Event Detail

Event ID 63086
Product Type Devices
Status Ongoing
Recalling Firm The Anspach Effort, Inc.
City Palm Beach Gardens
State FL
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-01-09
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
"***Electric Systems Foot Control with Direction and Irrigation***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium. Serial Numbers: E3230327501, E32305327502, E32305327503, E32305327504, E32305327505, E32305327601, E32305327602, E32305327603, E32305327604, and E32305327605 Class II The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. 10 units Z-0076-2013
"***Electric Systems Foot Control with Direction Only***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium. Serial Numbers: E28305271001, E28305271002, E28305271003, E28305271004, E2830527100, E28305271006, E28305271007, E28305271008, E2830521009, E28305271010, E29305270901, E29305270902, E29305270903, E29305271101, E29305271102, E29305271103, E29305271104, E29305271105, E29305271106, E29305271107, E29, Class II The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. 41 units Z-0075-2013
"***Electric Systems Foot Control****Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium. Serial Numbers: E31305311301, E31305311302, E31305311303, E31305311304, E31305311401, E31305311402, E31305311403, E31305311404, E31305311405, E32305311201, E32305311202, E33305311501, E33305311502, E33305311504, and E33305311505 Class II The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. 16 units Z-0074-2013
-
-