Event Detail

Event ID 63145
Product Type Devices
Status Completed
Recalling Firm SpineFrontier, Inc.
City Beverly
State MA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2011-12-23
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern US Nationwide - OR, FL, MA, TX, CO, OK, IL, and MO

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000. S-Lift Lock Shaft Part number; SI50001 Revisions: A, B Lots: 42613, 2478901R, 42615, 7625701 Class II It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal. Pin: The pin used to guide the Lock Shaft Inserter into the mating SI50002 Implant Holder can shear off under high-impact force during insertion and removal. 30 devices Z-1565-2014
-
-