Event Detail

Event ID 62564
Product Type Devices
Status Ongoing
Recalling Firm Atlas Spine, Inc.
City Jupiter
State FL
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2009-06-04
Initial Firm Notification of
Consignee or Public
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern USA Distribution to the state of Florida only.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Apelo Modular Screwdriver labeled in part: "APELO PEDICLE SCREW SYSTEM***Modular Screwdriver***Atlas Spine, Inc.***www.AtlasSpine.com*** 1555 Jupiter Park Drive #4***Jupiter, Florida 33458***561-741-1108***Tissue Sleeve***MFG Date: XX/XX/XXXX***Lot #: XXXX***Weldment***MFG Date: 03/23/2009***Sleeve Assembly***MFG Date: XX/XX/XXXX***Lot #: XXXX***Qty: 1 Rx Only***Non-Sterile. Sterilize Prior to Use*** Rev. A***. The Modular Screwdriver is used to insert modular pedicle screw implants which are part of the Apelo System. PN# 70175-001 CAT#2006-41-0000 Lot # 00686 Class II Atlas Spine, Inc. recalled their Apelo Modular Screwdriver, a component of the Apelo Pedicle Screw System Instrumentation, because of complaints received about the tip breaking. If the tip is fractured inside of the modular screw, the tip could become implanted in the body or if the entire screw is removed, it could cause a delay in the surgery. 4 pieces Z-2274-2012
-
-