Event Detail

Event ID 62648
Product Type Devices
Status Ongoing
Recalling Firm Stryker Spine
City Allendale
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-04-23
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution -- US, Switzerland, Netherlands, Brazil, Hong Kong, Germany, and Taiwan.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision. Lot Code: 093309 Class II Stryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu. 30 devices Z-2155-2012
-
-