Enforcement Report - Week of July 5, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62331 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Hill-Rom, Inc. |
| City | Batesville |
| State | IN |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-18 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution-USA (nationwide) and the countries of Canada, France, and Singapore. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital. | AdvantaTM 2 beds (P1190) produced between August 10, 2009 and October 6, 2011; bed S/N between K222AT9180 and M279AT4403 | Class II | During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail. | 7843 devices | Z-1864-2012 |
-