St. Jude Medical, Riata and Riata ST Silicone Endocardial Defibrillation Leads: Class 1 Recall - Failures with Lead Insulation
MedSun: Newsletter #68, January 2012
FDA MedWatch Safety Alert
FDA notified healthcare professionals of a Class I Recall of the St. Jude Medical Riata and Riata ST Silicone Endocardial Defibrillation Leads. The silicone insulation covering these defibrillation leads is at risk of premature abrasion. When abrasion occurs, the conductors inside the leads can come out (externalized) of the insulation. Leads with externalized conductors may develop electrical dysfunction and not work as intended. In the event the device does not work as intended, should a life-threatening heart rhythm occur, pacing or defibrillation therapy may not be delivered as intended. This may result in serious adverse events, including death.
The recall includes the following model numbers:
•Riata (8F) Silicone Endocardial Defibrillation Leads
Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592
•Riata ST (7Fr) Silicone Endocardial Defibrillation Leads
Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042
Additional Information:
FDA MedWatch Safety Alert. St. Jude Medical, Riata and Riata ST Silicone Endocardial Defibrillation Leads: Class 1 Recall - Failures with Lead Insulation. December 21, 2011.