Event Detail

Event ID 62586
Product Type Devices
Status Ongoing
Recalling Firm Roche Molecular Systems, Inc.
City Branchburg
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-02-01
Initial Firm Notification of
Consignee or Public
E-Mail
Distribution Pattern Nationwide Distribution (USA).

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample. Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503. Class II Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in process would be aborted and would require re-running. 2,132 boxes Z-2209-2012
-
-