Enforcement Report - Week of September 5, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62776 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Steris Corporation |
| City | Mentor |
| State | OH |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-08-10 |
| Initial Firm Notification of Consignee or Public | Visit |
| Distribution Pattern | Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor. | Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190 | Class II | STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field.. | 64 units | Z-2284-2012 |
-







