• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of September 12, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 62702
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc
City East Walpole
State MA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-07-25
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) and the countries of Canada, Argentina, Australia, Brazil, Chile, China, Croatia, Denmark, Ecuduar, Egypt,Guatemala, Hong Kong,India,Ireland, Israel,India, Japan,Korea, Mexico, Malyasia, Peru, Pakistan, Saudi Arabia,Singapore, Thailand,Uruguay, Viet Nam, and South Africa.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Siemens ADVIA Centaur Multi-Diluent 1 - Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems. Catalog Numbers: 2 Pack REF 07907174 (110312) 6 Pack REF 07293184 (110313). Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems. Lot # Expiry Date 2 Pack 76129 22 JUN 2012 77542 22 JUN 2012 78696 22 JUN 2012 80828 26 SEP 2012 80857 26 SEP 2012 81441 26 SEP 2012 82189 26 SEP 2012 83647 26 SEP 2012 84772 26 SEP 2012 86037 13 JAN 2013 86783 13 JAN 2013 88712 13 JAN 2013 90056 13 JAN 2013 91002 13 JAN 2013 93126 13 JAN 2013 6 Pack 76102 22 JUN 2012 76432 22 JUN 2012 77075 22 JUN 2012 78264 22 JUN 2012 79037 22 JUN 2012 81006 26 SEP 2012 82539 26 SEP 2012 83321 26 SEP 2012 84123 26 SEP 2012 85795 13 JAN 2013 86698 13 JAN 2013 88318 13 JAN 2013 89286 13 JAN 2013 90490 13 JAN 2013 91672 13 JAN 2013 93129 13 JAN 2013 Class II Decrease in onboard dilution recovery when using Multi-Diluent 1 that has been stored on board the ADVIA Centaur and ADVIA Centaur XP systems for for BNP and TSH3 Ultra assays 2 Packs = 15879 and 6 Packs = 7414 Z-2313-2012
-
-