Enforcement Report - Week of September 12, 2012
-
Back to Previous Page
Event Detail
| Event ID | 62949 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Quality Electrodynamics LLC |
| City | Mayfield Village |
| State | OH |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-07-27 |
| Initial Firm Notification of Consignee or Public | Other |
| Distribution Pattern | Worldwide distribution: USA (nationwide): Tennessee only and country of: Germany |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| TxRx Knee 15 Coil 3T. For use in conjunction with magnetic resonance scanners to produce diagnostic images of the knee that can be interpreted by a trained physician. | QED part numbers: Q7000057/QED000057/QSV000057 Siemens part numbers: 10606829 / 10606614 Serial #s: 1002-1012, 1016-1020, and 1022. | Class II | On July 27, 2012, it was reported to the firm that the Siemens 3T 15ch Knee Coil was not functional because the coil identification code for the device was incorrect. | 17 units | Z-2314-2012 |
-