• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of December 19, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 63312
Product Type Devices
Status Ongoing
Recalling Firm Roche Molecular Systems, Inc.
City Branchburg
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-25
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
AMPLILINK software, For use with COBAS® AmpliPrep instrument, COBAS® TaqMan® analyzer, COBAS® TaqMan® 48 analyzer, COBAS® AMPLICOR® analyzer, and Cobas p 630 instrument AMPLILINK software is used to integrate all instruments and analyzers offering a single user interface for PCR testing. AMPLILINK version 3.3 and 3.2 Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 884 units Z-0484-2013
Roche Applied Science MagNA Pure LC Version 3.0. The MagNA Pure LC (version 3.0) is a sample preparation instrument that can be used with the LightCycler instrument for automated isolation of nucleic acids in various types of biological materials. MagNA Pure LC version 3.0 Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 816 units Z-0485-2013
Roche MagNA Pure 96 System Version 1.0 (Software Version 2.0) for in vitro diagnostic use. The MagNA Pure 96 System (versions 1.0 and 2.0) is a sample preparation instrument for automated isolation of nucleic acids in various types of biological materials. This is a general purpose instrument. There are no corresponding assays. Roche MagNA Pure 96 System Version 1.0 (Software Version 2.0) Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 13 units Z-0486-2013
Roche Molecular Biochemicals LightCycler 1.2 with software version 3.5. LightCycler 1.2 (software version 3.5) is a Rapid PCR analyzer with real-time detection and/or quantification of a target NA (nucleic acid), as well as post-PCR analysis of the amplified NA by melting curve analysis Roche Molecular Biochemicals LightCycler 1.2 with software version 3.5. Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 3,772 units Z-0487-2013
Roche LightCycler 2.0 Instrument with software version 4.1. LightCycler 2.0 Instrument (software version 4.1) is a Rapid PCR analyzer with real-time detection and/or quantification of a target NA (nucleic acid), as well as post-PCR analysis of the amplified NA by melting curve analysis. Roche LightCycler 2.0 Instrument with software version 4.1. Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 1,213 units Z-0488-2013
cobas® 8000 modular analyzer series The COBAS 8000 data manager modular analyzer series (all versions) is a is a fully automated, random-access, software- controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in-vitro determinations using a wide variety of tests. cobas® 8000 modular analyzer series Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 77 units Z-0489-2013
cobas® Integra 400/400 Plus (Software version 3.4) The COBAS INTEGRA 400/400 plus system (software version 3.4) is a fully automated, computerized chemistry analyzer. It is intended for in vitro quantitative or qualitative determination of a wide range of analytes in various body fluids. cobas® Integra 400/400 Plus (Software version 3.4) Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 855 units. Z-0490-2013
DataInnovations (Instrument Manager) (version 8.04) interfaces middleware between larger Clinical Chemistry / Immunology instrument systems and LIS systems. Instrument Manager version 8.04. Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 1,015 units Z-0491-2013
DataCare GM DataCara GM is an information management system that provides data management functionality for managing blood glucose testing data. DataCare GM Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 479 units Z-0492-2013
DataCare POC  is an information management system that provides data management functionality for managing point-of -care testing data. DataCare POC Class II Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . 126 units Z-0493-2013
-
-