Enforcement Report - Week of August 15, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62448 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
| City | Newark |
| State | DE |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-07 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide distribution, including the countries of Australia, Austria, Belgium, Canada, Czech Republic, France, Germany, Italy, Netherlands, New Zealand, Slovakia, Spain, Switzerland and the United States, with nationwide distribution. Center Recall Depth is Retail. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| EasyLink(TM) Informatics System, Part Number 1000034805. EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-Vitro diagnostics devices. | Version 5.0 and Version 5.0 Service Pack 1 through 4 | Class II | There is a potential for the EasyLink system to display and transmit to the Laboratory Information System networks an incorrect result following the EasyLink systems application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex CA-7000 or CA-1500 coagulation analyzer does not match the EasyLink system settings on the following page. | 237 | Z-2172-2012 |
-