Enforcement Report - Week of August 29, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62571 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Medtronic Navigation, Inc. |
| City | Littleton |
| State | MA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-07-11 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Medtronic O-armĀ® Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R. | Serial Numbers: 109R, 118R, 124, 136, 159 ,170, 218, 222, 223, 236, 239 , 299, 309R, 402, 410, 412, 414. | Class II | A fault in a high voltage relay in the 0-armĀ® Imaging System might cause the relay to overheat and burn. | 17 units | Z-2244-2012 |
-