• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 29, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 56447
Product Type Devices
Status Ongoing
Recalling Firm Technidata S.A.
City Montbonnot St Martin
State
Country FR
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2010-07-29
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including Washington DC and the states of California, Nevada, Kentucky, Alabama, Rhode Island, Pennsylvania, and Louisiana and the country of Italy.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
TD-Synergy. Laboratory Information System. TD Lims product versions: LIS (ALLX)-21.041-_ML-5370, LIS (ALLX)-22.021-_ML-5374, TD-Synergy product versions: SYN (ALLX)- 03.13B-_ML-5368, SYN (ALLX)- 03.141-_ML-5369, SYN(LINUX)-06.011-_ML-5381, SYN (ALLX)-11.11B-_ML-5345, SYN (ALLX)-11.21B-_ML-5346 SYN (ALL)- 11.31C-_ML-5372, SYN (LINUX)-11.321-_FR-5382, SYN (ALL)-11.321-_ML-5373 Class II Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request. 10 Z-1345-2013
-
-