Enforcement Report - Week of July 18, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 62302
Product Type Devices
Status Ongoing
Recalling Firm Impac Medical Systems Inc
City Sunnyvale
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-08
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution -- USA, including the states of VA, CT, GA, NJ and NY and the countries of Belgium, Denmark, France, and the United Kingdom.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086 Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7. Class II Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero. 11 affected sites Z-1944-2012
-
-