• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 14, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 63270
Product Type Devices
Status Ongoing
Recalling Firm Diagnostica Stago, Inc.
City Parsippany
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-15
Initial Firm Notification of
Consignee or Public
E-Mail
Distribution Pattern Worldwide Distribution - USA (nationwide) and Canada

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy. K082248 D089976 Ref 58104 units Class II Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite®. 232 units US; 5 units Canada Z-0252-2013
-
-