Event Detail

Event ID 62925
Product Type Devices
Status Ongoing
Recalling Firm Advanced Sterilization Products
City Irvine
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-19
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model. Product Code: 10113 Class III Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing. 7,495 units Z-2337-2012
STERRAD NX Cassette, Batch # 12A078. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model. Product Code: 10133 Class III Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing. 7,550 units Z-2338-2012
-
-