Medical Device Problem Summaries
Here you will find medical device problems summarized from reports received by MedSun hospitals.
Medical Device Problem Summaries
Summary of MedSun Reports Describing Adverse Events With Vascular Hemostasis Devices |
Print Item E–mail Item |
Vascular hemostasis devices are used to stop bleeding from a blood vessel, usually after a medical procedure has opened an artery. During cardiac catheterization, balloon angioplasty or catheter-based procedures, the physician often will obtain access for the procedure through the femoral artery in the patient’s groin. The physician may also use the arteries in the arm, such as the radial or brachial artery. There are several types of hemostatic devices. Some administer an agent that speeds the body’s natural clotting process and stops the bleeding, while others involve a stitch in the femoral artery to close the incision site and stop the bleeding. [1]
Over the past year, MedSun has received 28 adverse event reports associated with vascular hemostasis devices [2]. The reports represent four manufacturers: 12 from Abbott Vascular, 9 from St. Jude Medical, Inc., 6 from AccessClosure, Inc., and 1 from Cardiva Medical Inc. The reports were submitted by 15 hospitals between June 5, 2008 and June 5, 2009. The reported device problems included:
• General Device Failure (8)
• Failure to deploy or improper deployment (4)
• Device did not hold (2)
• Sheath would not split (2)
• Suture material missing (2)
• Balloon burst (1)
• Device lodged in vessel (1)
• Tip broke off (1)
During this time period there were no reported deaths associated with these devices. The patient injuries listed below were reported in 15 of these 28 reports. Injuries reported include:
• Hematoma (7)
• Bleeding (4)
• Surgery (4)
Of the reports that listed patient age, none had a patient age listed as less than 21 years and 30 patient ages were listed as greater than 21 years. Of the reports that listed patient gender, a total of 15 female patients and a total of 13 male patients were involved. Reports may have involved more that one patient.
These MedSun reports contributed to FDA awareness of the device problems. The following recall, describing problems with failure of hemostasis are noted to be associated with the StarClose device manufactured by Abbott Vascular. Reported events involving specific StarClose devices may, or may not, be involved in the recall listed.
Recall # Z-0575-2007
A recall was initiated on January 31, 2007 by Abbott Vascular Inc. for the
StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677, Lot Numbers starting from 43034-6H to 47139-6H
The reason for the recall was premature release of the vessel locator wings, which stabilize the device prior to clip deployment, will result in no hemostasis. Recall online available,
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ID=50232
Other FDA Articles and Notices of interest discussing Vascular Hemostatis Devices authored by FDA staff.
FDA Patient Safety News
Complications Related to the Use of Vascular Hemostasis Devices, online available: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/psn/printer.cfm?id=97
Medical Device Safety Alert
Risk of Local Adverse Events following Cardiac Catheterization by Hemostasis Device Use - Phase II, online available: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/TipsandArticlesonDeviceSafety/ucm070057.htm
| Device | Device Identifiers | Event Description |
| Abbott Vascular Devices/ StarClose | Catalog # 14677/ Lot # 681696H | After an arteriogram, a StarClose device was used to close the artery (common femoral artery left), the device failed and tore a larger hole in the artery. The artery was repaired with suture. |
| Abbott Vascular Devices/ StarClose | No Device Identifiers | A StarClose device was being used on patient post cardiac cath procedure. When the patient got up to ambulate, the device did not hold and the patient had slight bleeding. A pressure device was applied and the patient was put on bedrest for 4 hours. Patient was discharged in a normal timeframe. The patient did not develop any other issues. |
| Abbott Vascular Devices/ StarClose | Catalog # 14677/ Lot # 642096H | The physician was attempting to deploy a vascular occlusion clip when the vascular closure system failed to deploy correctly. The end of the closure system wand twisted 90% and became lodged in the vein. The physician successfully removed the device and performed the procedure successfully with an additional closure system. Manufacturer requested that we return device to them for analysis. Manufacturer is to provide a written report. |
| Abbott Vascular Devices/ StarClose | Lot # 682696H | Following a closure of an aneurysm in the leg with a covered stent, an attempt was made to close the access in the left groin. StarClose was attempted by the Dr. without success. Pressure was held until hemostasis was obtained. |
| Abbott Vascular Devices/ Perclose Proglide | Catalog # 12673/Lot # 60026-6H | The Perclose device did not fire. A new device was opened and also did not fire. The patient developed a hematoma which required pressure for 20 minutes prior to discharge from the OR to the PACU. |
| Abbott Vascular Devices/ Perclose Proglide | Catalog # 12673/ Lot # 690436H | The device did not contain suture. |
| Abbott Vascular Devices/ Perclose Proglide | Catalog # 12673/ Lot # 690066H | Two Perclose vascular closure devices failed to deploy the appropriate stitch in the artery. The vessel was subsequently closed with a device by a different manufacturer. There was no patient harm. The devices were saved but only one packaging was saved. |
| Abbott Vascular Devices/ StarClose | Lot # 670046H | Procedure: Angiogram: Iliac/PTA/Stent. The patient's left common femoral artery was accessed for an AP aortogram and angiogram and a StarClose device was used for closure. There was some type of misfiring of the StarClose device which did not allow it to release. The physician then made an incision in the left groin to manually remove the StarClose device and then held pressure manually over the femoral artery for 15 minutes. A hemostatic agent was placed and the incision was sewn shut. The patient tolerated the procedures well. |
| Abbott Vascular Devices/ StarClose | Lot # 720476H | A 6F StarClose was flushed and prepped. A guide wire was advanced and the sheath was exchanged for a StarClose sheath. The guidewire and dilator were pulled out and the StarClose was inserted. The top button was clicked down and the trigger was advanced splitting the sheath, but the clip on the StarClose was not deploying. The Dr. tried several times to get the device to deploy. The system was abandoned and manual pressure was held until hemostasis. |
| Abbott Vascular Devices/ StarClose | Lot # 740266H | After the sheath was changed to a StarClose sheath, the device was inserted through sheath. It was properly snapped to sheath and raised up to begin splitting, but splitting did not work. It was then noticed that a leaflet on the end of StarClose was kinked and would not advance any further. After several attempts at pulling out the system with no success, the emergency release button was used and the StarClose was aborted. Manufacturer response: the device will be replaced |
| Abbott Vascular Devices/ StarClose | Catalog # 14677/ Lot # 700636H | A left femoral arteriogram was completed. The StarClose was used on left groin but failed requiring compression and application of a pressure device. Patient was not injured. |
| Abbott Vascular Devices/ StarClose | Catalog # 14677/ Lot# 720676H | During an arteriogram the vascular closure device malfunctioned. The sheath didn't split completely and the StarClose device did not deploy. The patient was not injured. |
| AccessClosure, Inc. / MYNX | Lot # F0808102 | The patient developed a hematoma about the size of egg after a MYNX closure device was used to close vessel status post procedure. Manual pressure and another brand of closure device were applied. The patient was discharged but was re-admitted 5 days later for fever in the right groin. |
| AccessClosure, Inc. / MYNX | No Device Identifiers | A MYNX device was applied to the right groin but device failed. The patient developed a hematoma, and was hypotensive in the Cath Lab. Dopamine was started and another type of wound closure device was applied. The patient was transferred to CCU for observation. The patient became hemodynamically stable and the hematoma resolved. The patient was discharged home. |
| AccessClosure, Inc. / MYNX | Lot # F0827308 | When the closure device was deployed, the balloon tip for the device broke off. |
| AccessClosure, Inc. / MYNX | No Device Identifiers | A MYNX closure device was deployed in the right femoral artery. Bleeding was noted at site and manual pressure was applied by the physician. Venous sheath was also disconnected at this time per the physician's order, with pressure to the venous site. The site continued to ooze blood. A pressure device was applied approximately 15 minutes later, per the physician's order. Hemostasis was obtained with the pressure device. |
| AccessClosure, Inc. / MYNX | Lot # F0823304 | Unsuccessful closure with the device. A MYNX balloon burst while pulling back on device through sheath. It is thought to be device malfunction and not operator error. |
| AccessClosure, Inc. / MYNX | Lot # F0824904 | As the physician was attempting to deploy the MYNX device, the balloon expanded but the physician was unable to withdraw the catheter to expose the sealant. After the MYNX was passed off the sterile field, the physician attempted to pull the catheter up, but was unable to do so. Blood loss during the procedure was within normal limits (50-100cc) and a total of 50 cc of radiopaque was used during the procedure. The patient had a palpable popliteal pulse on the right side post procedure and was transported to recovery in stable condition. The physician stated one of every two MYNX devices used is defective. |
| Cardiva Medical Inc. / Boomerang | No Device Identifiers | A boomerang device was disengaged but operator unable to retrieve boomerang device. Manual pressure was held and the Dr. was notified. Manual pressure was held until finally the Boomerang device came out. Manual pressure was continued. No injury to patient was noted. The Boomerang device was noted to have a bend in the wire when it was retrieved. |
| St. Jude Cardiology/ Angio-Seal Evolution | Catalog # C610134 | Radiology tech was placing Angio-Seal, post Cardiac Catheterization. Post placement of Angio-Seal, a wire was retained. This retained wire was identified while patient was undergoing a stent placement one month later. Intact wire was removed without incident or damage to patient. |
| St. Jude Cardiology/ Angio-Seal | No Device Identifiers | This report covers four incidences in which the patient's experienced hematomas. The incidences were within a two week period. |
| St. Jude Cardiology/ Angio-Seal | No Device Identifiers | Patient had a cath and the next day was sitting in bed and felt something pop in the right groin. The patient felt need to go to the bathroom, got up, felt weak, and fell to the floor. The patient was discovered to have a massive bleed with a retroperitoneal hematoma. The patient was taken emergently to surgery and was found to have bled from iliac artery. The surgeon indicated that the Angio-Seal had failed. |
| St. Jude Cardiology/ Angio-Seal | Catalog # 610130 / Lot # 2076405 | The patient underwent a cardiac cath through a right groin approach. Three weeks later the patient was noted to have decreased pulses in the right leg and was sent back to hospital for evaluation. Angiogram on this day revealed a large 90% flap obstruction of the common femoral artery. It was felt to be hostile and unstable and possibly could cause an acute vascular compromise down the right leg. Urgent surgery was performed. The flap was removed. At the end of the surgery, the patient had a palpable pulse. The pathology report states the mass that was removed was fibrin and blood. |
| St. Jude Cardiology/ Angio-Seal | No Device Identifiers | A right groin hematoma developed less than 18 hours after cardiac catheterization resulting in delayed discharge and patient discomfort. |
| St. Jude Cardiology/ Angio-Seal VIP | Catalog # 610310 | Vessel occlusion at the access site occurred during the cath lab visit. An Angio-Seal device was used as a closure device. The device was inserted per protocol but it malfunctioned. Upon attempting to cinch down collagen plug, the piece used to cinch the collagen plug was not able to be withdrawn. It appeared that the collagen plug was intra-arterial. An initial femoral angiogram revealed arterial occlusion. Subsequently vascular surgery was called emergently. The plug was able to be withdrawn but still remained intra-arterial; flow was restored to the femoral artery. The patient was taken emergently for a cut down and removal of the plug. This was done without immediate complication. |
| St. Jude Cardiology/ Angio-Seal | No Device Identifiers | Bleeding occurred after a percutaneous intervention with an Angio-Seal closure device. Manual pressure was held for fifteen minutes, and hemostasis was achieved. |
| St. Jude Cardiology/ Angio-Seal | Catalog # 610310/ Lot # 2023627 | The device was removed from the package and assembled; the dilator was inserted into the sheath. The physician removed the arterial access sheath over a guidewire. The Angio-Seal device was inserted into the artery over the existing guidewire as per standard operating procedure. The physician prepared the device to be deployed and noticed that the suture material was missing from the device. He abandoned the deployment, removed the device and applied manual pressure to the groin to obtain hemostasis. Hemostasis was successfully obtained with no sign of hematoma. The manufacturer was notified. Manufacturer representative came on site and re-inserviced staff and physicians on proper use and handling |
| St. Jude Cardiology/ Angio-Seal | Catalog # 610310/ Lot # 2657479 | A 6fr Angio-Seal was deployed per MD for vascular closure. The device did not deploy properly causing a tear/dissection of the left femoral artery. The patient was taken to OR for repair. |
FDA is currently looking into the other listed device problems.
[Note: The reports have been edited for clarity]
Additional Information:
[1] Tavris, D., Dey, S. & Albrecht-Gallauresi, B. (2005). Risk of local adverse events following cardiac catheterization by hemostasis device use — phase II. Journal of Invasive Cardiology,17 (2).
http://www.invasivecardiology.com/article/4937
[2] FDA MedSun Database
http://www.fda.gov/MedicalDevices/Safety/MedSunMedicalProductSafetyNetwork/ucm127686.htm
FDA Recall Notice. Class 2 Recall: StarClose. January 31, 2007.
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ID=50232
Preventing Serious Injuries from Hemostasis Devices. FDA Patient Safety News. April 2002.
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/psn/printer.cfm?id=97
Risk of Local Adverse Events following Cardiac Catheterization by Hemostasis Device Use - Phase II. FDA Medical Device Safety.
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/TipsandArticlesonDeviceSafety/ucm070057.htm
Summary of MedSun Reports Describing Adverse Events With Electrosurgical Cutting and Coagulation Devices: Fire Associated Reports |
Print Item E–mail Item |
An electrosurgical cutting and coagulation device (and its accessories) is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.[1]
Over the past 2 years, MedSun has received 13 reports where fire is association with the use of electrosurgical cutting and coagulation devices [2]. The reports represent devices across 8 manufacturers:
• 8 reports for Valley lab and
• 1 report for the following manufacturers: Arthrocare, Boston Scientific, Megadyne, Karl Storz, Gyrus, Integra LifeSciences Corporation, and Olympus.
The reports were submitted by 12 hospitals between May 2007 and May 2009.
The reported device problems in these reports include:
• 13 reports of fire
• 2 reports of spark
• 1 report of heat
• 1 report of arcing
None of reports involved a patient death. The patient injuries listed below were reported in 9 out of 13 reports:
• Burn
• Re-intubation
• Additional Therapy
• Wound
• Redness
• Tissue damage
Of the reports that listed patient age, 5 had a patient age listed as less than 21 years and 8 had a patient age listed as greater than 2 years. Of the reports that listed patient gender, a total of 4 reports involved female patients and a total of 7 reports involved male patients.
These MedSun reports contributed to FDA awareness of the device problems. FDA currently has this issue under review.
| Device | Device Identifiers | Event Description |
| Valley Lab, Brand name: FORCE FX, Type of device: ESU hand piece | Model# 34828 | The Bovie tip spontaneously irrupted into a small flame. There was no harm to patient or staff. The situation was controlled immediately |
| Karl Storz Endoscopy -America, Inc., Type of device: Monopolar ESU cord | Lot #:IE01 | During a laparoscopic appendectomy, the cautery device attached to a Maryland Kelly Dissector ignited at the level of the cord. The fire spread to the surgeon's gown and deeper into the surgeon's scrubs. The instrument was immediately removed from the patient and the fire was extinguished with the surgeon's hand. Water was doused on the surgeon's scrubs and gown and the patient's drapes. Oxygen was immediately shut off. No harm to the surgeon or patient. Later on, the case continued without incident. |
| Covidien LP, Formerly known as Valley Lab, Brand name: ESU, Type of Device: Hand switching suction coagulator | Lot #: 151616, Catalog #: E2610-6 | At the start of adenoidectomy surgery, a tongue blade and mouth gag were inserted and the mouth was opened for exposure. A red non-latex Robinson catheter was placed transnasally and brought out transorally to help elevate the palate for exposure of the nasopharynx and access to the nasopharynx. At the start of this adenoidectomy, a method of fulguration, using suction cautery, an apparent oxygen leak resulted in a spark. The red non-latex Robinson catheter ignited, and which rapidly began to smolder. A fire estimated to be the size of a fifty cent piece was observed in the oral cavity. The endotracheal tube and the red Robinson were immediately removed from the patient and the oral cavity was doused with saline, and then suctioned out. The patient was re-then intubated with an endotracheal tube. Immediate examination of the oral cavity and larynx revealed that the patient had a 0.5 cm x 2 cm burn on the left lateral tongue, as well as an 8 mm x 8 mm burn on the buccal mucosa of the right lateral lower lip. All just inferior to the crease of the oral commissure. The area was immediately iced. The burns continued to be iced and were classified as superficial and minor by the physician burn specialist. It was later treated with Neosporin. Despite no evidence of burn on the endotracheal tube, it was removed from the patient. The decision was made to confirm that there was no airway injury and direct laryngoscopy and bronchoscopy was performed with confirmation of no airway injury to the larynx or trachea to the level of the carina. Uncuffed ETT had a large air leak that increased when neck hyperextended; 02 flow rate was high at 89%-92% at the time of event; there was no soaked gauze throat packs in place; there was red non-latex Robinson catheter in mouth; and diluted Hibiclens was being used as a dental mirror defogging agent. |
| Covidien Valleylab, Device#1: Brand name: Force FX-C, Type of device: ESU unit, Device#2: Type of device: ESU hand piece | Product# Unknown | Patient under conscious sedation for insertion of portacath for chemotherapy into right external jugular vein. ESU pencil was being used to control bleeding. Patient was receiving 6L (liters) of oxygen via facemask and surgical site had been prepped with Chloraprep (which contained 70% alcohol). Anesthesiologist saw 2 separate flames: one behind patient's right ear and one on right side of facemask. Anesthesiologist pulled off facemask and patted flames with blue towel. Scrub poured a bowl of sterile saline on patient's face and surgeon pulled all drapes off of patient. The procedure was abandoned, and the patient examined and wounds dressed. The patient was taken into recovery and seen by a burn surgeon and plastic surgeon. Second-degree burns were found at right corner of mouth, at the base of her right ear and right side and back of her neck. They did not find any wounds in nose or mouth and the patient will have no respiratory consequence. An ophthalmologist found no injury to her eyes. It is not known for sure what caused the adverse event, but after discussion, the causation theory is that a spark or heat from the ESU pencil ignited the blue cotton drape that must have had remnants of the alcohol-based prep used to cleanse the skin for surgery. The oxygen likely pooled in high concentration around the face mask and near the table at the base of the neck and acted as an accelerant. Chloraprep with Tint, 26mL applicator, was being used at the time of event. |
| Valleylab, Device#1: Brand name: Suction Coagulator, Type of device: ESU hand piece; Device#2: Type of device: ESU unit | Device #1: Model #: E2505-10FR, Device#2: Model# Force FX | A patient was in the OR for PE tube placement and adenoidectomy. During the procedure it was noted that the tip of the suction catheter ignited. This was immediately recognized and the suction catheter was removed and the flame extinguished. Patient was assessed and no injury, including airway trauma or burn noted. The site reported the manufacturer told hospital that if the tip of the suction catheter becomes blocked with material, there is the potential for this material to ignite once the device is turned on. |
| MFR#1: Gyrus Medical, Inc. Device#1: Type of device: ESU Cable Accessory; MFR#2: Covidien Valleylab, Device#2: Brand name: Force FX, Type of device: ESU unit; MFR#3: Boston Scientific Corporation, Device#3: Brand name: Titanium Wedge electrosurgical Monopolar, Type of device: Electrode tip | Device#1: Model #: RAC-B Catalog #: 31880110, Device#2: product#: Unknown, Device#3: Catalog# 880-311 | Resectoscope cord caught fire while in use by physician. The fire was in the cord only, the tip was not involved. The duration of the fire was 2 seconds. The electrosurgical machine was set to maximum (120 W Coag/ 300 W cut) Patient had a 1 cm area of redness on penis. No treatment required. |
| Covidien Valleylab, Brand name: Electrosurgical pencil, Type of device: ESU hand piece | Model #: 145436 Lot #:146503 | Using general anesthesia via mask, and cautery, about 1 minute into procedure flames under drapes at right side of face. Second degree burns occurred pt face. |
| Integra LifeSciences Corporation: Brand name: Bipolar forceps | Model #:285-188 | Bipolar forceps burst into fire at junction where forceps is connected to the bipolar cord. Fire extinguished by itself. No harm to patient. Disposable bipolar cord came from flap pack. |
| Olympus Surgical and Industrial America Inc: Brand name: ESU active cable | Catalog #:AO35B | During laparoscopic cholecystectomy procedure, the end nearest to the cautery machine of the cautery cord caught fire and burned through. No harm to patient or staff. |
| MEGADYNE Medical Products, Inc., Brand name: Bovie tip, Type of device: ESU hand piece | Catalog # 139926 Lot # 813306, Model# +M1160001399261H | The device flared at the tip when with an attempt to use it outside the patient. The case was completed successfully using another device. |
| Covidien Valleylab, Type of device: ESU suction coagulator electrode | Model#:E2505-10FR Lot #:127510 | While the surgeon was cauterizing the adenoids, the oral cavity caught fire. Saline solution was poured into the mouth. There was no injury to the patient. |
| VALLEYLAB, Brand name: electrosurgical unit | Product# Unknown | During a tracheotomy the patient's O2 saturation began to drop. This required an increase in oxygen to 100%. A spillage of oxygen out of the mouth occurred into the area of the wound. There was a short flash of fire on the chest secondary to electrocautery. The fire was put out with sterile water. The patient's chest hair was singed. No apparent burn |
| Arthrocare Corporation, Brand name: Coblator II Procise EZ Wand with Integrated Cable, Type of Device: ESU Hand piece | Model# EC8870-01, Lot# L102770-A | Coblator being used for hemostasis in tonsillectomy procedure. The wand got hot, with apparent "arcing" to adjacent tissue, causing minor damage. |
Additional Information:
[1] Code of Federal Regulations Title 21
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfCFR/CFRSearch.cfm?FR=878.4400
[2] FDA MedSun
http://www.fda.gov/MedicalDevices/Safety/MedSunMedicalProductSafetyNetwork/ucm127686.htm
Summary of MedSun Reports Describing Adverse Events With Electrosurgical Cutting and Coagulation Devices: Patient Burns |
Print Item E–mail Item |
An electrosurgical cutting and coagulation device (and its accessories) are devices intended to remove tissue and control bleeding by use of high-frequency electrical current. [1]
Over the past 2 years, MedSun has received 36 reports where patients receive a burn in association with the use of electrosurgical cutting and coagulation devices [2]. These reports represent devices across 15 different manufacturers:
• 18 reports for Valleylab,
• 4 reports for Arthrocare,
• 3 reports for ConMed,
• 3 reports for Intuitive Surgical,
• 2 reports for Boston Scientific,
• 2 reports for Megadyne,
• 2 reports for Olsen, and
• 1 report for the following manufacturers: Angiodynamics, Mizuho, Salient, Cardima, Gyrus, Encision, Depuy Mitek, Cardinal Health.
The reports were submitted by 30 hospitals received Between May 2007 and May 2009.
The reported device problems in these reports include:
• Degraded insulation
• Arcing
• Spark
• Heat
• Holes
None of reports involve a patient death. The patient injuries listed below were reported in 18 of 36 reports noted to be associated with patient burn:
• 3 reports of 3rd degree burn
• 3 reports of tissue damage
• 5 reports of additional therapy/surgery
• 2 reports of aborted surgery
Of the reports that listed patient age, six reports had a patient age listed as less than 21 years and 30 had a patient age listed as greater than 21 years. Of the reports that listed patient gender, a total of 23 reports involved female patients and a total of 8 reports involved male patients.
These MedSun reports contributed to FDA awareness of the device problems.
| Device | Device Identifiers | Event Description |
| ANGIODYNAMICS, INC. Brand name: STARBURST SEMI-FLEX, Type of device: Catheter, ablation, RF, tumor | Lot # 952314, Model # 700-102-615 | Patient had femur osteoid osteoma. While undergoing radiofrequency ablation of femur tumor in interventional radiology a probe was placed, and it did not work at the site of the tumor. The probe did cause a burn of approximately 2 cm x 1 cm on the skin of the patient at the insertion site. Replaced probe with another one and it worked appropriately. Patient recovering without incident. |
| ValleyLab, Brand name: FORCE FX, Type of device: ESU | Product# Unknown | Physician wanted to utilize the harmonic scalpel yet pressed ESU foot pedal instead of harmonic scalpel foot pedal. Patient experienced burn of less than 0.5 cm on upper left abdomen when the ESU endoshears were mistakenly activated. |
| MFR#1: CONMED ELECTROSURGERY, Device#1: Type of device: ESU device, MFR#2: Olsen Medical, Device#2: Type of device: electrosurgical forceps | Device#1: Catalog # 20-1370, Model #5000, Lot #003163, Device#2: Catalog#: 20-1310 | The physician was performing a tonsillectomy. It was noticed that the patient had received a burn to the upper lip at the point of contact with the bipolar forceps. |
| MIZUHO AMERICA, INC., Brand name: BAY PROBE INSULATED BLUE WITH SPRING, Type of device: Transnasal Transsphenoeidal instrument set | Model # ISOLATED BAYONET SPRING 110MM ITEM 07-824-02 | During transnasal transsphenoidal (TNTS) surgery, the physician used the bay probe insulated blue with spring (bovie stick) inside the patient's nose. Physician noted a burn on the patient's left base of the nose, about 1 cm. Physician inspected the bay probe insulated blue with spring; there were two holes in the insulation. |
| SALIENT SURGICAL TECHNOLOGIES, INC., Brand name: AQUAMANTYS 2.3 BIPOLAR SEALER, Type of device: ESU | Device #1: Model # Aquamantys 2.3 | Pediatric patient taken to surgery for liver mass. A bipolar electrosurgical device was used during the procedure. A grounding pad was placed on the left shoulder. After the procedure, it was discovered that the patient sustained two burns to the left shoulder area. The first burn was a 1.5cm.x1cm full thickness burn and located on the superior aspect of the left shoulder, slightly posterior. The second burn was also full thickness, and measured 2cm x 3cm and was located on the upper back just lateral and inferior to the spine of the scapula. |
| ARTHROCARE CORPORATION, Brand name: Coblator II procise EZ wand with integrated cable, Type of device: ESU hand piece | Model # EIC8870-01, Lot # L123570-A | Surgeon using coblator for dissection/hemostasis during tonsillectomy. Small area of metal at the tip of the insulated hand piece became exposed and conducted energy to patient's tongue, resulting in approximately 1cm burn to the right proximal area. No treatment needed. |
| MFR#1: CARDIMA, Device#1:Brand name: Intellitemp Energy Management device, Type of device: Cardiac RF ablation, MFR#2: Covidien ValleyLab, Device#2: Brand name: Force FX, Type of device: ESU unit, Device#3: Type of device: ESU disperser electrode | Device#1:Model # 0075-04435, Catalog # 11-081008, Device#2 and 3: Product# Unknown | The patient was admitted for an atrial ablation procedure. The thoracotomy incision was made and the Valleylab Force FX Bovie was being used along with the Valleylab grounding pad. During the ablation portion of the procedure, the grounding pad was disconnected from the Valleylab Machine and connected to the Cardima Intellitemp Machine. The ablation equipment was then connected to the Valleylab Machine and after the ablation was complete, a burn was noted on the patient's right buttock, where the grounding pad was placed. |
| VALLEYLAB, Brand name: REM POLYHESIVE II, Type of device: ESU grounding pad | Catalog # E7507, Lot #125319 | The cautery/ESU grounding pad was placed on the patient for a port removal procedure. When taking the pad off the patient, a reddened area around the edge of the pad was seen with some breakdown of the skin at the cord entrance site. The wound was classified as a small blister or a burn injury at the site. |
| ARTHROCARE CORPORATION, Brand name: Arthrocare, Type of device: Arthroscopy wand | Catalog # ASC4250-01, Lot # A226570-A | When drapes were removed the physician observed a tract like appearance on the medial aspect of patient's left knee (operative leg). It was described as looking like an "inside out" burn. |
| ARTHROCARE CORPORATION, Brand name: Coblation hand piece, Type of device: ESU handpiece electrode | Lot#: B207570-A | The patient was undergoing a tonsillectomy and adenoidectomy. The surgeon was using the hand piece inside the tonsillar area to remove the tonsils and adenoids. The surgeon stated that while he was using the hand piece inside of the patient's throat he saw it "arc" causing a superficial burn. Both pieces of the apparatus were removed and sent to the biomedical department. The sales representative was also notified of the incident. |
| GYRUS MEDICAL, INC., Brand name: Everest Bipolar Bicoag, Type of Device: ESU Forceps | Model#: 3005 5MM, 33CM, Lot#: 7352014 | The physician was using the Everest bipolar cutting forceps when it was noted to be burning the tissue, sticking to the tissue and ripping it. The device was cleaned by the scrub nurse with no improvement. The procedure was stopped |
| COVIDIEN VALLEYLAB, Device#1: Brand name: FORCE FX-C, Type of device: ESU unit, Device#2: Type of device: ESU hand piece | Product# Unknown | Patient was under conscious sedation for the insertion of a portacath to infuse chemotherapy into right external jugular vein. ESU pencil was being used to control bleeding. The patient was receiving 6l (six liters) of oxygen via facemask and surgical site had been prepped with chloraprep (which contained 70% alcohol). The anesthesiologist saw two separate flames, one behind patient's right ear and one on right side of facemask. Anesthesiologist then pulled off the facemask and patted flames with blue towel. Scrub (staff member) poured a bowl of sterile saline on patient's face and surgeon pulled all drapes off of the patient. Procedure was abandoned, and the patient was examined with the wounds dressed. The patient was taken into recovery and seen by a burn surgeon and plastic surgeon. Second degree burns were seen at the right corner of mouth, at the base of her right ear and right side and back of her neck. They did not find any wounds in nose or mouth and the patient will have no respiratory consequence. An ophthalmologist found no injury to their eyes. It is not known for sure what caused the adverse event, but after discussion, the causation theory is that a spark or heat from the ESU pencil ignited the blue cotton drape that must have had remnants of the alcohol-based prep used to cleanse the skin for surgery. The oxygen likely pooled in high concentration around the face mask and near the table at the base of the neck and acted as an accelerant.. |
| ARTHROCARE CORPORATION, Brand name: Coblator II Procise XP, Type of device: ESU Coblation electrode | Catalog#: EIC8872-01, Lot#: R727470-A | Patient received anesthesia for tonsillectomy and adenoidectomy using a coblator. Equipment checked preoperatively for defects and visually appeared to be normal in appearance. Patient sustained burn to right side inner bucchal mucosa. Coblator wand inspected and found to have missing insulation on external surface of the wand. Injury documented with images on digital camera./TD> |
| CONMED ELECTROSURGERY, Type of device: ESU suction coagulator electrode | Lot#: 07FHL05 | Physician was doing an adenoidectomy on this patient. They were using the bovie and thinks there was a small break in the coating of the instrument that caused the patient burn. Dr. stated that there was a small area on the uvula that was affected. |
| COVIDIEN VALLEYLAB, Brand name: FORCE II, Type of device: ESU unit | Model#: 51500 | During a c-section the physician was holding a hemostat. When the cautery was used on the hemostat, it burned a hole in the glove, causing a small burn to the physician's thumb. The physician reported the burn to be the size of a pencil eraser. The hemostat was shiny, the first time the physician had seen it. No adverse outcome to patient. |
| VALLEYLAB, Device#1: Brand name: SOLID STATE ELECTROSURGERY FORCE IC, Device#2: Type of device: ESU needle tip electrode, Device#3: Type of device: ESU suction coagulation electrode | Device#1: Model# 590B; Device#2 and # Product# Uknown | Patient was scheduled for a tonsillectomy and adenoidectomy. Patient had a previous history of adenotonsillar hypertrophy that was thought to cause dysphagia for solid consistencies. During the surgery, yet after the dissection of the first tonsil, a 4-mm right lower lip burn was noticed. Ice was immediately placed on the area in addition to a layer of Vaseline. The cautery, both needle-tips and suction were disconnected and sent to biomed for evaluation. The rest of the procedure continued without incident. After the procedure was completed, the surgeon referred the child to a plastic surgery specialist. The surgeon involved also contacted the plastic surgeon to discuss the case and identify any immediate interventions. The biomed completed an inspection of the equipment. All equipment passed visual inspections, which included no evidence of nicks, cuts or abrasions. Tests revealed that the equipment met manufacturer's specifications of output and voltage. |
| MFR#1: ENCISION, INC., Device#1: Brand name: AEM, Type of device: Active electrode monitor, MFR#2: COVIEDEN VALLEYLAB, Device#2: Brand name: Force II, Type of device: ESU generator | /Device#1: Model# Encision, Device#2: Product# Unknown | Per the physician, the patient received a small electrical burn to the upper right quadrant of the abdomen. The grounding pad, as well as the cautery cords was properly connected. No alarms sounded and the equipment functioned without shutdown. The burns were received near the very end of the case, when the ball tip cautery was being used inside. The trocar was also in use, which has an eyepiece lens attached at a 90 degree angle. When the doctor moved the lens from the patient's abdomen, the scrub tech saw the red spots on the patient, and an investigation immediately began. The red spots were directly under the lens placement and were the width of the lens. The doctor had already finished with the cautery at this point. It was immediately removed and taken out of service. The company representative was contacted. The cautery cords were left intact in the machine for the clinical equipment staff to check and investigate further. There were no visible breaks in the insulating sheaths on the green cautery equipment. The OR supervisor came to the operating room to view the burn. No further treatment was required for the patient, as the physician instructed for the area to be left open. The burn area totaled approximately 1 inch in diameter, was reddened with about three small whitish streaks inside the reddened area. No blisters were observed. This event could not be duplicated. We tried to touch the cautery tip to the eyepiece but could not because of the inflexibility of the cautery. An active electrode monitoring device was also used during this case, and it did not alarm for any stray current. The ESU and the active electrode monitor were both checked by clinical equipment and both devices functioned within normal established limits. The active electrode monitor is under a service contract and was only checked for electrical safety. The ESU received a PM approximately five months prior and functioned within normal established limits at that time. |
| MFR#1: BOSTON SCIENTIFIC CORP., Device#1:Brand name: RF 3000, Type of device: RF Ablation System; MFR#2: Covidien Valleylab, Device#2; Brand name: Standard Polhesive, Type of device: ESU return electrode pad | Device#1: Catalog# CS545, Model# RF3000; Device#2: Catalog# E7506, Lot# 128795 | Patient underwent a liver ablation under CT guidance. After the ablation was done, the grounding pads were removed from patient's thighs. Once the grounding pads were removed, the burns were noted on the lower and upper part of the left thigh pad and the lower right thigh pad. The burn on the lower right thigh was worse than the left. |
| COVIDIEN VALLEYLAB, Brand name: FORCE FX, Type of device: ESU generator | Product# Unknown | A spark was noted at the time of the tracheostomy at the insertion site, while a bleeder was being cauterized. Oxygenation was stopped immediately, the endotracheal tube removed, and a tracheal tube inserted. The trachea was flushed with normal saline, cleaned and suctioned. A fiber optic evaluation of the trachea and bronchi was done, which revealed a possible second degree burn. |
| MFR#1: TRI-ANIM, Device#1:Brand name: Metzenbaum, Scissors, Type of device: Surgical Scissors, Device#2: Type of device: Scissors handle, MFR#2: Cardinal Health, Device#3: Brand name: Presource Custom, Type of device: ESU cable; MFR#3: ConMed Electrosurgery, Device#4: Brand name: SABER, Type of device: ESU unit, Device#5: Type of device: ESU Dispersive electrode | Device#1: Model# 810-01-21218, Lot# 44691, Device#2: Product# Unknown, Device#3: Catalog# SID23MCUB, Lot# 468976; Device#4 Model# 60-5600-02, Device#5 Catalog# 410-220 | Areas of endometriosis were excised, utilizing monopolar scissors, from the right and left posterior broad ligaments, the right posterior cul-de-sac, and the right and left anterior cul-de-sac. After completing the last part of the procedure, the right posterior cul-de-sac, the surgeon noticed a cautery injury on the posterior wall of the uterus, which was made hemostatic with bipolar cautery graspers. The surgeon then noted this on the medial pole of the right ovary and on an approximately 1.5 cm segment of the right fallopian tube. It appeared that somewhat less than half the diameter of the right tube had been cauterized. Fragments of black plastic material were noted in the pelvis, which seemed to be portions of insulation from the monopolar scissors. The scissors were removed and inspected. There were clear defects in the insulation on the scissor tip, and evidence of heat damage to the distal tip of the scissor shaft itself. The bovie machine settings were coag 40 and cutting 40. |
| COVIDIEN VALLEYLAB, Device#1: Brand name: COOL TIP, Type of device: RF ablation controller; Device#2: Brand name: Cool TIP, Type of device: Peristaltic pump; Device#3: Brand name: COOL TIP, Type of device: RF ablation system; Device#4: Brand name: COOL TIP, Type of device: RF ablation grounding pad; Device#5: Brand name: COOL TIP, Type of device: RF ablation electrode kit | Product# Unknown | A patient was treated with radio frequency ablation for an osteoid osteoma of the left femur. Upon completion, a burn under the grounding pad was noted. The patient also required surgery to remove a left thigh necrotic seroma status post radio frequency ablation (rfa). Two grounding pads had been used, one on each thigh. A draining sinus was also noted along the initial application port of the probe. This was resected during the subsequent surgery |
| DEPUY MITEK (A JOHNSON AND JOHNSON COMPANY), Brand name: VAPR, Type of device: ESU Irrigation, suction system | Product# Unknown | VAPR system was used for arthroscopic surgery. It was noted during surgery that the suction had not been set up correctly causing the heated fluid to pool and cause the burn. There was not a continuous flow of irrigate. Equipment was unrecoverable and had already been thrown away or cleaned. Patient complained of burning sensation on neck in post anesthesia care unit. A 6 x 3 inch area was noted with two blisters anterior to neck. Iodoform and gauze dressing applied. Family instructed to continue Silvadene and gauze dressing after discharge for ambulatory surgery. |
| VALLEYLAB, Brand name: BOVIE FX, Type of device: ESU device | Model# FX | Surgeon noted a bovie spark. After surgery the surgeon noted that there was a small hole on the small finger of his right glove. His small finger had a burn where the spark had gone through glove; burn treated with Silvadene; no other treatment needed. Equipment was checked by biomed; no problems found and it was returned to service. |
| BOSTON SCIENTIFIC, Device#1: Brand name: RF 300 RADIOFREQUENCY SYSTEM, Type of device: ESU, RF Ablation; Device#2: Brand name: Coaccess Introducers set, Type of device: ablation Electrode Introducer; Device#3: Brand name: Leveen Superslim Needle Electrodes, Type of device: RF Ablation Electrode | Device#1: Model#26-220, Catalog# M001262200, Device#2: Catalog# 26-225, Model# M001262250, Device#3: Catalog# 26-228, Model# M001262280 | At the end of the radio ablation procedure, the radiologist noted a posterior shoulder burn near the entrance of the radio frequency needle. Metal hemostat was used by radiologist (per recommendation of the Boston scientific sales representative that was present for the procedure), to stabilize the needle in the patient's lung, thus preventing the tip of needle from moving when ablating tumors. Visual examination of the radio frequency needle and introducer set with hemostat still attached was done following the procedure. A break in the insulation sheath of the introducer was noted |
| COVIDIEN (VALLEYLAB), Brand name: ESU Hand piece | Product# Unknown | The physician laid the hand piece, which had been previously used, on the patient's chest. It began to burn the skin. The physician immediately picked it up, but the skin was already burnt. The burn was a half inch long. The hand piece was not activated (the button was not pushed) and should not have been hot. |
| MEGADYNE MEDICAL PRODUCTS, INC., Brand name: EZ CLEAN, Type of device: ESU electrode | Catalog# 0014A, Lot# 71330 | During intraoperative use of bovie tip, 4" EZ clean caused burn on patient's left breast due to failure of protective insulation on tip. Patient has multiple dermal burns approximately 1cm in diameter on the left breast. |
| VALLEYLAB, Brand name: FORCE TRIAD, Type of device: ESU device | Product# Unknown | Patient was burned during her procedure by one of the instruments being used. Events discussed with MD who referred to them as 3rd degree burns along incision line and was able to excise the damaged tissue. According to the information provided by the OR techs, there was never any spark, or abnormality during the case until the end when the retractors were removed. At that time the area was noted to be red. This machine was brand new and had been tested and cleared upon delivery by biomedical engineering. Biomedical engineering has examined the equipment and they are unable to replicate the error / problem. |
| VALLEYLAB, Brand name: FORCE TRIAD, Type of device: ESU device | Product# Unknown | Patient was burned during her procedure by one of the instruments being used. Events discussed with the MD who referred to them as 3rd degree burns along the incision line and was able to excise any/all damaged tissue. According to the techs there was no sparks or abnormalities until the end of the procedure when the equipment flashed an error and stated notify biomed. The equipment was removed from service at this point. It was not until this case that they realized that there was an equip issue as this was patient #2 with the same burn as the case before. |
| COVIDIEN VALLEYLAB, Brand name: ESU PENCIL, Type of device: ESU Monopolar active electrode | Catalog# E2516W, Lot# 119864 | Small area of pt skin around incision site burned with use of cautery. The cautery was being used inside of neck and slipped. Small burn noted pink in color on upper left hand side of incision at end of case. Antibiotic treatment and 2x2 tegaderm placed. |
| OLSEN MEDICAL, Brand name: OLSEN, Type of device: Bovie Extender | Lot# 110802 | During a partial glossectomy, a superficial burn was noted just right of midline of the patient's lower lip and corresponding to the edge of the bovie extender. Extender was immediately removed. The inside rim of the extender, about 1/4 inch, was noted to have what appeared to be crack in the sheath. Char marks on exterior of extender and outer plastic coating appears melted. |
| INTUITIVE SURGICAL, INC., Brand name: DAVINCI SURGICAL SYSTEM, Type of device: ESU shears cover, Robotic device | PRODUCT # Unknown | During robotic assisted surgical procedure for laparoscopic hysterectomy with shears set at recommended level of 38 coag, arcing occurred causing burn to uterine tissue. No harm to patient since uterus being removed. Tip cover inspected before procedure and placed on shears according to company recommendations |
| INTUITIVE SURGICAL, INC., Brand name: DAVINCI SURGICAL SYSTEM, Type of device: ESU shears cover, Robotic device | Product# Unknown | Arc/spark occurred during use, a few minutes into case. Burn to uterus resulted. No harm to patient since uterus being removed. Tip cover then noted to have holes. Removed and replaced. A machine setting was at 38 coag per manufacturer recommendations. Also tip cover inspected prior to application using gage unit and applied as recommended. |
| INTUITIVE SURGICAL, INC., Brand name: DAVINCI SURGICAL SYSTEM, Type of device: ESU shears cover, Robotic device | Product# Unknown | Arc/spark occurred during robotic assisted surgery through scissor guard / accessory tip cover causing burn to uterus which was being removed. No harm then to patient. A setting for shears was at 38 coag which is recommended level. Tip cover inspected prior to application using gage provided by company. Also application of tip cover performed according to recommendations. |
| MFR#1: COVIDIEN VALLEYLAB; Device#1: Brand name: HANDSWITCHIN ROCKER SWITCH PENCIL, Type of device: ESU pencil; Device#2: Brand name: Force 2, Type of device: ESU generator; MFR#2: Megadyne Medical Products, Inc., Device#3: Brand name: E-Z Clean Blade Electrodes; Type of device: ESU electrode | Device#1:Catalog# E2515, Lot# 1387182013-03; Device#2: Model# Force 2-2PCH; Device#3: Catalog# 0012M, Lot# 2010-1172041 | An alternate site burn occurred at the left corner of the patient's mouth during a tonsillectomy. |
| COVIDEN LAB, Brand name: Electrosurgical pencil | Model# 145136, Lot# 146503 | Using general anesthesia via mask, about 1 minute into procedure flames under drapes at right side of face while using cautery. Cause 2nd degree burn to pt face |
| VALLEYLAB, Device#1: Brand name: Force FX, Type of device: ESU, Device #2: Brand name: Ligasure, Type of device: ESU vessel sealing | Device#1: Model# Force FX, Device#2: Model# Ligasure | Patient undergoing resection of large liver tumor. After surgery, patient had two burns on the left shoulder and back approximately 7 cm in diameter which required surgical debridement and flap closure. |
Additional Information:
[1] CFR – Code of Federal Regulations Title 21
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfCFR/CFRSearch.cfm?FR=878.4400
[2] FDA MedSun Database
http://www.fda.gov/MedicalDevices/Safety/MedSunMedicalProductSafetyNetwork/ucm127686.htm
All Medical Device Problem Summaries to Date
- (July 2009) Summary of MedSun Reports Describing Adverse Events With Vascular Hemostasis Devices
- (July 2009) Summary of MedSun Reports Describing Adverse Events With Electrosurgical Cutting and Coagulation Devices: Fire Associated Reports
- (July 2009) Summary of MedSun Reports Describing Adverse Events With Electrosurgical Cutting and Coagulation Devices: Patient Burns
- (June 2009) Summary of MedSun Reports Describing Adverse Events With Cardiac RF Ablation Catheters
- (May 2009) Summary of MedSun Reports Describing Adverse Events With Surgical Mesh Products for Hernia Repair
- (April 2009) Summary of MedSun Reports Describing Adverse Events With Ventricular Assist Devices
- (March 2009) Summary of MedSun Reports Describing Adverse Events With Integra LifeSciences Corp.'s MAYFIELD Skull Clamp
- (March 2009) Summary of MedSun Reports Describing Adverse Events With "Harmonic Scalpel" Ultrasonic Surgical Instruments
- (February 2009) Summary of MedSun Reports Describing Adverse Events: Large Volume Infusion Pumps and Issues Resulting In Over-infusion
- (February 2009) Summary of MedSun Reports Descrubing Adverse Events with the Cooper Surgical M-Style Mushroom Cup Vacuum Assisted Delivery System
- (January 2009) Summary of MedSun Reports Describing Adverse Events With the ConMed VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) Uterine Manipulator
- (December 2008) Summary of MedSun Reports Describing Adverse Events With Arthroscopic Pumps/Tubing
- (November 2008) Summary of MedSun Reports Describing Adverse Events With Long Term Intravascular Catheters and Related Article
- (October 2008) Summary of MedSun Reports Describing Adverse Events With Infusion Pumps in Pediatric Patients and Related Articles
- (September 2008) Summary of MedSun Reports Describing Adverse Events With Electrocardiograph Electrodes in Pediatric Patients and Related Reference
- (September 2008) Summary of MedSun Reports Describing Adverse Events With Thermal Regulating Systems in Pediatric Patients and Related Reference
- (August 2008) Summary of Medical Device Adverse Event Reports Describing Adverse Events With Umbilical Artery Catheters And Related Article
- (August 2008) Summary of MedSun Reports Describing Adverse Events With Infant Radiant Warmers and Neonatal Incubators And Related Recalls and Articles
- (June 2008) Summary of MedSun Reports Describing Adverse Events With High Frequency Oscillatory Ventilators
- (June 2008) Summary of MedSun Reports Describing Adverse Events With Cranial Perforators
- (May 2008) Summary of MedSun Reports Describing Adverse Events With Transcatheter Septal Occluders
- (November 2007) Summary of MedSun Reports Describing Adverse Events With da Vinci Surgical Devices
Updated July 3, 2009

