Enforcement Report - Week of August 8, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62371 |
|---|---|
| Product Type | Devices |
| Status | Completed |
| Recalling Firm | Abbott Point Of Care Inc. |
| City | Princeton |
| State | NJ |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2011-01-01 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | USA ( nationwide ) |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood. | List number 03P76-25 Lot number S10246A | Class II | i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer. | 52,750 cartridges | Z-2133-2012 |
-