Event Detail

Event ID 62709
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc.
City Newark
State DE
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-29
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern USA ( nationwide ) including the states of AZ, CA, FL, IL, KY, MD, MA, MN, MS, MO, NE, NH, NJ, NY, NC, OH, PA, TX, VA and WV.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
syngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments. Version VA11B and all previously released versions. Class II When using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall notification. QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient results will not be held as expected. The problem is encountered under the following conditions: 1 - The QC package within the syngo(R) system is being used to evaluate QC; 2 - The syngo(R) system is configured with Multi-Rule QC, and the Multi-level option is selected without following the instructions on page 2 of the recall letter. 44 Z-2305-2012
-
-