Event Detail

Event ID 62533
Product Type Devices
Status Ongoing
Recalling Firm Abbott Point Of Care Inc.
City Princeton
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-02-01
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Columbia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraquay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
i-Stat CK-MP cartridges Abbott Point of Care Inc. Abbott Park, IL 60064 USA The i-Stat CKMB test is an in vitro diagnostic test for the quantitative measurement of creatinine kinase MB mass in whole blood or plasma samples. CK-MB measurements can be used as an aid in the diagnosis and treatment of myocardial infarction (MI). List number 06F25-01, 06F25-02, 03P92-25, All lots Class II Abbott Point of Care has determined that the frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure. 368,149 cartridges US; 32,725 OUS Z-2171-2012
i-Stat BNP cartridges Abbott Point of Care Inc. Abbott Park, IL 60064 USA The i-Stat BNP test is an in vitro diagnostic test for the quantitative measurement of B-type Natriuertic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure. List number 06F30-01, 06F30-02, 03P93-25, 600-9010-25 All lots Class II Abbott Point of Care has determined that the frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure. 1,004,200 cartridges US; 54,200 cartridges OUS Z-2170-2012
i-Stat cTnl cartridges Abbott Point of Care Inc. Abbott Park, IL 60064 USA The i-Stat cTnl test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in heparinized whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. The cartridge is to be used with the i-Stat Analyzer, but not with the i-Stat Portable Clinical Analyzer or the Philips Medical System (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-Stat System, the cTnl test is to be used by trained health care professionals in accordance with a facility's policies and procedures. List number 06F15-03, 06F15-04, 03P90-23, 600-9009-25 All lots Class II Abbott Point of Care has determined that the frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure. 7,057,675 cartridges US ; 772,400 OUS Z-2169-2012
-
-