Enforcement Report - Week of June 27, 2012
-
Back to Previous Page
Event Detail
| Event ID | 61741 |
|---|---|
| Product Type | Devices |
| Status | Terminated |
| Recalling Firm | Medtronic Navigation, Inc. |
| City | Littleton |
| State | MA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2010-10-12 |
| Initial Firm Notification of Consignee or Public | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
| Distribution Pattern | Nationwide Distribution including the states of CO, IN, and MN. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-armĀ® Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging. | Serial Numbers: 125, 156, 234R. | Class II | Software Version 3.1.1 does not meet navigational accuracy. | 3 units | Z-1849-2012 |
-