Enforcement Report - Week of October 10, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63125 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Zimmer, Inc. |
| City | Warsaw |
| State | IN |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-09-07 |
| Initial Firm Notification of Consignee or Public | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
| Distribution Pattern | Worldwide Distribution - US Nationwide including the states of: AZ, CA, CO,FL, MN, IL, NJ, NY, OH, OR and the countries of Austrailia, Utd. Arab Emir., France, Germany, India, and Italy. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Cemented Tibial Drill Product Usage: The cemented tibial drill is used to prepare the tibial bone prior to the impaction of the applicable stemmed tibial broach. The cemented tibial drill has engraved lines that are size specific to guide the user as to the depth to insert the drill into the tibial bone. | Part 42-5399-018-00, Lots 62034571, 62089766 | Class II | Zimmer is initiating a recall of the Cemented Tibial Drill for the Zimmer Persona Personalized Knee System due to three reports being received as of August 29. 2012 for tibial cortex perforation during preparation of the tibial bone. The investigation into the reports that have been received found that the perforations are occurring while using the Cemented Tibial Drill. The perforations have occurred in the anterior medial tibial cortex, particularly when preparing for a size E tibial plate. As a result of this investigation, Zimmer is detailing below some important aspects within the surgical technique that mitigate the potential of perforating the tibial cortex. In addition, Zimmer is replacing the current Cemented Tibial Drill with a modified version that has relocated the depth indicators for determining how deep to drill for the different sizes of tibial plates. Concurrent with delivery of this notification, your sales representative is replacing your Cemented Tibial Drill. | 60 | Z-2493-2012 |
-