• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 5, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 63204
Product Type Devices
Status Ongoing
Recalling Firm Orthovita, Inc., dBA Stryker Orthobiologics.
City Malvern
State PA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-10
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0502 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2090-0502, Lot Number Q1006003 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 78 Z-1387-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0503, Lot Numbers E908003R and E908002 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 69 Z-1399-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 6", Part Number 2110-0524 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0524, Lot Number E909009R and E909009R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 12 Z-1398-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger, 8", Part Number 2110-0502 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0502, Lot Number E909008R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 17 Z-1397-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 6", Part Number 2110-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0504, Lot Number E909007R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 7 Z-1396-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 8", Part Number 2110-0525 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0525, Lot Numbers E909006R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 5 Z-1395-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Diamond Needle, 4", Part Number 2110-0506 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0506, Lot Number E909005R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 19 Z-1394-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger, 6", Part Number 2110-0501 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0501, Lot Numbers E909004R and E909004 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 414 Z-1393-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 8", Part Number 2110-0526 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0526, Lot Numbers E908004R and E908004 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 27 Z-1392-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 4", Part Number 2110-0505 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2110-0505, Lot Numbers E909002 and E909002R Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 35 Z-1391-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Syringe Delivery System, 6", Part Number 2090-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2090-0504, Lot Numbers A1006026 and A1111020 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 155 Z-1390-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0501 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2090-0501, Lot Numbers A1109035 and Q1006002 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 411 Z-1389-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 6", Part Number 2090-0505 Product Usage: Facilitate placement of bone cement in weakened or diminished bone Manufacturer Part Number 2090-0505, Lot Number A1006027 Class II There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. 164 Z-1388-2013
-
-