Enforcement Report - Week of August 8, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62566 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Exactech, Inc. |
| City | Gainesville |
| State | FL |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-15 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Equinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty | Sizes: 1.5 mm: Catalog Number 300-10-15 Serial Range 1653131-2381151 4.5 mm: Catalog Number 300-10-45 Serial Range 1622046-2381151 | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2115-2012 |
| Equinoxe Fixed Angle Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty. | Size: 0mm Catalog Number: 300-25-00 Serial Range: 1999574-2381151 | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2116-2012 |
| Equinoxe Reverse Compression Screw Intended for Primary and Reverse Total Shoulder Arthroplasty. | Sizes: 4.5x18mm Catalog # 320-20-18 Serial Range 1610445-2381151 4.5x22mm Catalog # 320-20-22 Serial Range 1610391-2381151 4.5x26mm Catalog # 320-20-26 Serial Range 1611237-2381151 4.5x30mm Catalog # 320-20-30 Serial Range 1617556-2381151 4.5x34mm Catalog # 320-20-34 Serial Range 1611290-2381151 4.5x38mm Catalog # 320-20-38 Serial Range 1610260-2381151 4.5x42mm Catalog # 320-20-42 Serial Range 1630006-2381151 4.5x46mm Catalog # 320-20-46 Serial Range 1644317-2381151 | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2117-2012 |
| Equinoxe Reverse Shoulder Fracture Adapter Tray Screw Intended for Primary and Reverse Total Shoulder Arthroplasty. | Catalog Number 324-11-00 Serial Range 1653706-2381151 | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2118-2012 |
| Equinoxe Reverse Glenosphere Locking Screw Intended for Primary and Reverse Total Shoulder Arthroplasty. | Catalog Number 324-15-05 Serial Range 1620549-2381151 | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2119-2012 |
| Octane-C Cervical, 6ยบ Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7). | Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catalog # 05-010-02-0607 All serial numbers 8 mm Catalog # 05-010-02-0608 All serial numbers 9 mm Catalog # 05-010-02-0609 All serial numbers 10mm Catalog # 05-010-02-0610 All serial numbers 11mm Catalog # 05-010-02-0611 All serial numbers 12mm Catalog # 05-010-02-0612 All serial numbers | Class II | Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year shelf life. | 54,877 units for all products in Recall Event | Z-2120-2012 |
-