Event Detail

Event ID 61742
Product Type Devices
Status Ongoing
Recalling Firm Terumo Cardiovascular Systems Corporation
City Ann Arbor
State MI
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-09
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
CPB CATHETER KIT- DR.VAN TRIGHT. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 80004-02 and Lot code: MG07, MK02, MM11, MN15, MP13, ND28, NE25, NN21, and PA23. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1842-2012
CUSTOM CPB CATHETER KIT-DR. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 80001-01 and Lot code: MG21, MM04, NA24, ND21, NK01, NM31, and PA30. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1841-2012
SPECIALTYCARE - HUDGENS. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 64625 and Lot code: NA17, NC07A, NG06, NG06A, and NL19. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1840-2012
24FR FLEX ART CANN NON STERILE. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 88813030 and Lot code: 0550551, 0568495, 0576844, 0586459, 0601958, 0626874, and 0635682. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1839-2012
SARNS FLEXIBLE ARTERIAL CANNULA: 8.0 MM (24 FR) 00 WITH 3/8" CONNECTOR, CM MARKING, 10" (25 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 144776 and Lot code: 0552165, 0560615, 0574341, 0586004, 0606472, 0614321, 0617360, 0622689, 0637775, 0640519, and 0653855. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1838-2012
SARNS FLEXIBLE ARTERIAL CANNULA: 8.0 MM (24 FR) WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 13030 and Lot code: 0553041, 0555967, 0557330, 0559885, 0562990, 0564951, 0567668, 0570456, 0571580, 0575365, 0580822, 0583186, 0585529, 0590858, 0593455, 0596125, 0600250, 0605504, 0609198, 0614183, 0620033, 0623325, 0626858, 0634096, 0636271, 0638011, 0640980, 0650900, 0653232, and 628985C. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1837-2012
SARNS FLEXIBLE ARTERIAL CANNULA: 7.3 MM (22 FR) 00 WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 13020 and Lot codes: 0550552, 0555854, 0557327, 0559103, 0562503, 0566861, 0570455, 0574279, 0577885, 0581606, 0585963, 0590856, 0593457, 0597308, 0605179, 0609199, 0614248, 0618969, 0626379, 0628950, 0634171, 0636332, 0640518, 0646396, 0653231, and 0666719. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1836-2012
SARNS FLEXIBLE ARTERIAL CANNULA: 6.7 MM (20 FR) 00 WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 13010 and Lot codes: 0590855, 0593791, 0597307, 0603718, 0616892, 0624831,0629341,0634937, 0636270, 0644444, 0646144, 0653230, 0666586, and 0669487. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1835-2012
SARNS HIGH-FLOW AORTIC ARCH CANNULA: 6.5 MM (20 FR) 00 WITH 3/8" CONNECTOR, 11" (28 CM) LONG, WITH XCOATING SURFACE COATING. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 12325X and Lot codes: 558946, 0589263, 0603353, 0640527, and 0653443. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1834-2012
SARNS HIGH-FLOW AORTIC ARCH CANNULA: 6.5 MM (20 FR) 00 WITH 3/8" CONNECTOR, 11" (28 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 12325 and Lot codes: 0551680, 0553010, 0557571, 0559882, 0562967, 0566831, 0569995, 0575870, 0579316, 0584969, 0589928, 0593454, 0595251, 0601331, 0606816, 0611597, 0617494, 0622167, 0626795, 0629156, 0632515, 0636330, 0640190, 0643858, 0649159, 0652241, 0656575, 0663510, and 621463C. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1833-2012
SARNS HIGH-FLOW AORTIC ARCH CANNULA: 5.2 MM (16 FR) 00 WITH 3/8" CONNECTOR, 7.5" (19 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Catalog number: 12315 and Lot codes: 0552019, 0557299, 0562142, 0572238, 0575869, 0579315, 0584831, 0590890, 0594622, 0598430, 0604045, 0612256, 0625339, 0640189, 0646473, 0652905, 0656574, 0666574, 0671713, and 593606C. Class II During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the particulate is still undetermined and while Terumo CVS has no other documented instances of this substance appearing during production, it has determined that any unit manufactured within the last three years on the same mold could potentially be affected. The particulate can be dislodged from the connector surface and either adhere to the gloves of the clinician in the sterile field, or fall into the pericardial well while the surgeon is connecting the cannula to the arterial line of the cardiopulmonary bypass circuit. 133,793 (each) Z-1832-2012
-
-