a total of seventeen positives and all seventeen were false positives.

BD Scientific group came to our lab ten days later following the conference call to run the samples to see if they could determine the problem. Lot numbers included 8296111, 8330466 and 829126. Awaiting BD's evaluation of the problem.

NEUROLOGY

Device:
Type: Electromyograph, Diagnostic
Manufacturer: Cadwell Laboratories
Brand: Cadwell Cascade
Model #: ES-5-100
Cat #: 190221-200
Problem:

30 minutes after starting the machine, a burning smell was noticed when attaching lead wires to the ES-5-100 stimulator box. The unit was immediately removed from service. The stimulator box was connected to a Cadwell Cascade Elite. Later investigation revealed the inside of the stimulator box was extensively burned and the case was partially melted. The concern here is this device was located under the surgical drape near the patient's foot and if a flame had protruded, it would set the drape on fire. There was a minor delay while the stimulator box was replaced. There was no harm to the patient.

There was a fairly large section of the circuit board that was charred and it is likely that there was a small flame inside the box at one point. Coincidentally, a second stimulator box in a different room failed on the same day, but not as extensively burned. According to the manufacturer, if the clinicians accidently plug the stimulator box into the wrong port (aux. port) on the main unit, this kind of burning failure is common. Apparently, if the clinician even plugs it in for a brief time, it will damage components and cause failure. The plugs are similar to standard computer VGA video connectors with one pin hole blocked and a corresponding missing pin on the plug to act as a key against misconnection. Close examination of the socket on the main unit revealed that the key plug was missing on one of the aux. connectors making it possible to misconnect easily. Clinicians report that it would be highly unlikely that the cable was misconnected. Another possibility is the cable that goes from the main unit to the stimulator may have been damaged by other equipment rolling over it, causing a short, but the cables tested OK.

Cadwell Cascade connections

Burned Circuit Board

Missing Key Plug

Stimulator Board

<br>Stimulator inside case


GENERAL HOSPITAL

Device:
Type: Iv Tubing
Manufacturer: B. Braun Medical, Inc.
Brand: Ultrasite
Model #: 470042
Lot #: 0061014595, 0061050990
Cat #: US1504HP
Other #: 4 from - Lot #: 0061014595 exp 08/11 and 2 from lot # 0061050990 02/12

Problem:

RN reports 6 sets of defective tubing over various dates. None of the tubing sets were actually used on patients. The issue was identified before use. In all cases fluid would not run through the tubing. All devices were retained and are available for evaluation purposes. There was no damage noted to the tubing upon visual inspection. Similar issues with this tubing have been experienced by our facility in the past, and previous inspections by B. Braun of the tubing from those reports indicated that there was too much solvent in the manifolds.


Device:
Type: Infusion Pump
Manufacturer: ALARIS Medical Systems, Inc.
Brand: 8100 Pump Module
Model #: 8100

Problem:

Upon powering up a module to begin an infusion on a patient, the module continually beeped “channel malfunction.” The channel malfunction warning indicator was on prior to connecting to the patient. A patient was not involved with this IV pump malfunction.

Biomed Assessment: The channel error alarm was caused by a defective upper pressure sensor. The sensor was replaced with a new one at no charge. After repair, the unit passed the manufacturer's test procedure and was returned to use.

Comment from FDA: See Alaris Recall online available at http://www.cardinalhealth.com/alaris/medical-device-recall/


Device:
Type: Pca Pump
Manufacturer: Hospira Global Medical Affairs
Brand: Gemstar
Model #: version 4.30

Problem:

The RN discovered that the PCA pump settings were different than what they had been previously when she checked on the patient. Staff was interviewed and had not changed the pump. Two staff members had been present when the new bag of Dilaudid was placed in the PCA pump, but the settings were different from what the RN discovered upon entry to the room. The patient denied altering the pump. However, a download of the pump showed that the settings had been changed twice - once earlier in the shift and then changed back to the MD's ordered settings and then again when the RN discovered different settings on the pump. The staff had also witnessed the patient manipulating the pump.

Anesthesia was notified and discontinued the PCA pump. The patient was changed to oral pain medications for pain control. Biomed checked the pump and it was functioning appropriately. It was deduced that the patient had learned the code to access the pump and was able to change the settings to bolus himself.

As a result of this event, the hospital changed the code from a 3 digit to a 5 digit non-sequential code. Education was provided to nursing staff so that when the code is entered into the machine, the RN turns the pump away from the patient's line of sight. The patient has since returned to the facility and has not been able to access the pump's settings.


Device:
Type: Iv Tubing
Manufacturer: ALARIS Medical Systems, Inc.
Brand: Alaris Burette Set
Lot #: 08035469
Cat #: 10779313

Problem:

A new TPN bag was primed for use, and was started on a patient. Approximately 4.5 hours later, the TPN was found on the floor near the pump. The pump did not alarm. The TPN tubing was attached to the patient with no visible air bubbles. When the tubing was removed from the pump, the TPN leaked from the part of the tubing that fits into top portion of the cassette.


Device:
Type: Pca Pump And Pcea Volume Infuser
Manufacturer: Hospira Global Medical Affairs
Brand: Gemstar
Model #: CE0050

Problem:

Our facility recently changed to a new lockbox product for PCA and PCEA (Patient Controlled Epidural Analgesia) volume infusers. Because of the new configuration of the lockbox, the way that the PCA and PCEA tubing is threaded through the Gemstar PCA and PCEA devices is different. If the IV tubing is threaded incorrectly, the tubing will be pinched and the device will not run properly. However, it is not intuitive to the nursing staff where the tubing should be threaded.

Staff suggested that it would be helpful if the Gemstar manufacturer could place an arrow with the phrase "thread here" on the device. In the interim, a sticker was created and placed on the device to indicate how it should be threaded. Additionally, staff has received education regarding how to thread the device with the new lock-boxes.

The manufacturer was notified and has provided a sticker template with directions for threading the device.


Device:
Type: Bottle, Collection, Vacuum
Manufacturer: Cardinal Health
Brand: Safe TCentesis
Lot #: L9C312F

Problem:

Safe T Centesis 6 fr. Cath drainage tray noted to have an air leak in the Y connector that connects tubing to drainage bottle – one way valve not working – no vacuum achieved.

Trays were pulled off the shelves, reported and returned to cardinal health. Cardinal health stated they had received no other reports of problems with this product. A new shipment was received 6 days later.

A second event occurred 9 days after first event. An air leak was again noted when the tubing was connected to the drainage bottle. The physician performing the procedure also noted that the scalpel in the tray was not sharp. Upon investigation, it was found that the new shipment was the same lot # that had previously been pulled and returned to cardinal health. Trays from the affected lot number have been pulled from all units pending further investigation.

Health professional's impression:
No adverse event reached the patient. The air leak was noted during the procedure and a different set was utilized.
======================
Manufacturer response for Chest tube system, Safe T Centesis
======================
States they have received no other reports of problems with the device.


Device:
Type: Syringe, Antistick
Manufacturer: Retractable Technologies
Brand: Vanishpoint Syringe
Cat #: 10301
Other #: 100301 (syringe - 3 cc)

Problem:

Retractable needle did not retract. This is not the first event of this happening with these needles. No patient injury.


Device:
Type: Infusion Pump
Manufacturer: ALARIS Medical Systems, Inc.
Brand: Medsystems Iii

Problem:

A patient was being transported into a medical helicopter for transport to another hospital, when the Alaris IV pump failed. The pump was infusing Norepinephrine when it "locked out," and would not pump. The pump was reporting "channel not available.” Another channel was selected, and the pump again indicated "channel not available.” The patient was quickly brought back into the hospital’s ICU, placed on a different IV pump, and reassessed. After clearance, the patient was taken to the helicopter and transferred to another hospital. The pump in question was immediately removed from service and sent to our Clinical Engineering department for analysis.


Device:
Type: Mattress, Flotation Therapy, Non-powered
Manufacturer: Kinetic Concepts, Inc.
Brand: First Step Select

Problem:

The elderly patient was identified as being at high risk for development of pressure ulcers and so as part of the pressure ulcer prevention protocol, the patient was placed on a First Step Select overlay mattress. Three days after placement of the mattress, a nurse was in the hallway outside of the patient's room and heard a noise come from the patient room. The nurse found the patient lying on the floor. The patient reported; was "trying to use the urinal and slid out of the bed onto the floor." The patient had no apparent injuries from the fall.

The overlay mattress has an air-filled chamber portion and a cover. Both of these items are made of a nylon-like material. This combination of slippery surfaces does have a "slip and slide" effect. The mattress overlay base is to be strapped to the frame of the bed by Velcro straps and the cover is to be tucked under the air mattress. Care is also taken to use the Hill Rom 840 bed instead of the Hill Rom Versa Care bed as the Versa Care beds do not have enough height to the side rails and a patient could easily go over the side rails due to the thickness of the overlay mattress. In this particular case, the patient was on a 840 bed and the nurse discovered that the mattress overlay was not anchored to the bed frame and the cover was not tucked in.

It is normally the practice for KCI to set up these mattresses, but not always. KCI will be contacted regarding this problem. Staff on the nursing unit on which this occurred noted that they have been using these mattresses for years and this is the first time they have encountered an overlay mattress that was not anchored properly to the bed frame.


Device 1:
Type: Iv Set
Manufacturer: Carmel Pharma Inc.
Brand: Phaseal Injector Luer Lock
Cat #: N31

Device 2:
Type: Iv Set
Manufacturer: Carmel Pharma Inc.
Brand: Phaseal Connector Luer Lock
Cat #: C40

Problem:

The patient bumped his infusion pump while walking through a doorway in the patient lounge. This caused the PhaSeal system to loosen because it extends out slightly from the infusion pump. Small drips of chemotherapy dripped on the floor, and some nurses noticed a small pool of chemotherapy when the patient stopped walking. A chemotherapy spill kit was obtained, and housekeeping was notified. The spill was contained. Patients and families were notified to stay out of the lounge and the small hallway. There was no patient injury. The PhaSeal Connector sticks out from the side of the pump and has contributed to disconnections and chemo spills in the past.


Device:
Type: Infusor, Pressure, For I. V. Bags
Manufacturer: Vital Signs
Brand: Infusable Pressure Infusor
Other #: Ref # IN8000

Problem:

Pressure bag pumped up to correct pressure; netting holding 500 cc bag ripped along side of bag. Pressure bag had to be replaced. No patient harm.

Additional information obtained from the site:
There was no patient harm, they just switched bags. There have been 3 more similar incidents before. The packaging was not saved, so there is no lot number available.

Device:
Type: Catheter, Intravascular, Therapeutic, Short-term
Manufacturer: Becton Dickinson
Brand: Bd Insyte Autoguard
Model #: 18 ga 1.16 in, 1.3 X 30mm
Lot #: 7340858
Cat #: ref 381444

Problem:
Nurse was attempting to place an 18 gauge intravenous catheter (IV start) into a Labor & Delivery patient's vein in left forearm. Nurse entered the patient's vein with the catheter and when attempting to advance the catheter, the hub broke off from the plastic angio. Fortunately the nurse was able to remove the catheter. Potential for detached catheter to have entered patient's vein.

Device:
Type: Tubing, Infusion
Manufacturer: ALARIS Medical Systems, Inc.
Brand: Alaris Se Pump Administration Set
Cat #: C72106M

Problem:

The infusion pump was programmed to infuse at 66 ml per hour. The potassium chloride infused the total amount in 15 minutes. Upon biomedical department investigation, the backflow preventer on the primary tubing set was not operating correctly. Approximately 6.5 ml of the potassium chloride secondary infusion was delivered to the patient, and the remaining fluid emptied into the primary bag.


ANESTHESIOLOGY

Device:
Type: Endotracheal Tube
Manufacturer: Mallinckrodt
Brand: Intermediate Hi-lo
Lot #: 081200 1153
Cat #: 86450

Problem:

Anesthesiologist intubated a patient for cardiac surgery and while taping the endotracheal tube in place outside the patient's mouth he noted an airway leak. He added more air via syringe to the tube cuff but again noted an airway leak. When he went to examine the pilot balloon (a small, in-line balloon that shows that the cuff/tubing/luer valve system is pressurized), he noticed that it had become disconnected from the cuff tubing. Close examination of the cut end shows repeated striations to the material like what would be obtained with a sawing motion.


Device:
Type: Anesthesia Breathing Circuit
Manufacturer: Medline Industries
Brand: Circuit, Adult Anesthesia
Model #: DYNJAA9832D
Lot #: 08LB2281
Cat #: DYNJAA9832D

Problem:

During the induction phase prior to surgery, the Anesthesiologist noticed that the 90 degree respiratory gas sample elbow "popped" off the parallel “wye” section of the patient circuit. The doctor tried to reattach the elbow, but the fit was very loose. The doctor obtained a different style sample gas elbow and it secured properly. The doctor noted this same problem with several patient circuits and notified the manufacturer to exchange all the patient circuits with this model and lot number.
======================
Manufacturer response for Anesthesia breathing circuit, Circuit, Adult Anesthesia
======================
Manufacturer replaced all the patient circuits with different circuit set.


Device:
Type: Anesthesia Unit
Manufacturer: GE Medical Systems, LLC
Brand: Datex Ohmeda S/5
Model #: AAUFO2

Problem:

Gas monitoring problems detected. On the display showing gas information, the displayed numbers went blank intermittently. No patient harm.

Biomedical engineering was contacted and worked with the manufacturer to perform a factory reset. The reset involved clearing the anesthesia unit's memory, then reloading all the default information and checking for proper operation. The reset did not correct the problem, as it recurred as an intermittent problem thirteen days later. At that time, the factory service technician, working with our Biomed department, replaced the display and the circuit board which feeds the gas information to the display. The problem has not returned since.


Device:
Type: Universal Neonatal Heated Ventilator Breathing Circuit
Manufacturer: Teleflex Medical Incorporated
Brand: Hudson Rci
Model #: Ref. 780-09
Lot #: 02324
Cat #: 02L08
Other #: Humidifier Unit, Model: concha Therm

Problem:

The far end of the inspiratory breathing limb was missing an additional temperature probe Tee connector. The second temperature probe was connected to the Y port. The inspiratory breathing limb touched the expiratory limb. Because the two limbs touched, the inspiratory limb melted and a leaky hole was created in the circuit. The expiratory limb also melted but did not have a hole.

The breathing circuit should be equipped with a temperature probe away from the Y connector. By connecting the temperature feedback at the Y connector, it can cause the heater to keep heating, since the feedback temperature at the Y always fluctuates due to the temperature change every time the patient breathes in and out.
======================
Manufacturer response for Universal Neonatal Heated Ventilator Breathing Circuit, Hudson RCI
======================
Recommended to replace with a different kind of circuit.

Device 1:
Type: Ventilator
Manufacturer: Newport Medical
Brand: E-500
Model #: E-500

Device 2:
Type: Ventilator
Manufacturer: NEWPORT MEDICAL
Brand: E-500
Model #: E-500

Device 3:
Type: Ventilator
Manufacturer: Newport Medical Instruments, Inc.
Brand: E-500
Model #: E-500
Problem:

Biomed Technician verified ventilator did not have an audio alarm. This happens when the ventilator is connected to our nurse call system. When the ventilator cable is plugged to the nurse call system, the nurse call system alarms and is unable to be reset. To reset the nurse call system, the biomed technician has to unplug the ventilator cable from the nurse call system. The ventilator-nurse call alarm board is locked and the ventilator has to be turned off. The RT has to bag the patient twice for 4 minutes to allow the ventilator to reset while it's off. The ventilator is then powered on again and subsequently reattached to the nurse call system.


CARDIOVASCULAR


Device:
Type: Catheter, Percutaneous, Slitter
Manufacturer: Medtronic Inc.
Brand: Attain Command
Model #: 6230UNI

Problem:

The new disposable Medtronic slitter pulled out the lead wire from the coronary sinus during a lead implantation. This event has happened two other times with other physicians since hospital started using this product this year. A new kit was used and the lead was successfully implanted. There was no patient injury identified. The manufacturer replaced the newly disposable slitters with the original universal slitters and will provide additional training. It was noted by the healthcare provider that the slitter handle in the new product is larger than the previous model.


Device:
Type: Ablation Console
Manufacturer: Cryocath Technologies, Inc.
Brand: Ablation Console
Model #: 10000003

Problem:

Attempted to use cryoabation on Ventricular Tachycardiac (VT) pt. System gave high refrigerant flow error. Cryocath umbilical cord and catheter exchanged, troubleshooting and error message remained. Unit remained on for a total of approx 25 minutes and continued to give error message when ablation was attempted. Unit was replaced with cryocath unit from Peds Dept and ablation was successfully administered after approx 1 min of warm-up time. Case delayed approx 30 minutes.
======================
Manufacturer response for Ablation unit, Ablation unit
======================
Device Tested by vendor but could not reproduce problem. Replaced the Valve board and the proportional valve responsible for refrigerant delivery in case there is an intermittent problem. Console is working properly and no problems detected.

Device:
Type: Electrode, Ecg, Monitoring
Manufacturer: 3M Healthcare
Brand: Red Dot
Model #: 2560
Lot #: 2009-09 CU
Cat #: 2560

Problem:

Nursing staff reported difficulty in obtaining good clear ECG signal including heart and respiration rates. Biomed discovered that the electrodes in use were very dry. Further investigation of unopened packages in the warehouse led to the suspicion of a faulty lot of electrodes from the manufacturer. The electrodes were not expired, and even unopened packs contained electrodes that were very dry and peeled away from the backing easily.

Sales representative from 3M came on the next business day and confirmed that the electrodes were defective but this was not a problem that 3M had been aware of previously. 3M provided replacement product.

Device:
Type: Lead, Pacemaker, Temporary
Manufacturer: Ethicon Inc.
Brand: Temporary Pacing Wire
Cat #: TPW 50
Problem:

The patient had a CABG X 4 and pacer wires were placed at the time. On post-op day 6 in anticipation of discharging the patient, the Physician Assistant (PA) pulled the wires. The patient appeared to seize followed by a full code arrest. CPR was administered and the chest was opened in the room. A rhythm was re-established and the patient was taken to the OR. The patient was placed on bypass. A laceration of the right coronary vein graft was repaired. The graft was lacerated by a clip that was holding the pacer wire to the right atrium. The patient was discharged to an acute rehab facility post-op day 17. He had problems with speech. He was impulsive and required verbal cues to respond to commands appropriately.


GASTROENTEOLOGY & UROLOGY


Device:
Type: Lap-band, Gastric
Manufacturer: Allergan, Inc.
Brand: Aps Access Port Lap-band System
Model #: B-2240

Problem:

Patient required replacement of lap band due to leakage in the actual band portion, which inflates with saline to resemble an inner tube. Upon removal of entire system, surgeon examined the band again by putting fluid into the system. There was an obvious leak coming from the band. The band was placed approximately 21 months ago.

Device:
Type: Valve, Stopcock, Nasogastric
Manufacturer: Bard Medical Division
Brand: Lopez Valve-adult
Lot #: NGSL0041
Cat #: 0056000

Problem:

Nursing staff report that they are experiencing a possible misconnection with the Adult Lopez Valve and nasogastric (NG) tubing. The universal adaptor on the Lopez Valve is designed to fit onto side "A" of the valve device. However, it easily fits onto side "B" as well. If the adaptor is placed on side B (the wrong side), then staff report that the NG tube leaks because it does not have as secure of a fitting due to the tapered side B tip. Instructions on the device packaging are clear.

Staff feels that the device could be improved if the universal adapter was not removable or if it was a different size such that a possible misconnection could not be made.
======================
Manufacturer response for NG valve, Lopez valve--adult
======================
Information contained in this report has been faxed to the manufacturer.


Device:
Type: Nasogastric Tube
Manufacturer: Covidien Kendall
Brand: Kendall Argyle Salem Sump
Model #: 8888264986

Problem:

After inserting the nasogastric tube, the nurse attempted to check the placement. The nurse was unable to do so. The nasogastric tube was removed and it was discovered there were no openings at the distal end of the tube. A new nasogastric tube was then inserted.


Device:
Type: Dialyzer, Hemodialysis
Manufacturer: Fresenius Medical Care North America
Lot #: 8NU803
Cat #: F180NR

Problem:

Small amount of blood found on the floor. Blood found leaking from venous end of dialyzer. Treatment paused and only arterial blood returned. Approximately 200 cc of normal saline with blood was lost. New system was set up, and treatment was resumed after ten minutes.


Device:
Type: Catheter, Retention, Barium Enema With Bag
Manufacturer: Medline
Brand: Bag, Enema Flip Top Dry
Cat #: DYND70102

Problem:

Patient told Risk Management gave himself the enema. He left the cap on but when he pulled it out, the cap was not on and he told the nurse this. The patient was provided the supplies and instructions. We discovered that the cap was left inside the patient. Cap was extracted in Medical Procedure unit (MPU).

We had a similar report earlier this year.

All of Enema kits from Medline packaging lacks instructions to remove the (dark blue) end cap prior to insertion. We have switched our products from Medline ( DYND70102) to MediChoice (ENBG1500), the product from MediChoice has clearer directions.
======================
Manufacturer response for bag, enema flip top dry, bag, enema flip top dry
======================
Working on adding instructions to remove the (dark blue) end cap prior to insertion.

Device:
Type: Catheter, Nephrostomy
Manufacturer: Cook Medical, Inc.
Brand: Ultrathane Mp Mac. Loc Rb 8. 5f 38 25cm 6sh Percutaneous Nephrostomy Tube
Lot #: 2260385
Problem:

A Cook 8.5 French percutaneous nephrostomy catheter was placed successfully. A month later, the patient who was 31 5/7 weeks pregnant experienced increased flank pain and the tube was not draining well. The catheter hub was cut and extensive attempts were made to pass two different wire guides through the tube. Wire passage was not possible secondary to complete obstruction. Attempts were made to remove without a wire. However, the pigtail was completely locked in place. The pigtail was unable to be unlocked. A sheath was placed and the nephrostomy tube straightened. However, the tube would not pass completely into the sheath. Approximately 7mm of the distal tip would not pass into the sheath.

A density consistent with sediment/calcification was noted at the catheter tip. Despite numerous attempts the catheter tip was unable to be removed, and upon final traction with the use of a hemostat the tip of the catheter broke and remained in the retroperitoneal tissue. The plan is to remove the broken tip of the catheter laparoscopically after the delivery of the infant.


EAR, NOSE & THROAT

Device:
Type: Cochlear Implant
Manufacturer: Cochlear Americas
Brand: Nucleus Freedom
Model #: CI24RE
Problem:

Patient is an adolescent with bilateral cochlear implants. His right side was implanted when he was a toddler. His left side was implanted ~3 years ago.

More recently, his older right implant has been functioning poorly for him. It does not provide him with useful sound and makes distracting noises. He notes that last fall, he began noticing a "bumblebee" bubbling sound in the right cochlear implant. In spite of adjustments to his MAPS, this would not go away. He had his processor upgraded on the right and this did not improve things. He now feels that he cannot tolerate this. He wears his processor so that he does not get in any trouble, but leaves it turned off.

His parents are very frustrated with this. They want him to benefit from both implants. They also feel that there has been a recent decline in his performance in school, which they think might be due to his right implant not being utilized. He underwent integrity tests by Cochlear Corporation ~5 months ago. The test was interpreted as normal.

Recently, the patient came in to have his existing right side device explanted, and replaced with a new one. The device was explanted as defective and sent to Biomedical Engineering.