Enforcement Report - Week of October 3, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63166 |
|---|---|
| Product Type | Devices |
| Status | Terminated |
| Recalling Firm | Toshiba American Medical Systems Inc |
| City | Tustin |
| State | CA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2008-05-13 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide distribution: USA including states of: MA, MD, and NV. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a gantry, patient couch (table) and peripheral cabinets used for data processing and display. This device is intended to produce cross sectional volume sets of the anatomy. Additionally the system, when used by a trained professional, is capable of proving data for dynamic and perfusion studies of organs and extremities. | Item number TSX-301 A/2D Serial Numbers: 2DA0792004, 2DA07X2005, 2DA07Y2010, 2DA0792002, 2DA07X2008, 2DA0842023. | Class II | The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function. | 3 units | Z-2468-2012 |
-