| Event ID | 62805 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | KCI USA, Inc. |
| City | San Antonio |
| State | TX |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-27 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution including AL, CA, CO, OK and GA. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction | Model Number 208030, Serial Numbers: 00157087, 00157091, 00157288, 00157375, 00157405, RRCKOO011, RRHKOO065, RRHKOO074, RRHKOO096, RRHK00109, and RRHKOO117. | Class II | The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers. | 282 units | Z-0409-2013 |
-