Event Detail

Event ID 61913
Product Type Devices
Status Ongoing
Recalling Firm Spacelabs Healthcare, Llc
City Issaquah
State WA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-11
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message. Version 1.6.0. Class II The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker. 114 Z-1831-2012
-
-