Event Detail
| Event ID | 63042 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Medical Solutions USA, Inc |
| City | Malvern |
| State | PA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-08-21 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Siemens Mobilett Mira. Mobile X-Ray System. | Model number 10273100 -- serial numbers 1034, 1038, 1041, 1042, 1043, 1053, 1057, 1070, 1079, 1081, 1088, 1098, 1099, 1102, 1103, 1105, 1106, 1107, 1108, and 1112. | Class II | During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to replace motor drive handles and circuit board connection cables on all affected Mobilett Mira systems. | 20 | Z-2322-2012 |
-