Highlighted Reports
MedSun: Newsletter #57, February 2011
This section contains a sample of reports from all the MedSun reports received during a particular period. The reports were submitted by MedSun Representatives. In some instances the reports have been summarized and/or edited for clarity. The entries that follow represent a cross section of device-related events submitted by MedSun reporters during the period November 1 through November 30 2010. All other reports can be searched under the ‘MedSun reports’ menu pane. Note: the two month delay is due to quality control and follow-up.
CARDIOVASCULAR
Device:
Type: External Defibrillator
Manufacturer: Physio Control, Inc.
Brand: Lifepak 20
Model#: LP20
Problem:
Patient had an AED defibrillator on a crash cart in the room as a secondary monitoring device (the defibrillator was powered off). Charge nurse went into the room to check the device. Patient was fully hooked up to AED defibrillator, i.e., pacer hands free defib pads attached to the patient, along with the EKG monitoring leads. Device was powered off (as originally intended) when the charge nurse went into the room. Device was turned on to check working condition. After the device (defibrillator) was powered up it delivered a shock (discharge from the defibrillator) to the patient. Device was found to be unplugged from electrical outlet. The nurse believed this appears to be the reason it (defibrillator) was off (powered off). Analysis of the log indicated that no problems were found with the device. Device recording data indicated the nurse had pressed the charge and discharge button, causing the defibrillator to discharge the patient.
Device:
Type: Opti QSxo2/cco Catheter
Manufacturer: Hospira, Inc.
Brand: Opti QSvo2/cco Catheter
Cat #: List # 52511-13
Problem:
Inconsistent Cardiac Output and Cardiac Input numbers. It does not correlate with patient's clinical picture.
======================
Health Professional's Impression
======================
Manufacturer response for Opti Q SxO2/CCO Catheter, Opti Q SvO2/CCO Catheter
======================
We had many discussions with the manufacture's representative. We followed their recommendations and we continue to have product malfunctions. This interferes with patient care.
Device:
Type: Telemetry Transmitter, Physiological
Manufacturer: Philips Medical Systems
Brand: Intellivue
Model#: M2601B
Cat #: M2601B
Other #: 453564007681, ABA037AAMS03
Problem:
RN sent Philips telemetry pack to Biomed because it persisted in displaying "low battery." Biomed changed AA batteries and got the same issue of "low battery" within 1-2 minutes. Tech opened battery case and on removing one of the two batteries found it to be extremely hot to touch. On further investigation it was found that the negative pole spring (the battery towards the out side of the case) had slipped over the top of retaining clip and come into contact with wire from the battery closer to the touch pad. The load on outside battery was greatly increased and overheated. The battery indicator was displaying "low battery" on the unit's screen. In a simulated situation it was demonstrated that the outside battery was heating up to ~ 196.5 F degrees. The plastic retaining clip "grinds down" over time. There is no apparent way to replace the clip. Patients have complained the pack gets hot but we had never discovered the cause until now.
Health Professional's Impression: Wear and tear on a plastic retaining clip that allows a negative pole battery spring to slip up over top of clip and come into contact with second AA battery. This is a subtle problem that nursing staff cannot easily visualize or fix.
See device images:

CLINICAL CHEMISTRY
Device:
Type: Analyzer, Chemistry,
Manufacturer: Beckman Coulter
Brand: Dxi 800
Model#: 973100
Cat #: 448229 AB
Problem:
Provider called the lab with a request to retest TSH samples on a few patients who had lower values than expected. After retesting the values were significantly higher. The lab did some troubleshooting and found that when the Lithium heparin tubes are inverted prior to sampling results for TSH, the results were significantly lower.
======================
Health Professional's Impression
======================
There is an interferent in the Lithium heparin tube that is causing falsely decreased values for TSH on the Beckman DXI.
EAR, NOSE, & THROAT
Device:
Type: Tube, Tympanic
Manufacturer: Gyrus ACMI, Inc.
Brand: Micron Titanium Bobbin Ventilation Tube
Model/Cat#: 14-5272
Lot #: 0625683013
Problem:
Patient had previous ossicular reconstructive surgery. After about two months, patient unable to hear. Physician brought patient to Operating Room to do another ossicular reconstruction and noted during surgery the prosthesis from the initial surgery was broken. Prosthesis removed and a new one implanted.
GENERAL & PLASTIC SURGERY
Device:
Type: Applier, Clip
Manufacturer: Ethicon Endo Surgery
Brand: Ligamax 5
Model/Cat#: EL5ML
Lot #: G4TU6E
Problem:
Surgeon was performing a laparoscopic cholecystectomy with intraoperative cholangiogram. When surgeon applied the first clip to the patient's artery, the Ligamax 5 mm Endoscopic Multiple Clip Applier would not release after firing. The clip applier was jammed and the device would not revert back to the open position. The concern was that the artery could have been damaged. The only other option would have been to open the abdomen and try to free the device. However, the surgeon was able to safely remove the clip off the artery with the clip applier still in the closed position. No injury was done to the patient, but this put the patient at risk and the surgeon had to make a decision to try to work the device off or open the patient up to try to remove it. This posed a significant issue during the procedure. Another clipper was used from the same lot number without a problem.
======================
Health Professional's Impression
======================
Device would not release after clipping artery.
Device:
Type: Endo Clip
Manufacturer: Covidien/US Surgical
Brand: Autosuture Endo Clip
Model#: REF 176630
Lot #: N0H0411
Problem:
Child admitted with lesion in his left lower chest. Child underwent a video assisted thorascopic surgery. During the procedure, after two clips had been placed on an artery, the clip applier jammed during the placement of the third clip. The surgeon replaced clip applier with another to place the third clip and then performed a partial transection of the artery. The first two clips; however had not closed the artery and the bleeding necessitated conversion of the thorascopic procedure to an open chest procedure.
Device:
Type: Instrument, Surgical
Manufacturer: Spineology Inc
Brand: Spineology EShaper Blade
Model#: 315-0014
Problem:
During a lumbar laminectomy and fusion L4-5, L5-S1 the tip of the Spineology Shaver blade broke off in the L5 disc space.
======================
Health Professional's Impression
======================
The doctor did not express dissatisfaction with the shaver blade.
======================
Manufacturer response for Sterile, single-use, surgical instrument subcomponent intended to be used in conjunction with the 1500E OptiMesh Implant System. This device is designed to aid in cavity creation prior to introduction of the 1500E OptiMesh Implant, Spineology E Shaper Blade
======================
Manufacturer requested broken instrument be returned. We have not returned to date. Apologized for their representative not notifying hospital central sterile staff that this instrument recently changed to single use one month prior to this event date. It is not known if this single use instrument was used prior to this event date.
Device:
Type: Tip Cover Accessory
Manufacturer: Intuitive Surgical
Brand: Davinci Tip Cover Accessory
Lot #: C10216
Cat #: 400180
Other #: Tip cover accessory made by Microtek Medical for the DaVinci systems
Problem:
The daVinci tip cover accessory had holes in it. This was discovered prior to use, and the lot was pulled from the shelf and returned to the manufacturer for credit. Had it been used, the potential existed for there to be arcing.
Device:
Type: Tray, Surgical
Manufacturer: Total Pack Management
Brand: Ir Cath Tray
Lot #: T005422
Cat #: TPM-IRCHFB-1
Other #: Manufacturer date: 10/2010
Problem:
Patient in room waiting for procedure for IVF filter. The first tray was opened and a "mucous looking particle" was observed in the small bowl; second tray opened and "plastic flecks" was observed in small bowl. There was a "metallic-like particle" observed in the large bowl (guide wire bowl). The procedure was delayed due to repeated set-ups. Clinicians pulled needed supplies by hand to complete the case.
Device:
Type: Tray, Surgical
Manufacturer: Medline
Brand: Medline Sterile Laceration Tray
Lot #: 10IB4116
Cat #: DYNDL1228
Problem:
We are reporting a total of 10 devices of the same lot and from the same manufacturer with this one report. MD reports that he opened a laceration tray and had difficulty opening and closing the hemostat needle holder in the kit. Another tray was opened and the same difficulty was encountered. There appears to be nothing wrong with the device visually. There are no problems with any of the other contents of the tray. A total of 10 laceration trays were found with this same problem. All were from the same lot number. The 11th tray that was opened had no difficulties. As expected, the hemostat needle holder opened and closed smoothly without difficulty. The 11th kit also was from the same lot number. None of the affected devices came into contact with the patient and there was no patient harm involved in this event.
======================
Health Professional's Impression
======================
The staff feels that the bolt/screw used to keep the two pieces of the device together may have been put on too tightly.
======================
Manufacturer response for Needle holder hemostat, Medline sterile Laceration Tray
======================
Manufacturer was contacted. We are currently awaiting return instructions and plan to send the device back to the manufacturer for their inspection and analysis.
GENERAL HOSPITAL
Device:
Type: Catheter, Picc
Manufacturer: BD Medical Systems
Brand: L-cath
Cat #: 384539
Problem:
PICC found to be leaking at catheter hub site. Clinician attempted to remove line at that time. Resistance felt and warm compress applied to patient for one hour. The second attempt caused PICC line to snap leaving 1cm above skin line. The third attempt caused the second line to snap leaving a section inside the patient. The remaining line was surgically retrieved the next day. Patient stable.
======================
Health Professional's Impression
======================
Device failed to pull out with minimal force.
======================
Manufacturer response for Peripherally Inserted Central Catheter (PICC), L-Cath
======================
Manufacturer provided Ref#58290 and sent pre-packaged mailer to have PICC returned for evaluation.
Device:
Type: Catheter, Picc
Manufacturer: Medical Components, Inc
Brand: Vascu-picc
Lot #: MBBK 950
Other #: Ref # MRVSP41003
Problem:
Leaking noted at the PICC catheter site. Small tear noted above hub on the clear part of the tubing. PICC removed.
Device:
Type: Filter, Infusion Line
Manufacturer: Pall Medical
Brand: Intravenous Total Nutrient Admixture Filter
Lot #: 016611
Problem:
TPN (total parenteral nutrition) filter and TPN tubing broke. The TPN had to be discontinued and surgeon called for another type of replacement fluids. Registered Dietician and Pharmacy states this has been happening frequently throughout the hospital, however not all are being reported. Pharmacy states that the filter and tubing are attached to the TPN under the hood and when everything is taken out of the hood the filter becomes very loose and requires tape to hold it secure as a "work around".
======================
Health Professional's Impression
======================
A possible faulty TPN tubing/filter. This is the 3rd of 4 reported and I've been told this is continually happening and not just in one area or by one Pharmacy tech connecting the tubing/filter to the TPN.
Device:
Type: Scale, Infant
Manufacturer: Cardinal Health
Brand: Detecto Scale
Model#: 6745
Cat #: 62450-300
Other #: PO# 10263100
Problem:
Baby Scale readings inaccurate by 5-10% wt differences. Biomed has calibrated and attempted to remedy the issue, and when the scale reads their test using a 10 pound weight it reads accurately, but when weighing a moving infant the weights are consistently off by 5-10% of the actual weight.
======================
Health Professional's Impression
======================
Scale does not read accurately and consistently especially with infant moving intermittently.
Device:
Type: Warmer, Infant Radiant
Manufacturer: Draeger
Brand: Babytherm 8004
Model#: Babytherm 8004
Problem:
Nursing reported the following: temperature probe connected to patient and bed and temp settings set to 36.5. Open warmer showed patient's temperature at 37.3, but still within normal limits. Heart rate was high so temperature was checked rectally which showed a temp of 39.0. Patient's temperature should have been regulated correctly with open warmer being connected to patient through temperature probe. The infant was removed and placed in another warmer; monitored vital signs closely.
It was determined that this overheating was caused by a known problem referred to as the Stem Effect. (Technical information available here http://bit.ly/eb6nZI)
Our facility’s current policy is to coil the temperature probe to avoid the effects of the Stem effect.
Our policy is as follows:
Thermoregulation for all patients admitted into radiant warmers and incubators:
1. Warm bed prior to admission. See CPG table Neutral thermal environment for temperature range to select. (This requires weight and age of patient)
2. Select proper skin temp probe.
3. For low-birthweight patients and those with fragile skin, place a barrier of duoderm between the temp probe cover and patient skin.
a) Cut the duoderm into a circle. Cut out the center portion. Place this against the skin.
b) Place the temp probe into the center of the duoderm against the skin.
c) Make a few coils of the temp probe wiring under the temp probe cover, and place this cover over the duoderm.
4. For patients without specialized skin care needs, place the temperature probe and coils of wiring on the sticky surface of the temperature probe cover and place this on the patient's body.
5. Ensure the incubator/ radiant warmer is set to skin control or servo control, rather than air temperature or manual.
6. Adjust control temperature to maintain baby within normal range. Measure axillary temperatures with cares.
Coiling prevents inadvertent cooling of the temperature probe tip, and helps to maintain a consistent temperature. Improper technique in placing skin probe can cause inadvertent over heating of the patient.
We originally found out about the stem effect in 1993. Manufacturer was involved at that time.
OBSTETRICS/GYNECOLOGY
Device:
Type: Clamp, Circumcision
Manufacturer: Cardinal Health
Brand: Gomco Circumcision Clamp
Cat #: 02-01-0500
Problem:
During circumcision with Gomco, it became difficult to remove the foreskin from the penis. This caused a splitting tear at the foreskin incision junction requiring six sutures to close.
PHYSICAL MEDICINE
Device:
Type: Heel Warmer, Infant
Manufacturer: Cardinal Health
Brand: Infant Heel Warmer With Tape
Cat #: 11460-010T
Other #: (01) 0 0630140 01739 4
Problem:
I was made aware of a medical device failure involving an infant warmer in the Family Care Center. The warmer's heating element had failed due to the introduction of a liquid while the element was heated. The introduction of the liquid to the heated element caused the liquid to instantly flash to the surrounding reflector and down upon the infant. Upon investigation with the Director of the Family Care Center, it was determined that the liquid was a chemical compound used in a product called "Infant Heel Warmer with Tape" and is manufactured by Cardinal Health. The Director provided the incident report that detailed the event with the following comments: "Heel warmer pack squeezed per instruction – heel pack popped open causing a 1 cm chemical burn on infant's abdomen. The chemical flew up into heating element of the infant warmer causing damage to the warmer". The Director went on to explain that the baby was under an oxygen hood, which covered most of the body. Without the hood in place the infant most likely would have been burned much more severely from head to abdomen. The incident will be reported to the FDA.
======================
Health Professional's Impression
======================
The health professional's comment was that they followed the instructions on the package. They feel the product and/or instructions for use are designed poorly.
Device:
Type: Heel Warmer, Infant
Manufacturer: DeRoyal
Cat #: HNICU-100
Problem:
Heel warmer was activated by phlebotomist. The device is hard to activate and requires a significant amount of pressure to do so.
The warmer ruptured, splashing the phlebotomist in the face and forearm. She experienced a burning sensation. She immediately flushed the affected areas with water and then went to the Emergency Room for care. The employee was noted to have erythema to her left cheek and minimal erythema to her forearm. She was sent home by the ED physician to shower with unscented soap and instructed to treat the areas with Aloe Vera or antibiotic ointment.
RADIOLOGY
Device:
Type: X-ray System, Rad/fluoro
Manufacturer: Siemens Medical Solutions USA, Inc.
Brand: Sireskop Sd
Model#: 8890415
Problem:
The footboard failed while the patient was at a 50 degree angle while undergoing a GI series. The patient slid (fell) to the floor, sustaining no injury. The footboard is rated for 400 lbs. BioMed tested the footboard with maximum weight load rating (400lbs) with calibrated weights to 90 degrees in various footboard positions and wiggling the footboard with weight applied. Footboard held without releasing.
The footboard latches may not have been completely engaged. There are green lines visible to the radiology staff which should provide a visible check that the latches are locked. These green lines are not too easy to see when a patient is on the table in various positions. There is a slight audible click heard when properly latched.
See device images:

~~~
Special Note: The lollipop icon distinguishes highlighted reports that describe medical device events involving neonatal or pediatric patients, or those events involving a medical device that is indicated for use in neonatal and pediatric patient populations. FDA defines pediatric patients as less than 21 years of age.