Enforcement Report - Week of July 18, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 61231
Product Type Devices
Status Ongoing
Recalling Firm Terumo Cardiovascular Systems Corporation
City Ann Arbor
State MI
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-22
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Terumo Advanced Perfusion System 1 Electronic O2 Blender/Analyzer. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas. catalog number: 801188 and Serial numbers: 0001-0006, 0008-00050, 00052-00123,00125-00152, 00154-0174,00176-00211, 00213-00241, 00245-00250, 00300-00356, 00358-00365, 00367-00449, 00451-00489, 00492-00522, 00524-00531, 00533-00649, 00651-00660, 00662-00825,0826-0830, 00831-01157, and 01160-01164. Class II Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob. 1095 Units Z-1962-2012
Terumo Advanced Perfusion System 1 System 1 Base, 100/120V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas. catalog number: 801763 and Serial numbers: 0011-0335, 0337-0610, 1001-1034,1037-1049, 1100, 1103-1106,1109-1445, and 1447-1450. Class II Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob. 648 units Z-1963-2012
Terumo Advanced Perfusion System 1 System 1 Base, 220/240 V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas. catalog number: 801764 and Serial numbers: 0006-0066, 0100-0311, 0313-0322,1001-1027,1100-1434,1436, 1440, and 1441. Class II Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob. 992 Units Z-1964-2012
-
-