Event Detail

Event ID 62738
Product Type Devices
Status Ongoing
Recalling Firm Roche Molecular Systems, Inc.
City Branchburg
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-04-23
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide (USA) Distribution

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Amplilink (AL) Software 3.2.3 for use with COBAS AMPLICOR, CAP/CA, COBAS TaqMan,and COBAS AmpliPrep/COBAS TaqMan. The software can be used with a Laboratory Information System. 05666678001 Class II When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order. 1292 instruments Z-2403-2012
-
-