• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 1, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 62534
Product Type Devices
Status Ongoing
Recalling Firm AGFA Corp.
City Greenville
State SC
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-07-10
Initial Firm Notification of
Consignee or Public
Telephone
Distribution Pattern Nationwide Distribution including IN, SC, OH, TX

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
IMPAX CV Reporting The Results Management (RM)/IMPAX CV reporting component facilitates the quick, effective creating of digital structured reports for : adult catheterization and echocardiography, pediatric echocardiography, congenital heat disease, nuclear cardiology and non-invasive vascular disease management. This webb-based application allows cardiology departments to view, edit and sign the structured reports. Software Version: RM 2.03 to 2.08, NIV 1.0 Class II When users selected "Left stenosis" in the "Graft Duplex Conclusion" section in IMPAX CV Reporting Non-Invasive(NIV) module, "Right stenosis" appeared in the sentence display within the printed representation of the report. 6 Z-2069-2012
-
-